CVE-2019-25719 (GCVE-0-2019-25719)
Vulnerability from cvelistv5 – Published: 2026-06-02 14:00 – Updated: 2026-06-03 15:43
VLAI
EPSS
VEX
Title
Dräger Infinity M540 VG4.1.1 Spoofing and DoS via Network Message Handling
Summary
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.
Severity
SSVC
Exploitation: none
Automatable: yes
Technical Impact: partial
CISA Coordinator (v2.0.3)
CWE
- CWE-924 - Improper Enforcement of Message Integrity During Transmission in a Communication Channel
Assigner
References
2 references
| URL | Tags |
|---|---|
| https://static.draeger.com/security/download/PSA-… | vendor-advisory |
| https://www.vulncheck.com/advisories/dr-ger-infin… | third-party-advisory |
Impacted products
2 products
| Vendor | Product | Version | |
|---|---|---|---|
| Dräger | Infinity Acute Care System |
Affected:
0 , ≤ VG4.1.1
(custom)
Affected: 0 , ≤ VG4.0.3 (custom) Unaffected: VG4.2 (custom) |
|
| Dräger | Standalone Infinity M540 patient monitor |
Affected:
0 , ≤ VG4.1.1
(custom)
Unaffected: VG4.2 (custom) |
Date Public
2019-09-12 00:00
{
"containers": {
"adp": [
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2019-25719",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "yes"
},
{
"Technical Impact": "partial"
}
],
"role": "CISA Coordinator",
"timestamp": "2026-06-02T14:56:18.300929Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2026-06-02T14:56:35.870Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "unknown",
"product": "Infinity Acute Care System",
"vendor": "Dr\u00e4ger",
"versions": [
{
"lessThanOrEqual": "VG4.1.1",
"status": "affected",
"version": "0",
"versionType": "custom"
},
{
"lessThanOrEqual": "VG4.0.3",
"status": "affected",
"version": "0",
"versionType": "custom"
},
{
"status": "unaffected",
"version": "VG4.2",
"versionType": "custom"
}
]
},
{
"defaultStatus": "unknown",
"product": "Standalone Infinity M540 patient monitor",
"vendor": "Dr\u00e4ger",
"versions": [
{
"lessThanOrEqual": "VG4.1.1",
"status": "affected",
"version": "0",
"versionType": "custom"
},
{
"status": "unaffected",
"version": "VG4.2",
"versionType": "custom"
}
]
}
],
"datePublic": "2019-09-12T00:00:00.000Z",
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "\u003cp\u003eDr\u00e4ger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.\u003c/p\u003e"
}
],
"value": "Dr\u00e4ger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality."
}
],
"metrics": [
{
"cvssV4_0": {
"Automatable": "NOT_DEFINED",
"Recovery": "NOT_DEFINED",
"Safety": "NOT_DEFINED",
"attackComplexity": "LOW",
"attackRequirements": "NONE",
"attackVector": "NETWORK",
"baseScore": 8.8,
"baseSeverity": "HIGH",
"exploitMaturity": "NOT_DEFINED",
"privilegesRequired": "NONE",
"providerUrgency": "NOT_DEFINED",
"subAvailabilityImpact": "NONE",
"subConfidentialityImpact": "NONE",
"subIntegrityImpact": "NONE",
"userInteraction": "NONE",
"valueDensity": "NOT_DEFINED",
"vectorString": "CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N",
"version": "4.0",
"vulnAvailabilityImpact": "HIGH",
"vulnConfidentialityImpact": "LOW",
"vulnIntegrityImpact": "LOW",
"vulnerabilityResponseEffort": "NOT_DEFINED"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
},
{
"cvssV3_1": {
"attackComplexity": "LOW",
"attackVector": "NETWORK",
"availabilityImpact": "HIGH",
"baseScore": 8.6,
"baseSeverity": "HIGH",
"confidentialityImpact": "LOW",
"integrityImpact": "LOW",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H",
"version": "3.1"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-924",
"description": "CWE-924 Improper Enforcement of Message Integrity During Transmission in a Communication Channel",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2026-06-03T15:43:03.688Z",
"orgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
"shortName": "VulnCheck"
},
"references": [
{
"tags": [
"vendor-advisory"
],
"url": "https://static.draeger.com/security/download/PSA-19-255-02_Product-Security-Advisory-IACS-VG4.1.pdf"
},
{
"tags": [
"third-party-advisory"
],
"url": "https://www.vulncheck.com/advisories/dr-ger-infinity-m540-vg4-spoofing-and-dos-via-network-message-handling"
}
],
"source": {
"discovery": "EXTERNAL"
},
"title": "Dr\u00e4ger Infinity M540 VG4.1.1 Spoofing and DoS via Network Message Handling",
"x_generator": {
"engine": "Vulnogram 1.0.2"
}
}
},
"cveMetadata": {
"assignerOrgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
"assignerShortName": "VulnCheck",
"cveId": "CVE-2019-25719",
"datePublished": "2026-06-02T14:00:52.578Z",
"dateReserved": "2026-06-02T13:54:01.021Z",
"dateUpdated": "2026-06-03T15:43:03.688Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.2",
"vulnerability-lookup:meta": {
"epss": {
"cve": "CVE-2019-25719",
"date": "2026-07-22",
"epss": "0.00132",
"percentile": "0.0315"
},
"nvd": "{\"cve\":{\"id\":\"CVE-2019-25719\",\"sourceIdentifier\":\"disclosure@vulncheck.com\",\"published\":\"2026-06-02T14:16:25.627\",\"lastModified\":\"2026-07-22T19:10:00.120\",\"vulnStatus\":\"Awaiting Analysis\",\"cveTags\":[],\"descriptions\":[{\"lang\":\"en\",\"value\":\"Dr\u00e4ger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.\"},{\"lang\":\"es\",\"value\":\"Los monitores de paciente Dr\u00e4ger Infinity Acute Care System y Standalone Infinity M540 que ejecutan versiones de software VG4.1.1, VG4.0.3 y anteriores contienen vulnerabilidades en el manejo de mensajes de red que permiten a atacantes adyacentes a la red suplantar o manipular datos y causar condiciones de denegaci\u00f3n de servicio. Los atacantes con acceso a un puerto de red Infinity habilitado o proximidad f\u00edsica a un punto de acceso inal\u00e1mbrico pueden modificar la configuraci\u00f3n del dispositivo, como los estados de alarma o los l\u00edmites de alarma, y sobrecargar el sistema con datos entrantes, lo que provoca que el dispositivo se reinicie y pierda la funcionalidad de red.\"}],\"affected\":[{\"source\":\"disclosure@vulncheck.com\",\"affectedData\":[{\"vendor\":\"Dr\u00e4ger\",\"product\":\"Infinity Acute Care System\",\"defaultStatus\":\"unknown\",\"versions\":[{\"version\":\"0\",\"lessThanOrEqual\":\"VG4.1.1\",\"versionType\":\"custom\",\"status\":\"affected\"},{\"version\":\"0\",\"lessThanOrEqual\":\"VG4.0.3\",\"versionType\":\"custom\",\"status\":\"affected\"},{\"version\":\"VG4.2\",\"versionType\":\"custom\",\"status\":\"unaffected\"}]},{\"vendor\":\"Dr\u00e4ger\",\"product\":\"Standalone Infinity M540 patient monitor\",\"defaultStatus\":\"unknown\",\"versions\":[{\"version\":\"0\",\"lessThanOrEqual\":\"VG4.1.1\",\"versionType\":\"custom\",\"status\":\"affected\"},{\"version\":\"VG4.2\",\"versionType\":\"custom\",\"status\":\"unaffected\"}]}]}],\"metrics\":{\"cvssMetricV40\":[{\"source\":\"disclosure@vulncheck.com\",\"type\":\"Secondary\",\"cvssData\":{\"version\":\"4.0\",\"vectorString\":\"CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X\",\"baseScore\":8.8,\"baseSeverity\":\"HIGH\",\"attackVector\":\"NETWORK\",\"attackComplexity\":\"LOW\",\"attackRequirements\":\"NONE\",\"privilegesRequired\":\"NONE\",\"userInteraction\":\"NONE\",\"vulnConfidentialityImpact\":\"LOW\",\"vulnIntegrityImpact\":\"LOW\",\"vulnAvailabilityImpact\":\"HIGH\",\"subConfidentialityImpact\":\"NONE\",\"subIntegrityImpact\":\"NONE\",\"subAvailabilityImpact\":\"NONE\",\"exploitMaturity\":\"NOT_DEFINED\",\"confidentialityRequirement\":\"NOT_DEFINED\",\"integrityRequirement\":\"NOT_DEFINED\",\"availabilityRequirement\":\"NOT_DEFINED\",\"modifiedAttackVector\":\"NOT_DEFINED\",\"modifiedAttackComplexity\":\"NOT_DEFINED\",\"modifiedAttackRequirements\":\"NOT_DEFINED\",\"modifiedPrivilegesRequired\":\"NOT_DEFINED\",\"modifiedUserInteraction\":\"NOT_DEFINED\",\"modifiedVulnConfidentialityImpact\":\"NOT_DEFINED\",\"modifiedVulnIntegrityImpact\":\"NOT_DEFINED\",\"modifiedVulnAvailabilityImpact\":\"NOT_DEFINED\",\"modifiedSubConfidentialityImpact\":\"NOT_DEFINED\",\"modifiedSubIntegrityImpact\":\"NOT_DEFINED\",\"modifiedSubAvailabilityImpact\":\"NOT_DEFINED\",\"Safety\":\"NOT_DEFINED\",\"Automatable\":\"NOT_DEFINED\",\"Recovery\":\"NOT_DEFINED\",\"valueDensity\":\"NOT_DEFINED\",\"vulnerabilityResponseEffort\":\"NOT_DEFINED\",\"providerUrgency\":\"NOT_DEFINED\"}}],\"cvssMetricV31\":[{\"source\":\"disclosure@vulncheck.com\",\"type\":\"Secondary\",\"cvssData\":{\"version\":\"3.1\",\"vectorString\":\"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H\",\"baseScore\":8.6,\"baseSeverity\":\"HIGH\",\"attackVector\":\"NETWORK\",\"attackComplexity\":\"LOW\",\"privilegesRequired\":\"NONE\",\"userInteraction\":\"NONE\",\"scope\":\"UNCHANGED\",\"confidentialityImpact\":\"LOW\",\"integrityImpact\":\"LOW\",\"availabilityImpact\":\"HIGH\"},\"exploitabilityScore\":3.9,\"impactScore\":4.7}],\"ssvcV203\":[{\"source\":\"134c704f-9b21-4f2e-91b3-4a467353bcc0\",\"ssvcData\":{\"timestamp\":\"2026-06-02T14:56:18.300929Z\",\"id\":\"CVE-2019-25719\",\"options\":[{\"exploitation\":\"none\"},{\"automatable\":\"yes\"},{\"technicalImpact\":\"partial\"}],\"role\":\"CISA Coordinator\",\"version\":\"2.0.3\"}}]},\"weaknesses\":[{\"source\":\"disclosure@vulncheck.com\",\"type\":\"Secondary\",\"description\":[{\"lang\":\"en\",\"value\":\"CWE-924\"}]}],\"references\":[{\"url\":\"https://static.draeger.com/security/download/PSA-19-255-02_Product-Security-Advisory-IACS-VG4.1.pdf\",\"source\":\"disclosure@vulncheck.com\"},{\"url\":\"https://www.vulncheck.com/advisories/dr-ger-infinity-m540-vg4-spoofing-and-dos-via-network-message-handling\",\"source\":\"disclosure@vulncheck.com\"}]}}",
"vulnrichment": {
"containers": "{\"adp\": [{\"title\": \"CISA ADP Vulnrichment\", \"metrics\": [{\"other\": {\"type\": \"ssvc\", \"content\": {\"id\": \"CVE-2019-25719\", \"role\": \"CISA Coordinator\", \"options\": [{\"Exploitation\": \"none\"}, {\"Automatable\": \"yes\"}, {\"Technical Impact\": \"partial\"}], \"version\": \"2.0.3\", \"timestamp\": \"2026-06-02T14:56:18.300929Z\"}}}], \"providerMetadata\": {\"orgId\": \"134c704f-9b21-4f2e-91b3-4a467353bcc0\", \"shortName\": \"CISA-ADP\", \"dateUpdated\": \"2026-06-02T14:56:27.160Z\"}}], \"cna\": {\"title\": \"Dr\\u00e4ger Infinity M540 VG4.1.1 Spoofing and DoS via Network Message Handling\", \"source\": {\"discovery\": \"EXTERNAL\"}, \"metrics\": [{\"format\": \"CVSS\", \"cvssV4_0\": {\"Safety\": \"NOT_DEFINED\", \"version\": \"4.0\", \"Recovery\": \"NOT_DEFINED\", \"baseScore\": 8.8, \"Automatable\": \"NOT_DEFINED\", \"attackVector\": \"NETWORK\", \"baseSeverity\": \"HIGH\", \"valueDensity\": \"NOT_DEFINED\", \"vectorString\": \"CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N\", \"exploitMaturity\": \"NOT_DEFINED\", \"providerUrgency\": \"NOT_DEFINED\", \"userInteraction\": \"NONE\", \"attackComplexity\": \"LOW\", \"attackRequirements\": \"NONE\", \"privilegesRequired\": \"NONE\", \"subIntegrityImpact\": \"NONE\", \"vulnIntegrityImpact\": \"LOW\", \"subAvailabilityImpact\": \"NONE\", \"vulnAvailabilityImpact\": \"HIGH\", \"subConfidentialityImpact\": \"NONE\", \"vulnConfidentialityImpact\": \"LOW\", \"vulnerabilityResponseEffort\": \"NOT_DEFINED\"}, \"scenarios\": [{\"lang\": \"en\", \"value\": \"GENERAL\"}]}, {\"format\": \"CVSS\", \"cvssV3_1\": {\"scope\": \"UNCHANGED\", \"version\": \"3.1\", \"baseScore\": 8.6, \"attackVector\": \"NETWORK\", \"baseSeverity\": \"HIGH\", \"vectorString\": \"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H\", \"integrityImpact\": \"LOW\", \"userInteraction\": \"NONE\", \"attackComplexity\": \"LOW\", \"availabilityImpact\": \"HIGH\", \"privilegesRequired\": \"NONE\", \"confidentialityImpact\": \"LOW\"}, \"scenarios\": [{\"lang\": \"en\", \"value\": \"GENERAL\"}]}], \"affected\": [{\"vendor\": \"Dr\\u00e4ger\", \"product\": \"Infinity Acute Care System\", \"versions\": [{\"status\": \"affected\", \"version\": \"0\", \"versionType\": \"custom\", \"lessThanOrEqual\": \"VG4.1.1\"}, {\"status\": \"affected\", \"version\": \"0\", \"versionType\": \"custom\", \"lessThanOrEqual\": \"VG4.0.3\"}, {\"status\": \"unaffected\", \"version\": \"VG4.2\", \"versionType\": \"custom\"}], \"defaultStatus\": \"unknown\"}, {\"vendor\": \"Dr\\u00e4ger\", \"product\": \"Standalone Infinity M540 patient monitor\", \"versions\": [{\"status\": \"affected\", \"version\": \"0\", \"versionType\": \"custom\", \"lessThanOrEqual\": \"VG4.1.1\"}, {\"status\": \"unaffected\", \"version\": \"VG4.2\", \"versionType\": \"custom\"}], \"defaultStatus\": \"unknown\"}], \"datePublic\": \"2019-09-12T00:00:00.000Z\", \"references\": [{\"url\": \"https://static.draeger.com/security/download/PSA-19-255-02_Product-Security-Advisory-IACS-VG4.1.pdf\", \"tags\": [\"vendor-advisory\"]}, {\"url\": \"https://www.vulncheck.com/advisories/dr-ger-infinity-m540-vg4-spoofing-and-dos-via-network-message-handling\", \"tags\": [\"third-party-advisory\"]}], \"x_generator\": {\"engine\": \"Vulnogram 1.0.2\"}, \"descriptions\": [{\"lang\": \"en\", \"value\": \"Dr\\u00e4ger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.\", \"supportingMedia\": [{\"type\": \"text/html\", \"value\": \"\u003cp\u003eDr\\u00e4ger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.\u003c/p\u003e\", \"base64\": false}]}], \"problemTypes\": [{\"descriptions\": [{\"lang\": \"en\", \"type\": \"CWE\", \"cweId\": \"CWE-924\", \"description\": \"CWE-924 Improper Enforcement of Message Integrity During Transmission in a Communication Channel\"}]}], \"providerMetadata\": {\"orgId\": \"83251b91-4cc7-4094-a5c7-464a1b83ea10\", \"shortName\": \"VulnCheck\", \"dateUpdated\": \"2026-06-03T15:43:03.688Z\"}}}",
"cveMetadata": "{\"cveId\": \"CVE-2019-25719\", \"state\": \"PUBLISHED\", \"dateUpdated\": \"2026-06-03T15:43:03.688Z\", \"dateReserved\": \"2026-06-02T13:54:01.021Z\", \"assignerOrgId\": \"83251b91-4cc7-4094-a5c7-464a1b83ea10\", \"datePublished\": \"2026-06-02T14:00:52.578Z\", \"assignerShortName\": \"VulnCheck\"}",
"dataType": "CVE_RECORD",
"dataVersion": "5.2"
}
}
}
Loading…
Loading…
Experimental. This forecast is provided for visualization only and may change without notice. Do not use it for operational decisions.
Forecast uses a logistic model when the trend is rising, or an exponential decay model when the trend is falling. Fitted via linearized least squares.
Sightings
| Author | Source | Type | Date | Other |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or observed by the user.
- Confirmed: The vulnerability has been validated from an analyst's perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
- Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
- Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
- Not confirmed: The user expressed doubt about the validity of the vulnerability.
- Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.
Loading…
The MITRE ATT&CK techniques below are AI-generated suggestions, inferred from the description of the
vulnerability by the CIRCL/vulnerability-attack-technique-classification-roberta-base
model, served locally by ML-Gateway.
They have not been verified by an analyst and are provided for guidance only.
Loading…
Loading…