FKIE_CVE-2018-14781

Vulnerability from fkie_nvd - Published: 2018-08-13 21:48 - Updated: 2026-06-17 01:41
Summary
Medtronic MiniMed MMT devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.

{
  "affected": [
    {
      "affectedData": [
        {
          "defaultStatus": "unaffected",
          "product": "MMT- 508 - MiniMed pump",
          "vendor": "Medtronic",
          "versions": [
            {
              "status": "affected",
              "version": "All versions"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "MMT \u2013 511 pump Paradigm",
          "vendor": "Medtronic",
          "versions": [
            {
              "status": "affected",
              "version": "All versions"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "MMT \u2013 512 / MMT \u2013 712 Paradigm x12",
          "vendor": "Medtronic",
          "versions": [
            {
              "status": "affected",
              "version": "All versions"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "MMT \u2013 515 / MMT \u2013 715 Paradigm x15",
          "vendor": "Medtronic",
          "versions": [
            {
              "status": "affected",
              "version": "All versions"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "MMT \u2013 522 / MMT \u2013 722 Paradigm REAL-TIME",
          "vendor": "Medtronic",
          "versions": [
            {
              "status": "affected",
              "version": "All versions"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "MMT \u2013 522(K) / MMT \u2013 722(K) Paradigm REAL-TIME",
          "vendor": "Medtronic",
          "versions": [
            {
              "status": "affected",
              "version": "All versions"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "MMT \u2013 523 / MMT \u2013 723 Paradigm Revel",
          "vendor": "Medtronic",
          "versions": [
            {
              "status": "affected",
              "version": "All versions"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "MMT \u2013 523(K) / MMT \u2013 723(K) Paradigm",
          "vendor": "Medtronic",
          "versions": [
            {
              "status": "affected",
              "version": "All versions"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "MMT \u2013 554 / MMT \u2013 754 MiniMed Veo",
          "vendor": "Medtronic",
          "versions": [
            {
              "status": "affected",
              "version": "All versions"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "MMT \u2013 551 / MMT \u2013 751 MiniMed 530G",
          "vendor": "Medtronic",
          "versions": [
            {
              "status": "affected",
              "version": "All versions"
            }
          ]
        }
      ],
      "source": "ics-cert@hq.dhs.gov"
    }
  ],
  "configurations": [
    {
      "nodes": [
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:medtronicdiabetes:508_minimed_insulin_pump_firmware:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "EF9C0D70-7924-414E-98EF-245E52ED8838",
              "vulnerable": true
            }
          ],
          "negate": false,
          "operator": "OR"
        },
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:medtronicdiabetes:508_minimed_insulin_pump:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "FF20B7E1-C7E3-45E6-A981-6A02807A2411",
              "vulnerable": false
            }
          ],
          "negate": false,
          "operator": "OR"
        }
      ],
      "operator": "AND"
    },
    {
      "nodes": [
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:medtronicdiabetes:522_paradigm_real-time_firmware:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "3A8E9BE8-E17F-490D-BCE7-92D43BE14574",
              "vulnerable": true
            }
          ],
          "negate": false,
          "operator": "OR"
        },
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:medtronicdiabetes:522_paradigm_real-time:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "CCD29F70-6322-4891-8036-6BDE7F98B08A",
              "vulnerable": false
            }
          ],
          "negate": false,
          "operator": "OR"
        }
      ],
      "operator": "AND"
    },
    {
      "nodes": [
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:medtronicdiabetes:722_paradigm_real-time_firmware:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "95CAA730-1862-4D91-B962-28A2D8EB07F2",
              "vulnerable": true
            }
          ],
          "negate": false,
          "operator": "OR"
        },
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:medtronicdiabetes:722_paradigm_real-time:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "AB82801B-A8AA-47C9-BC9E-A4FCCC4BF10C",
              "vulnerable": false
            }
          ],
          "negate": false,
          "operator": "OR"
        }
      ],
      "operator": "AND"
    },
    {
      "nodes": [
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:medtronicdiabetes:523_paradigm_revel_firmware:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "2C8DF53F-B5DB-4690-87E8-8CECF77783EE",
              "vulnerable": true
            }
          ],
          "negate": false,
          "operator": "OR"
        },
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:medtronicdiabetes:523_paradigm_revel:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "95A2BE4F-FE1B-496C-8237-73F8BF0639E6",
              "vulnerable": false
            }
          ],
          "negate": false,
          "operator": "OR"
        }
      ],
      "operator": "AND"
    },
    {
      "nodes": [
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:medtronicdiabetes:723_paradigm_revel_firmware:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "A1F8021A-E8CD-4B6A-979D-5E3F18EB6B4A",
              "vulnerable": true
            }
          ],
          "negate": false,
          "operator": "OR"
        },
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:medtronicdiabetes:723_paradigm_revel:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "AB2EF040-CE07-4430-8707-689666B7AE61",
              "vulnerable": false
            }
          ],
          "negate": false,
          "operator": "OR"
        }
      ],
      "operator": "AND"
    },
    {
      "nodes": [
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:medtronicdiabetes:523k_paradigm_revel_firmware:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "175AE115-73FA-4A1E-B8D6-185522D8AAEC",
              "vulnerable": true
            }
          ],
          "negate": false,
          "operator": "OR"
        },
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:medtronicdiabetes:523k_paradigm_revel:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "425712F3-B98B-4D0E-97B9-E94D5216A43F",
              "vulnerable": false
            }
          ],
          "negate": false,
          "operator": "OR"
        }
      ],
      "operator": "AND"
    },
    {
      "nodes": [
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:medtronicdiabetes:723k_paradigm_revel_firmware:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "83888F47-99B8-4D7F-9049-6B9168BBFA70",
              "vulnerable": true
            }
          ],
          "negate": false,
          "operator": "OR"
        },
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:medtronicdiabetes:723k_paradigm_revel:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "04627C77-7F76-4883-8C6D-77579F4AF28E",
              "vulnerable": false
            }
          ],
          "negate": false,
          "operator": "OR"
        }
      ],
      "operator": "AND"
    },
    {
      "nodes": [
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:medtronicdiabetes:551_minimed_530g_firmware:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "86BAFB4D-49CF-4855-8A84-D41BD80A16D4",
              "vulnerable": true
            }
          ],
          "negate": false,
          "operator": "OR"
        },
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:medtronicdiabetes:551_minimed_530g:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "2AE48E5F-E955-4351-BC2E-E6EF0EB40AAB",
              "vulnerable": false
            }
          ],
          "negate": false,
          "operator": "OR"
        }
      ],
      "operator": "AND"
    },
    {
      "nodes": [
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:medtronicdiabetes:751_minimed_530g_firmware:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "8AE64752-1F20-4F8B-9048-4B004A3B3CB2",
              "vulnerable": true
            }
          ],
          "negate": false,
          "operator": "OR"
        },
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:medtronicdiabetes:751_minimed_530g:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "B63924E6-F2F3-47A9-BCDC-5EC4B665DC9D",
              "vulnerable": false
            }
          ],
          "negate": false,
          "operator": "OR"
        }
      ],
      "operator": "AND"
    }
  ],
  "cveTags": [],
  "descriptions": [
    {
      "lang": "en",
      "value": "Medtronic MiniMed MMT \n\ndevices when paired with a remote controller and having the \u201ceasy bolus\u201d and \u201cremote bolus\u201d options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery."
    },
    {
      "lang": "es",
      "value": "En la v\u00e1lvula de insulina de Medtronic MMT 508 MiniMed, 522 / MMT - 722 Paradigm REAL-TIME, 523 / MMT - 723 Paradigm Revel, 523K / MMT - 723K Paradigm Revel y 551 / MMT - 751 MiniMed 530G, estos modelos, cuando se emparejan con un controlador remoto y tienen las opciones \"easy bolus\" y \"remote bolus\" habilitadas (no por defecto), son vulnerables a un ataque de captura y reproducci\u00f3n. Un atacante puede capturar las transmisiones inal\u00e1mbricas entre el controlador remoto y la v\u00e1lvula y reproducirlas para provocar una inyecci\u00f3n de insulina (bolus)."
    }
  ],
  "id": "CVE-2018-14781",
  "lastModified": "2026-06-17T01:41:37.633",
  "metrics": {
    "cvssMetricV2": [
      {
        "acInsufInfo": false,
        "baseSeverity": "LOW",
        "cvssData": {
          "accessComplexity": "MEDIUM",
          "accessVector": "ADJACENT_NETWORK",
          "authentication": "NONE",
          "availabilityImpact": "NONE",
          "baseScore": 2.9,
          "confidentialityImpact": "NONE",
          "integrityImpact": "PARTIAL",
          "vectorString": "AV:A/AC:M/Au:N/C:N/I:P/A:N",
          "version": "2.0"
        },
        "exploitabilityScore": 5.5,
        "impactScore": 2.9,
        "obtainAllPrivilege": false,
        "obtainOtherPrivilege": false,
        "obtainUserPrivilege": false,
        "source": "nvd@nist.gov",
        "type": "Primary",
        "userInteractionRequired": false
      }
    ],
    "cvssMetricV30": [
      {
        "cvssData": {
          "attackComplexity": "HIGH",
          "attackVector": "ADJACENT_NETWORK",
          "availabilityImpact": "NONE",
          "baseScore": 5.3,
          "baseSeverity": "MEDIUM",
          "confidentialityImpact": "NONE",
          "integrityImpact": "HIGH",
          "privilegesRequired": "NONE",
          "scope": "UNCHANGED",
          "userInteraction": "NONE",
          "vectorString": "CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N",
          "version": "3.0"
        },
        "exploitabilityScore": 1.6,
        "impactScore": 3.6,
        "source": "nvd@nist.gov",
        "type": "Primary"
      }
    ],
    "cvssMetricV31": [
      {
        "cvssData": {
          "attackComplexity": "HIGH",
          "attackVector": "ADJACENT_NETWORK",
          "availabilityImpact": "NONE",
          "baseScore": 5.3,
          "baseSeverity": "MEDIUM",
          "confidentialityImpact": "NONE",
          "integrityImpact": "HIGH",
          "privilegesRequired": "NONE",
          "scope": "UNCHANGED",
          "userInteraction": "NONE",
          "vectorString": "CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N",
          "version": "3.1"
        },
        "exploitabilityScore": 1.6,
        "impactScore": 3.6,
        "source": "ics-cert@hq.dhs.gov",
        "type": "Secondary"
      }
    ]
  },
  "published": "2018-08-13T21:48:01.227",
  "references": [
    {
      "source": "ics-cert@hq.dhs.gov",
      "tags": [
        "Third Party Advisory",
        "VDB Entry"
      ],
      "url": "http://www.securityfocus.com/bid/105044"
    },
    {
      "source": "ics-cert@hq.dhs.gov",
      "url": "https://global.medtronic.com/xg-en/product-security/security-bulletins/minimed.html"
    },
    {
      "source": "ics-cert@hq.dhs.gov",
      "tags": [
        "Third Party Advisory",
        "US Government Resource"
      ],
      "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-02"
    },
    {
      "source": "af854a3a-2127-422b-91ae-364da2661108",
      "tags": [
        "Third Party Advisory",
        "VDB Entry"
      ],
      "url": "http://www.securityfocus.com/bid/105044"
    },
    {
      "source": "af854a3a-2127-422b-91ae-364da2661108",
      "tags": [
        "Third Party Advisory",
        "US Government Resource"
      ],
      "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-02"
    }
  ],
  "sourceIdentifier": "ics-cert@hq.dhs.gov",
  "vulnStatus": "Modified",
  "weaknesses": [
    {
      "description": [
        {
          "lang": "en",
          "value": "CWE-294"
        }
      ],
      "source": "ics-cert@hq.dhs.gov",
      "type": "Secondary"
    },
    {
      "description": [
        {
          "lang": "en",
          "value": "CWE-287"
        }
      ],
      "source": "nvd@nist.gov",
      "type": "Primary"
    }
  ]
}



Log in or create an account to share your comment.




Tags
Taxonomy of the tags.


Loading…

Loading…

Loading…

Forecast uses a logistic model when the trend is rising, or an exponential decay model when the trend is falling. Fitted via linearized least squares.

Sightings

Author Source Type Date Other

Nomenclature

  • Seen: The vulnerability was mentioned, discussed, or observed by the user.
  • Confirmed: The vulnerability has been validated from an analyst's perspective.
  • Published Proof of Concept: A public proof of concept is available for this vulnerability.
  • Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
  • Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
  • Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
  • Not confirmed: The user expressed doubt about the validity of the vulnerability.
  • Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.

Loading…

Detection rules are retrieved from Rulezet.

Loading…

Loading…

Related by attack behaviour

Vulnerabilities whose description is nearest to this one in the vector space of the CIRCL/vulnerability-attack-technique-biencoder model. This is a similarity search over the bi-encoder space (plain cosine), not a classification, and it has no measured accuracy.


Loading…