ICSMA-19-255-01
Vulnerability from csaf_cisa - Published: 2019-09-12 00:00 - Updated: 2019-09-12 00:00Summary
Philips IntelliVue WLAN
Notes
CISA Disclaimer
This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov
Legal Notice
All information products included in https://us-cert.cisa.gov/ics are provided "as is" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.
Risk evaluation
Successful exploitation of these vulnerabilities may cause corruption of the IntelliVue WLAN firmware and impact to the data flow over the WLAN Version A and WLAN Version B wireless modules. This would lead to an inoperative condition alert at the device and Central Station.
Critical infrastructure sectors
Healthcare and Public Health
Countries/areas deployed
Worldwide
Company headquarters location
Netherlands
Recommended Practices
CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:
Recommended Practices
CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.
CISA also provides a section for control systems security recommended practices on the ICS webpage onus-cert.gov. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.
Recommended Practices
Additional mitigation guidance and recommended practices are publicly available on the ICS webpage on us-cert.gov in the Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.
Organizations observing any suspected malicious activity should follow their established internal procedures and report their findings to CISA for tracking and correlation against other incidents.
Exploitability
No known public exploits specifically target these vulnerabilities. These vulnerabilities are not exploitable remotely. High skill level is needed to exploit.
{
"document": {
"acknowledgments": [
{
"names": [
"Shawn Loveric"
],
"organization": "Finite State, Inc",
"summary": "reporting these vulnerabilities to Philips"
}
],
"category": "csaf_security_advisory",
"csaf_version": "2.0",
"distribution": {
"text": "Disclosure is not limited",
"tlp": {
"label": "WHITE",
"url": "https://us-cert.cisa.gov/tlp/"
}
},
"lang": "en-US",
"notes": [
{
"category": "general",
"text": "This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov",
"title": "CISA Disclaimer"
},
{
"category": "legal_disclaimer",
"text": "All information products included in https://us-cert.cisa.gov/ics are provided \"as is\" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.",
"title": "Legal Notice"
},
{
"category": "summary",
"text": "Successful exploitation of these vulnerabilities may cause corruption of the IntelliVue WLAN firmware and impact to the data flow over the WLAN Version A and WLAN Version B wireless modules. This would lead to an inoperative condition alert at the device and Central Station.",
"title": "Risk evaluation"
},
{
"category": "other",
"text": "Healthcare and Public Health",
"title": "Critical infrastructure sectors"
},
{
"category": "other",
"text": "Worldwide",
"title": "Countries/areas deployed"
},
{
"category": "other",
"text": "Netherlands",
"title": "Company headquarters location"
},
{
"category": "general",
"text": "CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:",
"title": "Recommended Practices"
},
{
"category": "general",
"text": "CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.\nCISA also provides a section for control systems security recommended practices on the ICS webpage onus-cert.gov. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.",
"title": "Recommended Practices"
},
{
"category": "general",
"text": "Additional mitigation guidance and recommended practices are publicly available on the ICS webpage on us-cert.gov in the Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.\nOrganizations observing any suspected malicious activity should follow their established internal procedures and report their findings to CISA for tracking and correlation against other incidents.",
"title": "Recommended Practices"
},
{
"category": "other",
"text": "No known public exploits specifically target these vulnerabilities. These vulnerabilities are not exploitable remotely. High skill level is needed to exploit.",
"title": "Exploitability"
}
],
"publisher": {
"category": "coordinator",
"contact_details": "Email: CISAservicedesk@cisa.dhs.gov;\n Toll Free: 1-888-282-0870",
"name": "CISA",
"namespace": "https://www.cisa.gov/"
},
"references": [
{
"category": "self",
"summary": "ICS Advisory ICSMA-19-255-01 JSON",
"url": "https://raw.githubusercontent.com/cisagov/CSAF/develop/csaf_files/OT/white/2019/icsma-19-255-01.json"
},
{
"category": "self",
"summary": "ICS Advisory ICSMA-19-255-01 Web Version",
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-19-255-01"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-19-255-01"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://www.us-cert.gov/sites/default/files/recommended_practices/NCCIC_ICS-CERT_Defense_in_Depth_2016_S508C.pdf"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://www.us-cert.gov/ics/tips/ICS-TIP-12-146-01B"
}
],
"title": "Philips IntelliVue WLAN",
"tracking": {
"current_release_date": "2019-09-12T00:00:00.000000Z",
"generator": {
"engine": {
"name": "CISA CSAF Generator",
"version": "1.0.0"
}
},
"id": "ICSMA-19-255-01",
"initial_release_date": "2019-09-12T00:00:00.000000Z",
"revision_history": [
{
"date": "2019-09-12T00:00:00.000000Z",
"legacy_version": "Initial",
"number": "1",
"summary": "ICSMA-19-255-01 Philips IntelliVue WLAN"
}
],
"status": "final",
"version": "1"
}
},
"product_tree": {
"branches": [
{
"branches": [
{
"branches": [
{
"category": "product_version",
"name": "Version B, Firmware A.01.09, Part #: N/A (Replaced by Version C)",
"product": {
"name": "IntelliVue MP monitors MX800/700/600 ((865240/41/42) WLAN: Version B, Firmware A.01.09, Part #: N/A (Replaced by Version C)",
"product_id": "CSAFPID-0001"
}
}
],
"category": "product_name",
"name": "IntelliVue MP monitors MX800/700/600 ((865240/41/42) WLAN"
},
{
"branches": [
{
"category": "product_version",
"name": "Version B, Firmware A.01.09, Part #: N/A (Replaced by Version C)",
"product": {
"name": "IntelliVue MP monitors MP2/X2 (M8102A/M3002A) WLAN: Version B, Firmware A.01.09, Part #: N/A (Replaced by Version C)",
"product_id": "CSAFPID-0002"
}
}
],
"category": "product_name",
"name": "IntelliVue MP monitors MP2/X2 (M8102A/M3002A) WLAN"
},
{
"branches": [
{
"category": "product_version",
"name": "Version A, Firmware A.03.09, Part #: M8096-67501",
"product": {
"name": "IntelliVue MP monitors MP5/5SC (M8105A/5AS) WLAN: Version A, Firmware A.03.09, Part #: M8096-67501",
"product_id": "CSAFPID-0003"
}
}
],
"category": "product_name",
"name": "IntelliVue MP monitors MP5/5SC (M8105A/5AS) WLAN"
},
{
"branches": [
{
"category": "product_version",
"name": "Version A, Firmware A.03.09",
"product": {
"name": "IntelliVue MP monitors MP20-MP90 (M8001A/2A/3A/4A/5A/7A/8A/10A) WLAN: Version A Firmware A.03.09",
"product_id": "CSAFPID-0004"
}
}
],
"category": "product_name",
"name": "IntelliVue MP monitors MP20-MP90 (M8001A/2A/3A/4A/5A/7A/8A/10A) WLAN"
}
],
"category": "vendor",
"name": "Philips"
}
]
},
"vulnerabilities": [
{
"cve": "CVE-2019-13530",
"cwe": {
"id": "CWE-259",
"name": "Use of Hard-coded Password"
},
"notes": [
{
"category": "summary",
"text": "An attacker can use these credentials to login via ftp and upload a malicious firmware. CVE-2019-13530 has been assigned to this vulnerability. A CVSS v3 base score of 6.4 has been calculated; the CVSS vector string is (AV:A/AC:H/PR:H/UI:N/S:U/C:H/I:H/A:H).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2019-13530"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:A/AC:H/PR:H/UI:N/S:U/C:H/I:H/A:H"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "WLAN Version B is obsolete. Philips recommends users update to the WLAN assembly Version C wireless module in affected IntelliVue monitors.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
]
},
{
"category": "vendor_fix",
"details": "WLAN Version C with current firmware of B.00.31 is not vulnerable to the described attack.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
]
},
{
"category": "vendor_fix",
"details": "WLAN Version A is a software patch from Philips estimated to be available in Incenter by the end of 2019.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
]
},
{
"category": "vendor_fix",
"details": "Wireless network access should be controlled by authentication and authorization (e.g., WPA2), which are supported by Philips.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
]
},
{
"category": "vendor_fix",
"details": "As an additional mitigation measure, Philips recommends implementing a firewall rule on the user \u0027s wireless network to restrict access control via FTP. Users should implement role-based access controls to control physical access to the system. Refer to the Security for Clinical Networks Guide and the IntelliVue Network Configuration Guide for additional security recommendations.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
]
},
{
"category": "vendor_fix",
"details": "Please see the Philips product security website for the latest security information for Philips products:https://www.philips.com/productsecurity",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
],
"url": "https://www.philips.com/productsecurity"
},
{
"category": "vendor_fix",
"details": "Where additional information is needed, follow this link to existing cybersecurity in medical device guidance issued by the FDA.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
],
"url": "https://www.fda.gov/MedicalDevices/DigitalHealth/ucm373213.htm"
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 6.4,
"baseSeverity": "MEDIUM",
"vectorString": "CVSS:3.0/AV:A/AC:H/PR:H/UI:N/S:U/C:H/I:H/A:H",
"version": "3.0"
},
"products": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
]
}
]
},
{
"cve": "CVE-2019-13534",
"cwe": {
"id": "CWE-494",
"name": "Download of Code Without Integrity Check"
},
"notes": [
{
"category": "summary",
"text": "The product downloads source code or an executable from a remote location and executes the code without sufficiently verifying the origin and integrity of the code. CVE-2019-13534 has been assigned to this vulnerability. A CVSS v3 base score of 6.4 has been calculated; the CVSS vector string is (AV:A/AC:H/PR:H/UI:N/S:U/C:H/I:H/A:H).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2019-13534"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:A/AC:H/PR:H/UI:N/S:U/C:H/I:H/A:H"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "WLAN Version B is obsolete. Philips recommends users update to the WLAN assembly Version C wireless module in affected IntelliVue monitors.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
]
},
{
"category": "vendor_fix",
"details": "WLAN Version C with current firmware of B.00.31 is not vulnerable to the described attack.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
]
},
{
"category": "vendor_fix",
"details": "WLAN Version A is a software patch from Philips estimated to be available in Incenter by the end of 2019.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
]
},
{
"category": "vendor_fix",
"details": "Wireless network access should be controlled by authentication and authorization (e.g., WPA2), which are supported by Philips.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
]
},
{
"category": "vendor_fix",
"details": "As an additional mitigation measure, Philips recommends implementing a firewall rule on the user \u0027s wireless network to restrict access control via FTP. Users should implement role-based access controls to control physical access to the system. Refer to the Security for Clinical Networks Guide and the IntelliVue Network Configuration Guide for additional security recommendations.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
]
},
{
"category": "vendor_fix",
"details": "Please see the Philips product security website for the latest security information for Philips products:https://www.philips.com/productsecurity",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
],
"url": "https://www.philips.com/productsecurity"
},
{
"category": "vendor_fix",
"details": "Where additional information is needed, follow this link to existing cybersecurity in medical device guidance issued by the FDA.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
],
"url": "https://www.fda.gov/MedicalDevices/DigitalHealth/ucm373213.htm"
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 6.4,
"baseSeverity": "MEDIUM",
"vectorString": "CVSS:3.0/AV:A/AC:H/PR:H/UI:N/S:U/C:H/I:H/A:H",
"version": "3.0"
},
"products": [
"CSAFPID-0001",
"CSAFPID-0002",
"CSAFPID-0003",
"CSAFPID-0004"
]
}
]
}
]
}
Loading…
Loading…
Sightings
| Author | Source | Type | Date |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or observed by the user.
- Confirmed: The vulnerability has been validated from an analyst's perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
- Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
- Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
- Not confirmed: The user expressed doubt about the validity of the vulnerability.
- Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.
Loading…
Loading…