CVE-2025-14096 (GCVE-0-2025-14096)
Vulnerability from cvelistv5 – Published: 2025-12-17 12:19 – Updated: 2025-12-17 14:43
VLAI
EPSS
VEX
Title
Credential Disclosure vulnerability in Radiometer Products
Summary
A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.
Other related CVE's are CVE-2025-14095 & CVE-2025-14097.
Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.
Required Configuration for Exposure:
Attacker requires physical access to the analyzer.
Temporary work Around:
Only authorized people can physically access the analyzer.
Permanent solution:
Local Radiometer representatives will contact all affected customers to discuss a permanent solution.
Exploit Status:
Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.
Severity
8.4 (High)
SSVC
Exploitation: none
Automatable: no
Technical Impact: total
CISA Coordinator · CISA-ADP (v2.0.3)
Decision recorded 2025-12-17 14:42 UTC
Assigner
References
1 reference
Impacted products
3 products
| Vendor | Product | Version | CPE status | |
|---|---|---|---|---|
| Radiometer Medical Aps | ABL90 FLEX and ABL90 FLEX PLUS Analyzers |
Affected:
Windows 7 Operating system
(Radiometer Customized Windows Operating System)
Affected: Windows XP Operating system (Radiometer Customized Windows Operating System) |
guessed | |
| Radiometer Medical Aps | AQT90 FLEX Analyzers |
Affected:
Windows 7 Operating system
(Radiometer Customized Windows Operating System)
Affected: Windows XP Operating system (Radiometer Customized Windows Operating System) |
guessed | |
| Radiometer Medical Aps | ABL800 BASIC and ABL800 FLEX analyzers |
Affected:
Windows 7 Operating system
(Radiometer Customized Windows Operating System)
Affected: Windows XP Operating system (Radiometer Customized Windows Operating System) |
guessed |
{
"containers": {
"adp": [
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2025-14096",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "total"
}
],
"role": "CISA Coordinator",
"timestamp": "2025-12-17T14:42:57.843796Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2025-12-17T14:43:16.523Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "unaffected",
"platforms": [
"Windows XP",
"Windows 7"
],
"product": "ABL90 FLEX and ABL90 FLEX PLUS Analyzers",
"vendor": "Radiometer Medical Aps",
"versions": [
{
"status": "affected",
"version": "Windows 7 Operating system",
"versionType": "Radiometer Customized Windows Operating System"
},
{
"status": "affected",
"version": "Windows XP Operating system",
"versionType": "Radiometer Customized Windows Operating System"
}
]
},
{
"defaultStatus": "unaffected",
"platforms": [
"Windows 7",
"Windows XP"
],
"product": "AQT90 FLEX Analyzers",
"vendor": "Radiometer Medical Aps",
"versions": [
{
"status": "affected",
"version": "Windows 7 Operating system",
"versionType": "Radiometer Customized Windows Operating System"
},
{
"status": "affected",
"version": "Windows XP Operating system",
"versionType": "Radiometer Customized Windows Operating System"
}
]
},
{
"defaultStatus": "unaffected",
"platforms": [
"Windows 7",
"Windows XP"
],
"product": "ABL800 BASIC and ABL800 FLEX analyzers",
"vendor": "Radiometer Medical Aps",
"versions": [
{
"status": "affected",
"version": "Windows 7 Operating system",
"versionType": "Radiometer Customized Windows Operating System"
},
{
"status": "affected",
"version": "Windows XP Operating system",
"versionType": "Radiometer Customized Windows Operating System"
}
]
}
],
"configurations": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "Requires physical access to the Analyzer.\n\n\u003cbr\u003e"
}
],
"value": "Requires physical access to the Analyzer."
}
],
"credits": [
{
"lang": "en",
"type": "reporter",
"value": "Florian Hauser and Fabian Weber from CODE WHITE GmbH"
}
],
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.\u003cbr\u003e\u003cbr\u003eOther related CVE\u0027s are CVE-2025-14095 \u0026amp; CVE-2025-14097.\u003cbr\u003e\u003cbr\u003e\n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eAffected customers have been informed about this vulnerability. This CVE is being published to provide transparency.\u003c/span\u003e\n\n\u003cbr\u003e\u003cbr\u003e\n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eRequired Configuration for Exposure:\u003cbr\u003e\u003c/span\u003e\n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eAttacker requires physical access to the analyzer.\u003c/span\u003e\n\n\u003cbr\u003e\u003cbr\u003eTemporary work Around:\u003cbr\u003eOnly authorized people can physically access the analyzer.\u003cbr\u003e\u003cbr\u003ePermanent solution:\u003cbr\u003eLocal Radiometer representatives will contact all affected customers to discuss a permanent solution.\u003cbr\u003e\u003cbr\u003eExploit Status:\u003cbr\u003e\n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eResearchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.\u003c/span\u003e\n\n\u003cbr\u003e\u003cbr\u003e"
}
],
"value": "A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.\n\nOther related CVE\u0027s are CVE-2025-14095 \u0026 CVE-2025-14097.\n\n\n\nAffected customers have been informed about this vulnerability. This CVE is being published to provide transparency.\n\n\n\n\n\nRequired Configuration for Exposure:\n\n\nAttacker requires physical access to the analyzer.\n\n\n\nTemporary work Around:\nOnly authorized people can physically access the analyzer.\n\nPermanent solution:\nLocal Radiometer representatives will contact all affected customers to discuss a permanent solution.\n\nExploit Status:\n\n\nResearchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication."
}
],
"exploits": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.\n\n\u003cbr\u003e"
}
],
"value": "Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication."
}
],
"impacts": [
{
"capecId": "CAPEC-69",
"descriptions": [
{
"lang": "en",
"value": "CAPEC-69 \u2014 Target Programs with Elevated Privileges"
}
]
}
],
"metrics": [
{
"cvssV3_1": {
"attackComplexity": "LOW",
"attackVector": "LOCAL",
"availabilityImpact": "HIGH",
"baseScore": 8.4,
"baseSeverity": "HIGH",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
"version": "3.1"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-798",
"description": "CWE-798: Use of Hard-coded Credentials",
"lang": "en",
"type": "CWE"
}
]
},
{
"descriptions": [
{
"cweId": "CWE-250",
"description": "CWE-250: Execution with Unnecessary Privileges",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2025-12-17T13:03:05.696Z",
"orgId": "46b595e9-1acc-41cb-9398-adaf98d37a9b",
"shortName": "Radiometer"
},
"references": [
{
"url": "https://www.radiometer.com/myradiometer"
}
],
"solutions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "Local Radiometer representatives will contact all affected customers to discuss a permanent solution.\n\n\n\n\u003cbr\u003e"
}
],
"value": "Local Radiometer representatives will contact all affected customers to discuss a permanent solution."
}
],
"source": {
"discovery": "EXTERNAL"
},
"timeline": [
{
"lang": "en",
"time": "2025-12-17T13:01:00.000Z",
"value": "CVE Published"
}
],
"title": "Credential Disclosure vulnerability in Radiometer Products",
"workarounds": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "Only authorized people can physically access the analyzer.\n\n\n\n\u003cbr\u003e"
}
],
"value": "Only authorized people can physically access the analyzer."
}
],
"x_generator": {
"engine": "Vulnogram 0.5.0"
}
}
},
"cveMetadata": {
"assignerOrgId": "46b595e9-1acc-41cb-9398-adaf98d37a9b",
"assignerShortName": "Radiometer",
"cveId": "CVE-2025-14096",
"datePublished": "2025-12-17T12:19:27.489Z",
"dateReserved": "2025-12-05T10:50:00.566Z",
"dateUpdated": "2025-12-17T14:43:16.523Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.2",
"vulnerability-lookup:meta": {
"epss": {
"cve": "CVE-2025-14096",
"date": "2026-10-19",
"epss": "0.00135",
"percentile": "0.0264"
},
"nvd": {
"cve": {
"affected": [
{
"affectedData": [
{
"defaultStatus": "unaffected",
"platforms": [
"Windows XP",
"Windows 7"
],
"product": "ABL90 FLEX and ABL90 FLEX PLUS Analyzers",
"vendor": "Radiometer Medical Aps",
"versions": [
{
"status": "affected",
"version": "Windows 7 Operating system",
"versionType": "Radiometer Customized Windows Operating System"
},
{
"status": "affected",
"version": "Windows XP Operating system",
"versionType": "Radiometer Customized Windows Operating System"
}
]
},
{
"defaultStatus": "unaffected",
"platforms": [
"Windows 7",
"Windows XP"
],
"product": "AQT90 FLEX Analyzers",
"vendor": "Radiometer Medical Aps",
"versions": [
{
"status": "affected",
"version": "Windows 7 Operating system",
"versionType": "Radiometer Customized Windows Operating System"
},
{
"status": "affected",
"version": "Windows XP Operating system",
"versionType": "Radiometer Customized Windows Operating System"
}
]
},
{
"defaultStatus": "unaffected",
"platforms": [
"Windows 7",
"Windows XP"
],
"product": "ABL800 BASIC and ABL800 FLEX analyzers",
"vendor": "Radiometer Medical Aps",
"versions": [
{
"status": "affected",
"version": "Windows 7 Operating system",
"versionType": "Radiometer Customized Windows Operating System"
},
{
"status": "affected",
"version": "Windows XP Operating system",
"versionType": "Radiometer Customized Windows Operating System"
}
]
}
],
"source": "46b595e9-1acc-41cb-9398-adaf98d37a9b"
}
],
"cveTags": [],
"descriptions": [
{
"lang": "en",
"value": "A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.\n\nOther related CVE\u0027s are CVE-2025-14095 \u0026 CVE-2025-14097.\n\n\n\nAffected customers have been informed about this vulnerability. This CVE is being published to provide transparency.\n\n\n\n\n\nRequired Configuration for Exposure:\n\n\nAttacker requires physical access to the analyzer.\n\n\n\nTemporary work Around:\nOnly authorized people can physically access the analyzer.\n\nPermanent solution:\nLocal Radiometer representatives will contact all affected customers to discuss a permanent solution.\n\nExploit Status:\n\n\nResearchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication."
}
],
"id": "CVE-2025-14096",
"lastModified": "2026-06-17T08:35:19.237",
"metrics": {
"cvssMetricV31": [
{
"cvssData": {
"attackComplexity": "LOW",
"attackVector": "LOCAL",
"availabilityImpact": "HIGH",
"baseScore": 8.4,
"baseSeverity": "HIGH",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
"version": "3.1"
},
"exploitabilityScore": 2.5,
"impactScore": 5.9,
"source": "46b595e9-1acc-41cb-9398-adaf98d37a9b",
"type": "Secondary"
}
],
"ssvcV203": [
{
"source": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"ssvcData": {
"id": "CVE-2025-14096",
"options": [
{
"exploitation": "none"
},
{
"automatable": "no"
},
{
"technicalImpact": "total"
}
],
"role": "CISA Coordinator",
"timestamp": "2025-12-17T14:42:57.843796Z",
"version": "2.0.3"
}
}
]
},
"published": "2025-12-17T13:15:57.890",
"references": [
{
"source": "46b595e9-1acc-41cb-9398-adaf98d37a9b",
"url": "https://www.radiometer.com/myradiometer"
}
],
"sourceIdentifier": "46b595e9-1acc-41cb-9398-adaf98d37a9b",
"vulnStatus": "Deferred",
"weaknesses": [
{
"description": [
{
"lang": "en",
"value": "CWE-250"
},
{
"lang": "en",
"value": "CWE-798"
}
],
"source": "46b595e9-1acc-41cb-9398-adaf98d37a9b",
"type": "Secondary"
}
]
}
},
"redhat_vex": {
"current_release_date": "2025-12-18T12:39:55+00:00",
"cve": "CVE-2025-14096",
"id": "CVE-2025-14096",
"initial_release_date": "2025-01-01T00:00:00+00:00",
"product_status:known_not_affected": "1",
"source": "Red Hat CSAF VEX",
"status": "final",
"title": "Credential Disclosure vulnerability in Radiometer Products",
"url": "https://security.access.redhat.com/data/csaf/v2/vex/2025/cve-2025-14096.json",
"version": "3"
},
"vulnrichment": {
"containers": {
"adp": [
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2025-14096",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "total"
}
],
"role": "CISA Coordinator",
"timestamp": "2025-12-17T14:42:57.843796Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2025-12-17T14:43:12.844Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "unaffected",
"platforms": [
"Windows XP",
"Windows 7"
],
"product": "ABL90 FLEX and ABL90 FLEX PLUS Analyzers",
"vendor": "Radiometer Medical Aps",
"versions": [
{
"status": "affected",
"version": "Windows 7 Operating system",
"versionType": "Radiometer Customized Windows Operating System"
},
{
"status": "affected",
"version": "Windows XP Operating system",
"versionType": "Radiometer Customized Windows Operating System"
}
]
},
{
"defaultStatus": "unaffected",
"platforms": [
"Windows 7",
"Windows XP"
],
"product": "AQT90 FLEX Analyzers",
"vendor": "Radiometer Medical Aps",
"versions": [
{
"status": "affected",
"version": "Windows 7 Operating system",
"versionType": "Radiometer Customized Windows Operating System"
},
{
"status": "affected",
"version": "Windows XP Operating system",
"versionType": "Radiometer Customized Windows Operating System"
}
]
},
{
"defaultStatus": "unaffected",
"platforms": [
"Windows 7",
"Windows XP"
],
"product": "ABL800 BASIC and ABL800 FLEX analyzers",
"vendor": "Radiometer Medical Aps",
"versions": [
{
"status": "affected",
"version": "Windows 7 Operating system",
"versionType": "Radiometer Customized Windows Operating System"
},
{
"status": "affected",
"version": "Windows XP Operating system",
"versionType": "Radiometer Customized Windows Operating System"
}
]
}
],
"configurations": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "Requires physical access to the Analyzer.\n\n\u003cbr\u003e"
}
],
"value": "Requires physical access to the Analyzer."
}
],
"credits": [
{
"lang": "en",
"type": "reporter",
"value": "Florian Hauser and Fabian Weber from CODE WHITE GmbH"
}
],
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.\u003cbr\u003e\u003cbr\u003eOther related CVE\u0027s are CVE-2025-14095 \u0026amp; CVE-2025-14097.\u003cbr\u003e\u003cbr\u003e\n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eAffected customers have been informed about this vulnerability. This CVE is being published to provide transparency.\u003c/span\u003e\n\n\u003cbr\u003e\u003cbr\u003e\n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eRequired Configuration for Exposure:\u003cbr\u003e\u003c/span\u003e\n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eAttacker requires physical access to the analyzer.\u003c/span\u003e\n\n\u003cbr\u003e\u003cbr\u003eTemporary work Around:\u003cbr\u003eOnly authorized people can physically access the analyzer.\u003cbr\u003e\u003cbr\u003ePermanent solution:\u003cbr\u003eLocal Radiometer representatives will contact all affected customers to discuss a permanent solution.\u003cbr\u003e\u003cbr\u003eExploit Status:\u003cbr\u003e\n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eResearchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.\u003c/span\u003e\n\n\u003cbr\u003e\u003cbr\u003e"
}
],
"value": "A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.\n\nOther related CVE\u0027s are CVE-2025-14095 \u0026 CVE-2025-14097.\n\n\n\nAffected customers have been informed about this vulnerability. This CVE is being published to provide transparency.\n\n\n\n\n\nRequired Configuration for Exposure:\n\n\nAttacker requires physical access to the analyzer.\n\n\n\nTemporary work Around:\nOnly authorized people can physically access the analyzer.\n\nPermanent solution:\nLocal Radiometer representatives will contact all affected customers to discuss a permanent solution.\n\nExploit Status:\n\n\nResearchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication."
}
],
"exploits": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.\n\n\u003cbr\u003e"
}
],
"value": "Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication."
}
],
"impacts": [
{
"capecId": "CAPEC-69",
"descriptions": [
{
"lang": "en",
"value": "CAPEC-69 \u2014 Target Programs with Elevated Privileges"
}
]
}
],
"metrics": [
{
"cvssV3_1": {
"attackComplexity": "LOW",
"attackVector": "LOCAL",
"availabilityImpact": "HIGH",
"baseScore": 8.4,
"baseSeverity": "HIGH",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
"version": "3.1"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-798",
"description": "CWE-798: Use of Hard-coded Credentials",
"lang": "en",
"type": "CWE"
}
]
},
{
"descriptions": [
{
"cweId": "CWE-250",
"description": "CWE-250: Execution with Unnecessary Privileges",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2025-12-17T13:03:05.696Z",
"orgId": "46b595e9-1acc-41cb-9398-adaf98d37a9b",
"shortName": "Radiometer"
},
"references": [
{
"url": "https://www.radiometer.com/myradiometer"
}
],
"solutions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "Local Radiometer representatives will contact all affected customers to discuss a permanent solution.\n\n\n\n\u003cbr\u003e"
}
],
"value": "Local Radiometer representatives will contact all affected customers to discuss a permanent solution."
}
],
"source": {
"discovery": "EXTERNAL"
},
"timeline": [
{
"lang": "en",
"time": "2025-12-17T13:01:00.000Z",
"value": "CVE Published"
}
],
"title": "Credential Disclosure vulnerability in Radiometer Products",
"workarounds": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "Only authorized people can physically access the analyzer.\n\n\n\n\u003cbr\u003e"
}
],
"value": "Only authorized people can physically access the analyzer."
}
],
"x_generator": {
"engine": "Vulnogram 0.5.0"
}
}
},
"cveMetadata": {
"assignerOrgId": "46b595e9-1acc-41cb-9398-adaf98d37a9b",
"assignerShortName": "Radiometer",
"cveId": "CVE-2025-14096",
"datePublished": "2025-12-17T12:19:27.489Z",
"dateReserved": "2025-12-05T10:50:00.566Z",
"dateUpdated": "2025-12-17T14:43:16.523Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.2"
}
}
}
Loading…
Loading…
Drag nodes to arrange them. Double-click a related vulnerability to open it, or a source to visit its link. The Physics button on the left tunes the layout and the simulation; the legend at the bottom left hides or shows nodes by kind and by sighting type.
Experimental. This forecast is provided for visualization only and may change without notice. Do not use it for operational decisions.
Forecast uses a logistic model when the trend is rising, or an exponential decay model when the trend is falling. Fitted via linearized least squares.
Sightings
| Author | Source | Type | Date | Other |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or observed by the user.
- Confirmed: The vulnerability has been validated from an analyst's perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
- Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
- Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
- Not confirmed: The user expressed doubt about the validity of the vulnerability.
- Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.
Vulnerability timeline
Sightings and recorded dates, grouped by day · UTC
Loading timeline…
Loading…
Loading…
The MITRE ATT&CK techniques below are AI-generated suggestions, inferred from the description of the
vulnerability by the CIRCL/vulnerability-attack-technique-classification-roberta-base
model, served locally by ML-Gateway.
They have not been verified by an analyst and are provided for guidance only.
The approach is described in our paper Mapping CVEs to MITRE ATT&CK Techniques: A Curated Gold-Set Classifier and the Limits of LLM-Assisted Label Expansion.
Browse all ATT&CK techniques and the vulnerabilities related to each.
The approach is described in our paper Mapping CVEs to MITRE ATT&CK Techniques: A Curated Gold-Set Classifier and the Limits of LLM-Assisted Label Expansion.
Browse all ATT&CK techniques and the vulnerabilities related to each.
Loading…
Related by attack behaviour
Vulnerabilities whose description is nearest to this one in the vector space of the CIRCL/vulnerability-attack-technique-biencoder model. This is a similarity search over the bi-encoder space (plain cosine), not a classification, and it has no measured accuracy.