ICSMA-21-152-01

Vulnerability from csaf_cisa - Published: 2021-06-01 00:00 - Updated: 2021-12-14 00:00
Summary
Hillrom Medical Device Management (Update B)

Notes

CISA Disclaimer
This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov
Legal Notice
All information products included in https://us-cert.cisa.gov/ics are provided "as is" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.
Risk evaluation
Successful exploitation of these vulnerabilities could allow an attacker to cause memory corruption and remotely execute arbitrary code.
Critical infrastructure sectors
Healthcare and Public Health
Countries/areas deployed
Worldwide
Company headquarters location
United States
Recommended Practices
CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should: Minimize network exposure for all control system devices and/or systems, and ensure they are not accessible from the Internet; Locate control system networks and remote devices behind firewalls and isolate them from the business network; When remote access is required, use secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as its connected devices.
Recommended Practices
CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures. CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.
Recommended Practices
Additional mitigation guidance and recommended practices are publicly available on the ICS webpage on cisa.gov in the Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.
Exploitability
No known public exploits specifically target these vulnerabilities. These vulnerabilities have high attack complexity.

{
  "document": {
    "acknowledgments": [
      {
        "names": [
          "Itamar Cohen-Matalon"
        ],
        "organization": "Medigate Research Labs",
        "summary": "reporting these vulnerabilities to Hillrom"
      }
    ],
    "category": "csaf_security_advisory",
    "csaf_version": "2.0",
    "distribution": {
      "text": "Disclosure is not limited",
      "tlp": {
        "label": "WHITE",
        "url": "https://us-cert.cisa.gov/tlp/"
      }
    },
    "lang": "en-US",
    "notes": [
      {
        "category": "general",
        "text": "This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov",
        "title": "CISA Disclaimer"
      },
      {
        "category": "legal_disclaimer",
        "text": "All information products included in https://us-cert.cisa.gov/ics are provided \"as is\" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.",
        "title": "Legal Notice"
      },
      {
        "category": "summary",
        "text": "Successful exploitation of these vulnerabilities could allow an attacker to cause memory corruption and remotely execute arbitrary code.",
        "title": "Risk evaluation"
      },
      {
        "category": "other",
        "text": "Healthcare and Public Health",
        "title": "Critical infrastructure sectors"
      },
      {
        "category": "other",
        "text": "Worldwide",
        "title": "Countries/areas deployed"
      },
      {
        "category": "other",
        "text": "United States",
        "title": "Company headquarters location"
      },
      {
        "category": "general",
        "text": "CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should: Minimize network exposure for all control system devices and/or systems, and ensure they are not accessible from the Internet; Locate control system networks and remote devices behind firewalls and isolate them from the business network; When remote access is required, use secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as its connected devices.",
        "title": "Recommended Practices"
      },
      {
        "category": "general",
        "text": "CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.\nCISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.",
        "title": "Recommended Practices"
      },
      {
        "category": "general",
        "text": "Additional mitigation guidance and recommended practices are publicly available on the ICS webpage on cisa.gov in the Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.",
        "title": "Recommended Practices"
      },
      {
        "category": "other",
        "text": "No known public exploits specifically target these vulnerabilities. These vulnerabilities have high attack complexity.",
        "title": "Exploitability"
      }
    ],
    "publisher": {
      "category": "coordinator",
      "contact_details": "Email: CISAservicedesk@cisa.dhs.gov;\n Toll Free: 1-888-282-0870",
      "name": "CISA",
      "namespace": "https://www.cisa.gov/"
    },
    "references": [
      {
        "category": "self",
        "summary": "ICS Advisory ICSMA-21-152-01 JSON",
        "url": "https://raw.githubusercontent.com/cisagov/CSAF/develop/csaf_files/OT/white/2021/icsma-21-152-01.json"
      },
      {
        "category": "self",
        "summary": "ICS Advisory ICSMA-21-152-01 Web Version",
        "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-21-152-01"
      },
      {
        "category": "external",
        "summary": "Recommended Practices",
        "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-21-152-01"
      },
      {
        "category": "external",
        "summary": "Recommended Practices",
        "url": "https://www.cisa.gov/uscert/sites/default/files/recommended_practices/NCCIC_ICS-CERT_Defense_in_Depth_2016_S508C.pdf"
      }
    ],
    "title": "Hillrom Medical Device Management (Update B)",
    "tracking": {
      "current_release_date": "2021-12-14T00:00:00.000000Z",
      "generator": {
        "engine": {
          "name": "CISA CSAF Generator",
          "version": "1.0.0"
        }
      },
      "id": "ICSMA-21-152-01",
      "initial_release_date": "2021-06-01T00:00:00.000000Z",
      "revision_history": [
        {
          "date": "2021-06-01T00:00:00.000000Z",
          "legacy_version": "Initial",
          "number": "1",
          "summary": "ICSMA-21-152-01 Hillrom Medical Device Management"
        },
        {
          "date": "2021-12-14T00:00:00.000000Z",
          "legacy_version": "A",
          "number": "2",
          "summary": "ICSMA-21-152-01 Hillrom Medical Device Management (Update A)"
        }
      ],
      "status": "final",
      "version": "2"
    }
  },
  "product_tree": {
    "branches": [
      {
        "branches": [
          {
            "branches": [
              {
                "category": "product_version_range",
                "name": "\u003c 1.52",
                "product": {
                  "name": "Welch Allyn Connex Spot Monitor (CSM): versions prior to v1.52",
                  "product_id": "CSAFPID-0001"
                }
              }
            ],
            "category": "product_name",
            "name": "Welch Allyn Connex Spot Monitor (CSM)"
          },
          {
            "branches": [
              {
                "category": "product_version_range",
                "name": "\u003c 2.43.02",
                "product": {
                  "name": "Welch Allyn Connex Integrated Wall System (CIWS): versions prior to v2.43.02",
                  "product_id": "CSAFPID-0002"
                }
              }
            ],
            "category": "product_name",
            "name": "Welch Allyn Connex Integrated Wall System (CIWS)"
          },
          {
            "branches": [
              {
                "category": "product_version_range",
                "name": "\u003c 2.43.02",
                "product": {
                  "name": "Welch Allyn Connex Vital Signs Monitor (CVSM): versions prior to v2.43.02",
                  "product_id": "CSAFPID-0003"
                }
              }
            ],
            "category": "product_name",
            "name": "Welch Allyn Connex Vital Signs Monitor (CVSM)"
          },
          {
            "branches": [
              {
                "category": "product_version_range",
                "name": "\u003c 1.8.4 Service Pack 01",
                "product": {
                  "name": "Welch Allyn Connex Central Station (CS): versions prior to v1.8.4 Service Pack 01",
                  "product_id": "CSAFPID-0004"
                }
              }
            ],
            "category": "product_name",
            "name": "Welch Allyn Connex Central Station (CS)"
          },
          {
            "branches": [
              {
                "category": "product_version_range",
                "name": "\u003c 1.11.00",
                "product": {
                  "name": "Welch Allyn Spot Vital Signs 4400 Device (Spot 4400) / Welch Allyn Spot 4400 Vital Signs Extended Care Device: versions prior to v1.11.00",
                  "product_id": "CSAFPID-0005"
                }
              }
            ],
            "category": "product_name",
            "name": "Welch Allyn Spot Vital Signs 4400 Device (Spot 4400) / Welch Allyn Spot 4400 Vital Signs Extended Care Device"
          },
          {
            "branches": [
              {
                "category": "product_version_range",
                "name": "\u003c 3.2",
                "product": {
                  "name": "Welch Allyn Software Development Kit (SDK): versions prior to v3.2",
                  "product_id": "CSAFPID-0006"
                }
              }
            ],
            "category": "product_name",
            "name": "Welch Allyn Software Development Kit (SDK)"
          },
          {
            "branches": [
              {
                "category": "product_version_range",
                "name": "\u003c 1.10",
                "product": {
                  "name": "Welch Allyn Service Tool: versions prior to v1.10",
                  "product_id": "CSAFPID-0007"
                }
              }
            ],
            "category": "product_name",
            "name": "Welch Allyn Service Tool"
          },
          {
            "branches": [
              {
                "category": "product_version_range",
                "name": "\u003c 5.3",
                "product": {
                  "name": "Welch Allyn Connex Device Integration Suite - Network Connectivity Engine (NCE): versions prior to v5.3",
                  "product_id": "CSAFPID-0008"
                }
              }
            ],
            "category": "product_name",
            "name": "Welch Allyn Connex Device Integration Suite - Network Connectivity Engine (NCE)"
          },
          {
            "branches": [
              {
                "category": "product_version_range",
                "name": "\u003c 1.7.0.0",
                "product": {
                  "name": "Welch Allyn Service Monitor: versions prior to v1.7.0.0",
                  "product_id": "CSAFPID-0009"
                }
              }
            ],
            "category": "product_name",
            "name": "Welch Allyn Service Monitor"
          }
        ],
        "category": "vendor",
        "name": "Hillrom"
      }
    ]
  },
  "vulnerabilities": [
    {
      "cve": "CVE-2021-27410",
      "cwe": {
        "id": "CWE-787",
        "name": "Out-of-bounds Write"
      },
      "notes": [
        {
          "category": "summary",
          "text": "The affected product is vulnerable to an out-of-bounds write, which may result in corruption of data or code execution. CVE-2021-27410 has been assigned to this vulnerability. A CVSS v3 base score of 5.9 has been calculated; the CVSS vector string is (AV:A/AC:H/PR:L/UI:N/S:U/C:L/I:H/A:L).",
          "title": "Vulnerability Summary"
        }
      ],
      "product_status": {
        "known_affected": [
          "CSAFPID-0001",
          "CSAFPID-0002",
          "CSAFPID-0003",
          "CSAFPID-0004",
          "CSAFPID-0005",
          "CSAFPID-0006",
          "CSAFPID-0007",
          "CSAFPID-0008",
          "CSAFPID-0009"
        ]
      },
      "references": [
        {
          "category": "external",
          "summary": "web.nvd.nist.gov",
          "url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2021-27410"
        },
        {
          "category": "external",
          "summary": "www.first.org",
          "url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:A/AC:H/PR:L/UI:N/S:U/C:L/I:H/A:L"
        }
      ],
      "remediations": [
        {
          "category": "vendor_fix",
          "details": "Welch Allyn Service Tool: v1.10",
          "product_ids": [
            "CSAFPID-0007"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "Welch Allyn Software Development Kit (SDK): v3.2",
          "product_ids": [
            "CSAFPID-0006"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "Welch Allyn Connex Central Station (CS): v1.8.4 Service Pack 01 (released November 2022)",
          "product_ids": [
            "CSAFPID-0004"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "Welch Allyn Connex Device Integration Suite - Network Connectivity Engine (NCE): v5.3 (released September 2021)",
          "product_ids": [
            "CSAFPID-0008"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "Welch Allyn Spot Vital Signs 4400 Device (Spot 4400) / Welch Allyn Spot 4400 Vital Signs Extended Care Device: v1.11.00 (Released October 2021)",
          "product_ids": [
            "CSAFPID-0005"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "Welch Allyn Service Monitor: v1.7.0.0",
          "product_ids": [
            "CSAFPID-0009"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "Welch Allyn Connex Vital Signs Monitor (CVSM): v2.43.02",
          "product_ids": [
            "CSAFPID-0003"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "Welch Allen Connex Integrated Wall System (CIWS): v2.43.02",
          "product_ids": [
            "CSAFPID-0002"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "Welch Allyn Connex Spot Monitor (CSM): v1.52",
          "product_ids": [
            "CSAFPID-0001"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "Welch Allyn Spot Vital Signs 4400 Device (Spot 4400) / Welch Allyn Spot 4400 Vital Signs Extended Care Device: v1.11.00 (available Fall 2021)",
          "product_ids": [
            "CSAFPID-0005"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "Hillrom recommends users to upgrade to the latest versions of their products. Information on how to update these products to their new versions can be found on the Hillrom disclosure page.",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009"
          ],
          "url": "https://www.hillrom.com/en/responsible-disclosures/"
        },
        {
          "category": "mitigation",
          "details": "Apply proper network and physical security controls.",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009"
          ]
        },
        {
          "category": "mitigation",
          "details": "Apply authentication for server access.",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009"
          ]
        },
        {
          "category": "mitigation",
          "details": "Apply data execution prevention (DEP) where applicable to help prevent shellcode from running. Address space layout randomization (ASLR) is built into standard Windows and Linux distributions.",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009"
          ]
        }
      ],
      "scores": [
        {
          "cvss_v3": {
            "baseScore": 5.9,
            "baseSeverity": "MEDIUM",
            "vectorString": "CVSS:3.0/AV:A/AC:H/PR:L/UI:N/S:U/C:L/I:H/A:L",
            "version": "3.0"
          },
          "products": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009"
          ]
        }
      ]
    },
    {
      "cve": "CVE-2021-27408",
      "cwe": {
        "id": "CWE-125",
        "name": "Out-of-bounds Read"
      },
      "notes": [
        {
          "category": "summary",
          "text": "The affected product is vulnerable to an out-of-bounds read, which can cause information leakage leading to arbitrary code execution if chained to the out-of-bounds write vulnerability. CVE-2021-27408 has been assigned to this vulnerability. A CVSS v3 base score of 5.9 has been calculated; the CVSS vector string is (AV:A/AC:H/PR:L/UI:N/S:U/C:L/I:H/A:L).",
          "title": "Vulnerability Summary"
        }
      ],
      "product_status": {
        "known_affected": [
          "CSAFPID-0001",
          "CSAFPID-0002",
          "CSAFPID-0003",
          "CSAFPID-0004",
          "CSAFPID-0005",
          "CSAFPID-0006",
          "CSAFPID-0007",
          "CSAFPID-0008",
          "CSAFPID-0009"
        ]
      },
      "references": [
        {
          "category": "external",
          "summary": "web.nvd.nist.gov",
          "url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2021-27408"
        },
        {
          "category": "external",
          "summary": "www.first.org",
          "url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:A/AC:H/PR:L/UI:N/S:U/C:L/I:H/A:L"
        }
      ],
      "remediations": [
        {
          "category": "vendor_fix",
          "details": "Welch Allyn Service Tool: v1.10",
          "product_ids": [
            "CSAFPID-0007"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "Welch Allyn Software Development Kit (SDK): v3.2",
          "product_ids": [
            "CSAFPID-0006"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "Welch Allyn Connex Central Station (CS): v1.8.4 Service Pack 01 (released November 2022)",
          "product_ids": [
            "CSAFPID-0004"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "Welch Allyn Connex Device Integration Suite - Network Connectivity Engine (NCE): v5.3 (released September 2021)",
          "product_ids": [
            "CSAFPID-0008"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "Welch Allyn Spot Vital Signs 4400 Device (Spot 4400) / Welch Allyn Spot 4400 Vital Signs Extended Care Device: v1.11.00 (Released October 2021)",
          "product_ids": [
            "CSAFPID-0005"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "Welch Allyn Service Monitor: v1.7.0.0",
          "product_ids": [
            "CSAFPID-0009"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "Welch Allyn Connex Vital Signs Monitor (CVSM): v2.43.02",
          "product_ids": [
            "CSAFPID-0003"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "Welch Allen Connex Integrated Wall System (CIWS): v2.43.02",
          "product_ids": [
            "CSAFPID-0002"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "Welch Allyn Connex Spot Monitor (CSM): v1.52",
          "product_ids": [
            "CSAFPID-0001"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "Welch Allyn Spot Vital Signs 4400 Device (Spot 4400) / Welch Allyn Spot 4400 Vital Signs Extended Care Device: v1.11.00 (available Fall 2021)",
          "product_ids": [
            "CSAFPID-0005"
          ]
        },
        {
          "category": "vendor_fix",
          "details": "Hillrom recommends users to upgrade to the latest versions of their products. Information on how to update these products to their new versions can be found on the Hillrom disclosure page.",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009"
          ],
          "url": "https://www.hillrom.com/en/responsible-disclosures/"
        },
        {
          "category": "mitigation",
          "details": "Apply proper network and physical security controls.",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009"
          ]
        },
        {
          "category": "mitigation",
          "details": "Apply authentication for server access.",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009"
          ]
        },
        {
          "category": "mitigation",
          "details": "Apply data execution prevention (DEP) where applicable to help prevent shellcode from running. Address space layout randomization (ASLR) is built into standard Windows and Linux distributions.",
          "product_ids": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009"
          ]
        }
      ],
      "scores": [
        {
          "cvss_v3": {
            "baseScore": 5.9,
            "baseSeverity": "MEDIUM",
            "vectorString": "CVSS:3.0/AV:A/AC:H/PR:L/UI:N/S:U/C:L/I:H/A:L",
            "version": "3.0"
          },
          "products": [
            "CSAFPID-0001",
            "CSAFPID-0002",
            "CSAFPID-0003",
            "CSAFPID-0004",
            "CSAFPID-0005",
            "CSAFPID-0006",
            "CSAFPID-0007",
            "CSAFPID-0008",
            "CSAFPID-0009"
          ]
        }
      ]
    }
  ]
}


Log in or create an account to share your comment.




Tags
Taxonomy of the tags.


Loading…

Loading…

Loading…

Sightings

Author Source Type Date

Nomenclature

  • Seen: The vulnerability was mentioned, discussed, or observed by the user.
  • Confirmed: The vulnerability has been validated from an analyst's perspective.
  • Published Proof of Concept: A public proof of concept is available for this vulnerability.
  • Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
  • Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
  • Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.