FKIE_CVE-2021-4479

Vulnerability from fkie_nvd - Published: 2026-06-02 20:16 - Updated: 2026-07-22 19:10
Summary
Dräger Atlan A350 versions 1.00 up to and including 1.01 contains an improper input handling vulnerability that allows attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can transmit malformed data to overload the internal processor, gradually disrupting device operation over several hours and causing loss of data transmission, delayed display of real-time curves, and deviation between displayed airway pressure values and screen curves.
Impacted products
Vendor Product Version

{
  "affected": [
    {
      "affectedData": [
        {
          "defaultStatus": "unknown",
          "product": "Atlan A350",
          "vendor": "Dr\u00e4ger",
          "versions": [
            {
              "lessThanOrEqual": "1.01",
              "status": "affected",
              "version": "1.00",
              "versionType": "custom"
            }
          ]
        }
      ],
      "source": "disclosure@vulncheck.com"
    }
  ],
  "cveTags": [],
  "descriptions": [
    {
      "lang": "en",
      "value": "Dr\u00e4ger Atlan A350 versions 1.00 up to and including 1.01 contains an improper input handling vulnerability that allows attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can transmit malformed data to overload the internal processor, gradually disrupting device operation over several hours and causing loss of data transmission, delayed display of real-time curves, and deviation between displayed airway pressure values and screen curves."
    },
    {
      "lang": "es",
      "value": "Las versiones de software de Dr\u00e4ger Atlan A350 1.00 a 1.01 contienen una vulnerabilidad de manejo de entrada incorrecto que permite a los atacantes causar una denegaci\u00f3n de servicio mediante el env\u00edo de datos espec\u00edficamente dise\u00f1ados y no conformes con Medibus a trav\u00e9s de la interfaz Medibus. Los atacantes pueden transmitir datos malformados para sobrecargar el procesador interno, interrumpiendo gradualmente el funcionamiento del dispositivo durante varias horas y causando la p\u00e9rdida de transmisi\u00f3n de datos, el retraso en la visualizaci\u00f3n de curvas en tiempo real y la desviaci\u00f3n entre los valores de presi\u00f3n de las v\u00edas respiratorias mostrados y las curvas en pantalla."
    }
  ],
  "id": "CVE-2021-4479",
  "lastModified": "2026-07-22T19:10:00.120",
  "metrics": {
    "cvssMetricV31": [
      {
        "cvssData": {
          "attackComplexity": "HIGH",
          "attackVector": "NETWORK",
          "availabilityImpact": "LOW",
          "baseScore": 4.0,
          "baseSeverity": "MEDIUM",
          "confidentialityImpact": "NONE",
          "integrityImpact": "NONE",
          "privilegesRequired": "NONE",
          "scope": "CHANGED",
          "userInteraction": "NONE",
          "vectorString": "CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:N/I:N/A:L",
          "version": "3.1"
        },
        "exploitabilityScore": 2.2,
        "impactScore": 1.4,
        "source": "disclosure@vulncheck.com",
        "type": "Secondary"
      }
    ],
    "cvssMetricV40": [
      {
        "cvssData": {
          "Automatable": "NOT_DEFINED",
          "Recovery": "NOT_DEFINED",
          "Safety": "NOT_DEFINED",
          "attackComplexity": "HIGH",
          "attackRequirements": "NONE",
          "attackVector": "NETWORK",
          "availabilityRequirement": "NOT_DEFINED",
          "baseScore": 6.3,
          "baseSeverity": "MEDIUM",
          "confidentialityRequirement": "NOT_DEFINED",
          "exploitMaturity": "NOT_DEFINED",
          "integrityRequirement": "NOT_DEFINED",
          "modifiedAttackComplexity": "NOT_DEFINED",
          "modifiedAttackRequirements": "NOT_DEFINED",
          "modifiedAttackVector": "NOT_DEFINED",
          "modifiedPrivilegesRequired": "NOT_DEFINED",
          "modifiedSubAvailabilityImpact": "NOT_DEFINED",
          "modifiedSubConfidentialityImpact": "NOT_DEFINED",
          "modifiedSubIntegrityImpact": "NOT_DEFINED",
          "modifiedUserInteraction": "NOT_DEFINED",
          "modifiedVulnAvailabilityImpact": "NOT_DEFINED",
          "modifiedVulnConfidentialityImpact": "NOT_DEFINED",
          "modifiedVulnIntegrityImpact": "NOT_DEFINED",
          "privilegesRequired": "NONE",
          "providerUrgency": "NOT_DEFINED",
          "subAvailabilityImpact": "LOW",
          "subConfidentialityImpact": "NONE",
          "subIntegrityImpact": "NONE",
          "userInteraction": "NONE",
          "valueDensity": "NOT_DEFINED",
          "vectorString": "CVSS:4.0/AV:N/AC:H/AT:N/PR:N/UI:N/VC:N/VI:N/VA:L/SC:N/SI:N/SA:L/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X",
          "version": "4.0",
          "vulnAvailabilityImpact": "LOW",
          "vulnConfidentialityImpact": "NONE",
          "vulnIntegrityImpact": "NONE",
          "vulnerabilityResponseEffort": "NOT_DEFINED"
        },
        "source": "disclosure@vulncheck.com",
        "type": "Secondary"
      }
    ],
    "ssvcV203": [
      {
        "source": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
        "ssvcData": {
          "id": "CVE-2021-4479",
          "options": [
            {
              "exploitation": "none"
            },
            {
              "automatable": "no"
            },
            {
              "technicalImpact": "partial"
            }
          ],
          "role": "CISA Coordinator",
          "timestamp": "2026-06-03T12:50:30.624302Z",
          "version": "2.0.3"
        }
      }
    ]
  },
  "published": "2026-06-02T20:16:28.383",
  "references": [
    {
      "source": "disclosure@vulncheck.com",
      "url": "https://static.draeger.com/security/download/2021-05-04-PSA-21-124-1-Security-Advisory-Atlan.pdf"
    },
    {
      "source": "disclosure@vulncheck.com",
      "url": "https://www.vulncheck.com/advisories/drager-atlan-a350-dos-via-medibus-interface"
    }
  ],
  "sourceIdentifier": "disclosure@vulncheck.com",
  "vulnStatus": "Awaiting Analysis",
  "weaknesses": [
    {
      "description": [
        {
          "lang": "en",
          "value": "CWE-1286"
        }
      ],
      "source": "disclosure@vulncheck.com",
      "type": "Secondary"
    }
  ]
}



Log in or create an account to share your comment.




Tags
Taxonomy of the tags.


Loading…

Loading…

Loading…

Forecast uses a logistic model when the trend is rising, or an exponential decay model when the trend is falling. Fitted via linearized least squares.

Sightings

Author Source Type Date Other

Nomenclature

  • Seen: The vulnerability was mentioned, discussed, or observed by the user.
  • Confirmed: The vulnerability has been validated from an analyst's perspective.
  • Published Proof of Concept: A public proof of concept is available for this vulnerability.
  • Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
  • Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
  • Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
  • Not confirmed: The user expressed doubt about the validity of the vulnerability.
  • Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.

Loading…

Detection rules are retrieved from Rulezet.

Loading…

Loading…

Loading…