ZOLL Defibrillator Dashboard
Notes
CISA Disclaimer
This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov
Legal Notice
All information products included in https://us-cert.cisa.gov/ics are provided "as is" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.
Risk evaluation
Successful exploitation of these vulnerabilities could allow remote code execution, allow an attacker to gain access to credentials, or impact confidentiality, integrity, and availability of the application.
Critical infrastructure sectors
Healthcare and Public Health
Countries/areas deployed
Worldwide
Company headquarters location
United States
Recommended Practices
CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:
Recommended Practices
CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.
CISA also provides a section for control systems security recommended practices on the ICS webpage onus-cert.cisa.gov. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.
Recommended Practices
Additional mitigation guidance and recommended practices are publicly available on the ICS webpage on us-cert.cisa.gov in the Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.
Organizations observing any suspected malicious activity should follow their established internal procedures and report their findings to CISA for tracking and correlation against other incidents.
{
"document": {
"acknowledgments": [
{
"names": [
"an anonymous party"
],
"summary": "reporting these vulnerabilities to CISA"
}
],
"category": "csaf_security_advisory",
"csaf_version": "2.0",
"distribution": {
"text": "Disclosure is not limited",
"tlp": {
"label": "WHITE",
"url": "https://us-cert.cisa.gov/tlp/"
}
},
"lang": "en-US",
"notes": [
{
"category": "general",
"text": "This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov",
"title": "CISA Disclaimer"
},
{
"category": "legal_disclaimer",
"text": "All information products included in https://us-cert.cisa.gov/ics are provided \"as is\" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.",
"title": "Legal Notice"
},
{
"category": "summary",
"text": "Successful exploitation of these vulnerabilities could allow remote code execution, allow an attacker to gain access to credentials, or impact confidentiality, integrity, and availability of the application.",
"title": "Risk evaluation"
},
{
"category": "other",
"text": "Healthcare and Public Health",
"title": "Critical infrastructure sectors"
},
{
"category": "other",
"text": "Worldwide",
"title": "Countries/areas deployed"
},
{
"category": "other",
"text": "United States",
"title": "Company headquarters location"
},
{
"category": "general",
"text": "CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:",
"title": "Recommended Practices"
},
{
"category": "general",
"text": "CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.\nCISA also provides a section for control systems security recommended practices on the ICS webpage onus-cert.cisa.gov. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.",
"title": "Recommended Practices"
},
{
"category": "general",
"text": "Additional mitigation guidance and recommended practices are publicly available on the ICS webpage on us-cert.cisa.gov in the Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.\nOrganizations observing any suspected malicious activity should follow their established internal procedures and report their findings to CISA for tracking and correlation against other incidents.",
"title": "Recommended Practices"
}
],
"publisher": {
"category": "coordinator",
"contact_details": "Email: CISAservicedesk@cisa.dhs.gov;\n Toll Free: 1-888-282-0870",
"name": "CISA",
"namespace": "https://www.cisa.gov/"
},
"references": [
{
"category": "self",
"summary": "ICS Advisory ICSMA-21-161-01 JSON",
"url": "https://raw.githubusercontent.com/cisagov/CSAF/develop/csaf_files/OT/white/2021/icsma-21-161-01.json"
},
{
"category": "self",
"summary": "ICS Advisory ICSMA-21-161-01 Web Version",
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-21-161-01"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://us-cert.cisa.gov/ics/alerts/ICS-ALERT-10-301-01"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://us-cert.cisa.gov/sites/default/files/recommended_practices/NCCIC_ICS-CERT_Defense_in_Depth_2016_S508C.pdf"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://us-cert.cisa.gov/ics/tips/ICS-TIP-12-146-01B"
}
],
"title": "ZOLL Defibrillator Dashboard",
"tracking": {
"current_release_date": "2021-06-10T00:00:00.000000Z",
"generator": {
"engine": {
"name": "CISA CSAF Generator",
"version": "1.0.0"
}
},
"id": "ICSMA-21-161-01",
"initial_release_date": "2021-06-10T00:00:00.000000Z",
"revision_history": [
{
"date": "2021-06-10T00:00:00.000000Z",
"legacy_version": "Initial",
"number": "1",
"summary": "ICSMA-21-161-01 ZOLL Defibrillator Dashboard"
}
],
"status": "final",
"version": "1"
}
},
"product_tree": {
"branches": [
{
"branches": [
{
"branches": [
{
"category": "product_version_range",
"name": "\u003c 2.2",
"product": {
"name": "Defibrillator Dashboard: All versions prior to 2.2",
"product_id": "CSAFPID-0001"
}
}
],
"category": "product_name",
"name": "Defibrillator Dashboard"
}
],
"category": "vendor",
"name": "ZOLL"
}
]
},
"vulnerabilities": [
{
"cve": "CVE-2021-27489",
"cwe": {
"id": "CWE-434",
"name": "Unrestricted Upload of File with Dangerous Type"
},
"notes": [
{
"category": "summary",
"text": "The web application allows a non-administrative user to upload a malicious file. This file could allow an attacker to remotely execute arbitrary commands.CVE-2021-27489 has been assigned to this vulnerability. A CVSS v3 base score of 9.9 has been assigned; the CVSS vector string is (AV:N/AC:L/PR:L/UI:N/S:C/C:H/I:H/A:H).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2021-27489"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:L/UI:N/S:C/C:H/I:H/A:H"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "ZOLL recommends users of affected versions mitigate these vulnerabilities by upgrading to the latest version of Defibrillator Dashboard Version 2.2 or later. For upgrade assistance or questions, please contact ZOLL Data Products Technical Support by phone at 800-348-9011 option one, or via email at supportdata@zoll.com",
"product_ids": [
"CSAFPID-0001"
]
},
{
"category": "vendor_fix",
"details": "ZOLL recommends users of affected versions keep in mind data on the defibrillator device should be considered the source of accurate data, should there be any discrepancy with the Defibrillator Dashboard.\nUsers should perform frequent local checks to confirm readiness of the devices as per manuals.",
"product_ids": [
"CSAFPID-0001"
]
},
{
"category": "vendor_fix",
"details": "ZOLL recommends users disable the password autocomplete function on browsers accessing the Defibrillator Dashboard.",
"product_ids": [
"CSAFPID-0001"
]
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 9.9,
"baseSeverity": "CRITICAL",
"vectorString": "CVSS:3.0/AV:N/AC:L/PR:L/UI:N/S:C/C:H/I:H/A:H",
"version": "3.0"
},
"products": [
"CSAFPID-0001"
]
}
]
},
{
"cve": "CVE-2021-27481",
"cwe": {
"id": "CWE-321",
"name": "Use of Hard-coded Cryptographic Key"
},
"notes": [
{
"category": "summary",
"text": "The affected products utilize an encryption key in the data exchange process, which is hardcoded. This could allow an attacker to gain access to sensitive information.CVE-2021-27481 has been assigned to this vulnerability. A CVSS v3 base score of 7.1 has been assigned; the CVSS vector string is (AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:N).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2021-27481"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:N"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "ZOLL recommends users of affected versions mitigate these vulnerabilities by upgrading to the latest version of Defibrillator Dashboard Version 2.2 or later. For upgrade assistance or questions, please contact ZOLL Data Products Technical Support by phone at 800-348-9011 option one, or via email at supportdata@zoll.com",
"product_ids": [
"CSAFPID-0001"
]
},
{
"category": "vendor_fix",
"details": "ZOLL recommends users of affected versions keep in mind data on the defibrillator device should be considered the source of accurate data, should there be any discrepancy with the Defibrillator Dashboard.\nUsers should perform frequent local checks to confirm readiness of the devices as per manuals.",
"product_ids": [
"CSAFPID-0001"
]
},
{
"category": "vendor_fix",
"details": "ZOLL recommends users disable the password autocomplete function on browsers accessing the Defibrillator Dashboard.",
"product_ids": [
"CSAFPID-0001"
]
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 7.1,
"baseSeverity": "HIGH",
"vectorString": "CVSS:3.0/AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:N",
"version": "3.0"
},
"products": [
"CSAFPID-0001"
]
}
]
},
{
"cve": "CVE-2021-27487",
"cwe": {
"id": "CWE-312",
"name": "Cleartext Storage of Sensitive Information"
},
"notes": [
{
"category": "summary",
"text": "The affected products contain credentials stored in plaintext. This could allow an attacker to gain access to sensitive information.CVE-2021-27487 has been assigned to this vulnerability. A CVSS v3 base score of 7.1 has been assigned; the CVSS vector string is (AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:N).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2021-27487"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:N"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "ZOLL recommends users of affected versions mitigate these vulnerabilities by upgrading to the latest version of Defibrillator Dashboard Version 2.2 or later. For upgrade assistance or questions, please contact ZOLL Data Products Technical Support by phone at 800-348-9011 option one, or via email at supportdata@zoll.com",
"product_ids": [
"CSAFPID-0001"
]
},
{
"category": "vendor_fix",
"details": "ZOLL recommends users of affected versions keep in mind data on the defibrillator device should be considered the source of accurate data, should there be any discrepancy with the Defibrillator Dashboard.\nUsers should perform frequent local checks to confirm readiness of the devices as per manuals.",
"product_ids": [
"CSAFPID-0001"
]
},
{
"category": "vendor_fix",
"details": "ZOLL recommends users disable the password autocomplete function on browsers accessing the Defibrillator Dashboard.",
"product_ids": [
"CSAFPID-0001"
]
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 7.1,
"baseSeverity": "HIGH",
"vectorString": "CVSS:3.0/AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:N",
"version": "3.0"
},
"products": [
"CSAFPID-0001"
]
}
]
},
{
"cve": "CVE-2021-27479",
"cwe": {
"id": "CWE-79",
"name": "Improper Neutralization of Input During Web Page Generation (\u0027Cross-site Scripting\u0027)"
},
"notes": [
{
"category": "summary",
"text": "The affected product \u0027s web application could allow a low privilege user to inject parameters to contain malicious scripts to be executed by higher privilege users.CVE-2021-27479 has been assigned to this vulnerability. A CVSS v3 base score of 4.6 has been assigned; the CVSS vector string is (AV:N/AC:L/PR:L/UI:R/S:U/C:L/I:L/A:N).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2021-27479"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:L/UI:R/S:U/C:L/I:L/A:N"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "ZOLL recommends users of affected versions mitigate these vulnerabilities by upgrading to the latest version of Defibrillator Dashboard Version 2.2 or later. For upgrade assistance or questions, please contact ZOLL Data Products Technical Support by phone at 800-348-9011 option one, or via email at supportdata@zoll.com",
"product_ids": [
"CSAFPID-0001"
]
},
{
"category": "vendor_fix",
"details": "ZOLL recommends users of affected versions keep in mind data on the defibrillator device should be considered the source of accurate data, should there be any discrepancy with the Defibrillator Dashboard.\nUsers should perform frequent local checks to confirm readiness of the devices as per manuals.",
"product_ids": [
"CSAFPID-0001"
]
},
{
"category": "vendor_fix",
"details": "ZOLL recommends users disable the password autocomplete function on browsers accessing the Defibrillator Dashboard.",
"product_ids": [
"CSAFPID-0001"
]
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 4.6,
"baseSeverity": "MEDIUM",
"vectorString": "CVSS:3.0/AV:N/AC:L/PR:L/UI:R/S:U/C:L/I:L/A:N",
"version": "3.0"
},
"products": [
"CSAFPID-0001"
]
}
]
},
{
"cve": "CVE-2021-27485",
"cwe": {
"id": "CWE-257",
"name": "Storing Passwords in a Recoverable Format"
},
"notes": [
{
"category": "summary",
"text": "The application allows users to store their passwords in a recoverable format, which could allow an attacker to retrieve the credentials from the web browser.CVE-2021-27485 has been assigned to this vulnerability. A CVSS v3 base score of 7.1 has been assigned; the CVSS vector string is (AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:N).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2021-27485"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:N"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "ZOLL recommends users of affected versions mitigate these vulnerabilities by upgrading to the latest version of Defibrillator Dashboard Version 2.2 or later. For upgrade assistance or questions, please contact ZOLL Data Products Technical Support by phone at 800-348-9011 option one, or via email at supportdata@zoll.com",
"product_ids": [
"CSAFPID-0001"
]
},
{
"category": "vendor_fix",
"details": "ZOLL recommends users of affected versions keep in mind data on the defibrillator device should be considered the source of accurate data, should there be any discrepancy with the Defibrillator Dashboard.\nUsers should perform frequent local checks to confirm readiness of the devices as per manuals.",
"product_ids": [
"CSAFPID-0001"
]
},
{
"category": "vendor_fix",
"details": "ZOLL recommends users disable the password autocomplete function on browsers accessing the Defibrillator Dashboard.",
"product_ids": [
"CSAFPID-0001"
]
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 7.1,
"baseSeverity": "HIGH",
"vectorString": "CVSS:3.0/AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:N",
"version": "3.0"
},
"products": [
"CSAFPID-0001"
]
}
]
},
{
"cve": "CVE-2021-27483",
"cwe": {
"id": "CWE-269",
"name": "Improper Privilege Management"
},
"notes": [
{
"category": "summary",
"text": "The affected products contain insecure filesystem permissions that could allow a lower privilege user to escalate privileges to an administrative level user.CVE-2021-27483 has been assigned to this vulnerability. A CVSS v3 base score of 5.3 has been assigned; the CVSS vector string is (AV:L/AC:L/PR:L/UI:N/S:U/C:L/I:L/A:L).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2021-27483"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:L/AC:L/PR:L/UI:N/S:U/C:L/I:L/A:L"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "ZOLL recommends users of affected versions mitigate these vulnerabilities by upgrading to the latest version of Defibrillator Dashboard Version 2.2 or later. For upgrade assistance or questions, please contact ZOLL Data Products Technical Support by phone at 800-348-9011 option one, or via email at supportdata@zoll.com",
"product_ids": [
"CSAFPID-0001"
]
},
{
"category": "vendor_fix",
"details": "ZOLL recommends users of affected versions keep in mind data on the defibrillator device should be considered the source of accurate data, should there be any discrepancy with the Defibrillator Dashboard.\nUsers should perform frequent local checks to confirm readiness of the devices as per manuals.",
"product_ids": [
"CSAFPID-0001"
]
},
{
"category": "vendor_fix",
"details": "ZOLL recommends users disable the password autocomplete function on browsers accessing the Defibrillator Dashboard.",
"product_ids": [
"CSAFPID-0001"
]
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 5.3,
"baseSeverity": "MEDIUM",
"vectorString": "CVSS:3.0/AV:L/AC:L/PR:L/UI:N/S:U/C:L/I:L/A:L",
"version": "3.0"
},
"products": [
"CSAFPID-0001"
]
}
]
}
]
}
Loading…
Loading…
Sightings
| Author | Source | Type | Date |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or observed by the user.
- Confirmed: The vulnerability has been validated from an analyst's perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
- Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
- Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
- Not confirmed: The user expressed doubt about the validity of the vulnerability.
- Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.
Loading…
Loading…