GCVE Workshop - 22 September 2026 (14:00-18:00), Luxembourg Before The Vulnopticon Conference - Registration
Search

Find a vulnerability

Search criteria Use this form to refine search results.
Full-text search supports keyword queries with ranking and filtering.
You can combine vendor, product, and sources to narrow results.
Enable “Apply ordering” to sort by date instead of relevance.

    2 vulnerabilities by Sooma

    CVE-2026-85104 (GCVE-0-2026-85104)

    Vulnerability from nvd – Published: 2026-09-16 13:38 – Updated: 2026-09-17 19:10
    VLAI
    Title
    Brain stimulation parameters can be modified via Bluetooth in Sooma
    Summary
    In Sooma 2GEN brain stimulator, an attacker within Bluetooth range can make unauthenticated changes to brain stimulation parameters.
    SSVC
    Exploitation: none Automatable: no Technical Impact: partial
    CISA Coordinator · CISA-ADP (v2.0.3)
    Decision recorded 2026-09-17 19:10 UTC
    CWE
    • CWE-924 - Improper enforcement of message integrity during transmission in a communication channel
    References
    Impacted products
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "metrics": [
              {
                "other": {
                  "content": {
                    "id": "CVE-2026-85104",
                    "options": [
                      {
                        "Exploitation": "none"
                      },
                      {
                        "Automatable": "no"
                      },
                      {
                        "Technical Impact": "partial"
                      }
                    ],
                    "role": "CISA Coordinator",
                    "timestamp": "2026-09-17T19:10:32.914600Z",
                    "version": "2.0.3"
                  },
                  "type": "ssvc"
                }
              }
            ],
            "providerMetadata": {
              "dateUpdated": "2026-09-17T19:10:41.232Z",
              "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
              "shortName": "CISA-ADP"
            },
            "title": "CISA ADP Vulnrichment"
          }
        ],
        "cna": {
          "affected": [
            {
              "defaultStatus": "unknown",
              "product": "Sooma tDCS Home Therapy",
              "vendor": "Sooma",
              "versions": [
                {
                  "status": "affected",
                  "version": "2Gen"
                }
              ]
            }
          ],
          "credits": [
            {
              "lang": "en",
              "type": "finder",
              "value": "A. C. Buglione"
            }
          ],
          "descriptions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "\u0026nbsp;In Sooma 2GEN brain stimulator, an attacker within Bluetooth range can make unauthenticated changes to brain stimulation\nparameters."
                }
              ],
              "value": "In Sooma 2GEN brain stimulator, an attacker within Bluetooth range can make unauthenticated changes to brain stimulation\nparameters."
            }
          ],
          "impacts": [
            {
              "capecId": "CAPEC-212",
              "descriptions": [
                {
                  "lang": "en",
                  "value": "CAPEC-212 Functionality Misuse"
                }
              ]
            }
          ],
          "metrics": [
            {
              "cvssV4_0": {
                "Automatable": "NOT_DEFINED",
                "Recovery": "NOT_DEFINED",
                "Safety": "NOT_DEFINED",
                "attackComplexity": "LOW",
                "attackRequirements": "NONE",
                "attackVector": "ADJACENT",
                "baseScore": 5.3,
                "baseSeverity": "MEDIUM",
                "exploitMaturity": "NOT_DEFINED",
                "privilegesRequired": "NONE",
                "providerUrgency": "NOT_DEFINED",
                "subAvailabilityImpact": "NONE",
                "subConfidentialityImpact": "NONE",
                "subIntegrityImpact": "NONE",
                "userInteraction": "NONE",
                "valueDensity": "NOT_DEFINED",
                "vectorString": "CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:L/VA:L/SC:N/SI:N/SA:N",
                "version": "4.0",
                "vulnAvailabilityImpact": "LOW",
                "vulnConfidentialityImpact": "NONE",
                "vulnIntegrityImpact": "LOW",
                "vulnerabilityResponseEffort": "NOT_DEFINED"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-924",
                  "description": "CWE-924 Improper enforcement of message integrity during transmission in a communication channel",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2026-09-16T13:38:30.097Z",
            "orgId": "db4dfee8-a97e-4877-bfae-eba6d14a2166",
            "shortName": "NCSC-FI"
          },
          "references": [
            {
              "url": "https://shop.soomamedical.com/pages/all-products"
            }
          ],
          "source": {
            "discovery": "UNKNOWN"
          },
          "title": "Brain stimulation parameters can be modified via Bluetooth in Sooma",
          "x_generator": {
            "engine": "Vulnogram 1.0.5"
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "db4dfee8-a97e-4877-bfae-eba6d14a2166",
        "assignerShortName": "NCSC-FI",
        "cveId": "CVE-2026-85104",
        "datePublished": "2026-09-16T13:38:30.097Z",
        "dateReserved": "2026-09-03T06:44:36.452Z",
        "dateUpdated": "2026-09-17T19:10:41.232Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.2"
    }

    CVE-2026-85104 (GCVE-0-2026-85104)

    Vulnerability from cvelistv5 – Published: 2026-09-16 13:38 – Updated: 2026-09-17 19:10
    VLAI
    Title
    Brain stimulation parameters can be modified via Bluetooth in Sooma
    Summary
    In Sooma 2GEN brain stimulator, an attacker within Bluetooth range can make unauthenticated changes to brain stimulation parameters.
    SSVC
    Exploitation: none Automatable: no Technical Impact: partial
    CISA Coordinator · CISA-ADP (v2.0.3)
    Decision recorded 2026-09-17 19:10 UTC
    CWE
    • CWE-924 - Improper enforcement of message integrity during transmission in a communication channel
    References
    Impacted products
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "metrics": [
              {
                "other": {
                  "content": {
                    "id": "CVE-2026-85104",
                    "options": [
                      {
                        "Exploitation": "none"
                      },
                      {
                        "Automatable": "no"
                      },
                      {
                        "Technical Impact": "partial"
                      }
                    ],
                    "role": "CISA Coordinator",
                    "timestamp": "2026-09-17T19:10:32.914600Z",
                    "version": "2.0.3"
                  },
                  "type": "ssvc"
                }
              }
            ],
            "providerMetadata": {
              "dateUpdated": "2026-09-17T19:10:41.232Z",
              "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
              "shortName": "CISA-ADP"
            },
            "title": "CISA ADP Vulnrichment"
          }
        ],
        "cna": {
          "affected": [
            {
              "defaultStatus": "unknown",
              "product": "Sooma tDCS Home Therapy",
              "vendor": "Sooma",
              "versions": [
                {
                  "status": "affected",
                  "version": "2Gen"
                }
              ]
            }
          ],
          "credits": [
            {
              "lang": "en",
              "type": "finder",
              "value": "A. C. Buglione"
            }
          ],
          "descriptions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "\u0026nbsp;In Sooma 2GEN brain stimulator, an attacker within Bluetooth range can make unauthenticated changes to brain stimulation\nparameters."
                }
              ],
              "value": "In Sooma 2GEN brain stimulator, an attacker within Bluetooth range can make unauthenticated changes to brain stimulation\nparameters."
            }
          ],
          "impacts": [
            {
              "capecId": "CAPEC-212",
              "descriptions": [
                {
                  "lang": "en",
                  "value": "CAPEC-212 Functionality Misuse"
                }
              ]
            }
          ],
          "metrics": [
            {
              "cvssV4_0": {
                "Automatable": "NOT_DEFINED",
                "Recovery": "NOT_DEFINED",
                "Safety": "NOT_DEFINED",
                "attackComplexity": "LOW",
                "attackRequirements": "NONE",
                "attackVector": "ADJACENT",
                "baseScore": 5.3,
                "baseSeverity": "MEDIUM",
                "exploitMaturity": "NOT_DEFINED",
                "privilegesRequired": "NONE",
                "providerUrgency": "NOT_DEFINED",
                "subAvailabilityImpact": "NONE",
                "subConfidentialityImpact": "NONE",
                "subIntegrityImpact": "NONE",
                "userInteraction": "NONE",
                "valueDensity": "NOT_DEFINED",
                "vectorString": "CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:L/VA:L/SC:N/SI:N/SA:N",
                "version": "4.0",
                "vulnAvailabilityImpact": "LOW",
                "vulnConfidentialityImpact": "NONE",
                "vulnIntegrityImpact": "LOW",
                "vulnerabilityResponseEffort": "NOT_DEFINED"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-924",
                  "description": "CWE-924 Improper enforcement of message integrity during transmission in a communication channel",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2026-09-16T13:38:30.097Z",
            "orgId": "db4dfee8-a97e-4877-bfae-eba6d14a2166",
            "shortName": "NCSC-FI"
          },
          "references": [
            {
              "url": "https://shop.soomamedical.com/pages/all-products"
            }
          ],
          "source": {
            "discovery": "UNKNOWN"
          },
          "title": "Brain stimulation parameters can be modified via Bluetooth in Sooma",
          "x_generator": {
            "engine": "Vulnogram 1.0.5"
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "db4dfee8-a97e-4877-bfae-eba6d14a2166",
        "assignerShortName": "NCSC-FI",
        "cveId": "CVE-2026-85104",
        "datePublished": "2026-09-16T13:38:30.097Z",
        "dateReserved": "2026-09-03T06:44:36.452Z",
        "dateUpdated": "2026-09-17T19:10:41.232Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.2"
    }