CVE-2025-14096 (GCVE-0-2025-14096)
Vulnerability from cvelistv5 – Published: 2025-12-17 12:19 – Updated: 2025-12-17 14:43
VLAI
EPSS
VEX
Title
Credential Disclosure vulnerability in Radiometer Products
Summary
A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.
Other related CVE's are CVE-2025-14095 & CVE-2025-14097.
Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.
Required Configuration for Exposure:
Attacker requires physical access to the analyzer.
Temporary work Around:
Only authorized people can physically access the analyzer.
Permanent solution:
Local Radiometer representatives will contact all affected customers to discuss a permanent solution.
Exploit Status:
Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.
Severity
8.4 (High)
SSVC
Exploitation: none
Automatable: no
Technical Impact: total
CISA Coordinator · CISA-ADP (v2.0.3)
Decision recorded 2025-12-17 14:42 UTC
Assigner
References
1 reference
Impacted products
3 products
| Vendor | Product | Version | CPE status | |
|---|---|---|---|---|
| Radiometer Medical Aps | ABL90 FLEX and ABL90 FLEX PLUS Analyzers |
Affected:
Windows 7 Operating system
(Radiometer Customized Windows Operating System)
Affected: Windows XP Operating system (Radiometer Customized Windows Operating System) |
guessed | |
| Radiometer Medical Aps | AQT90 FLEX Analyzers |
Affected:
Windows 7 Operating system
(Radiometer Customized Windows Operating System)
Affected: Windows XP Operating system (Radiometer Customized Windows Operating System) |
guessed | |
| Radiometer Medical Aps | ABL800 BASIC and ABL800 FLEX analyzers |
Affected:
Windows 7 Operating system
(Radiometer Customized Windows Operating System)
Affected: Windows XP Operating system (Radiometer Customized Windows Operating System) |
guessed |
{
"containers": {
"adp": [
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2025-14096",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "total"
}
],
"role": "CISA Coordinator",
"timestamp": "2025-12-17T14:42:57.843796Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2025-12-17T14:43:16.523Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "unaffected",
"platforms": [
"Windows XP",
"Windows 7"
],
"product": "ABL90 FLEX and ABL90 FLEX PLUS Analyzers",
"vendor": "Radiometer Medical Aps",
"versions": [
{
"status": "affected",
"version": "Windows 7 Operating system",
"versionType": "Radiometer Customized Windows Operating System"
},
{
"status": "affected",
"version": "Windows XP Operating system",
"versionType": "Radiometer Customized Windows Operating System"
}
]
},
{
"defaultStatus": "unaffected",
"platforms": [
"Windows 7",
"Windows XP"
],
"product": "AQT90 FLEX Analyzers",
"vendor": "Radiometer Medical Aps",
"versions": [
{
"status": "affected",
"version": "Windows 7 Operating system",
"versionType": "Radiometer Customized Windows Operating System"
},
{
"status": "affected",
"version": "Windows XP Operating system",
"versionType": "Radiometer Customized Windows Operating System"
}
]
},
{
"defaultStatus": "unaffected",
"platforms": [
"Windows 7",
"Windows XP"
],
"product": "ABL800 BASIC and ABL800 FLEX analyzers",
"vendor": "Radiometer Medical Aps",
"versions": [
{
"status": "affected",
"version": "Windows 7 Operating system",
"versionType": "Radiometer Customized Windows Operating System"
},
{
"status": "affected",
"version": "Windows XP Operating system",
"versionType": "Radiometer Customized Windows Operating System"
}
]
}
],
"configurations": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "Requires physical access to the Analyzer.\n\n\u003cbr\u003e"
}
],
"value": "Requires physical access to the Analyzer."
}
],
"credits": [
{
"lang": "en",
"type": "reporter",
"value": "Florian Hauser and Fabian Weber from CODE WHITE GmbH"
}
],
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.\u003cbr\u003e\u003cbr\u003eOther related CVE\u0027s are CVE-2025-14095 \u0026amp; CVE-2025-14097.\u003cbr\u003e\u003cbr\u003e\n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eAffected customers have been informed about this vulnerability. This CVE is being published to provide transparency.\u003c/span\u003e\n\n\u003cbr\u003e\u003cbr\u003e\n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eRequired Configuration for Exposure:\u003cbr\u003e\u003c/span\u003e\n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eAttacker requires physical access to the analyzer.\u003c/span\u003e\n\n\u003cbr\u003e\u003cbr\u003eTemporary work Around:\u003cbr\u003eOnly authorized people can physically access the analyzer.\u003cbr\u003e\u003cbr\u003ePermanent solution:\u003cbr\u003eLocal Radiometer representatives will contact all affected customers to discuss a permanent solution.\u003cbr\u003e\u003cbr\u003eExploit Status:\u003cbr\u003e\n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eResearchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.\u003c/span\u003e\n\n\u003cbr\u003e\u003cbr\u003e"
}
],
"value": "A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.\n\nOther related CVE\u0027s are CVE-2025-14095 \u0026 CVE-2025-14097.\n\n\n\nAffected customers have been informed about this vulnerability. This CVE is being published to provide transparency.\n\n\n\n\n\nRequired Configuration for Exposure:\n\n\nAttacker requires physical access to the analyzer.\n\n\n\nTemporary work Around:\nOnly authorized people can physically access the analyzer.\n\nPermanent solution:\nLocal Radiometer representatives will contact all affected customers to discuss a permanent solution.\n\nExploit Status:\n\n\nResearchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication."
}
],
"exploits": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.\n\n\u003cbr\u003e"
}
],
"value": "Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication."
}
],
"impacts": [
{
"capecId": "CAPEC-69",
"descriptions": [
{
"lang": "en",
"value": "CAPEC-69 \u2014 Target Programs with Elevated Privileges"
}
]
}
],
"metrics": [
{
"cvssV3_1": {
"attackComplexity": "LOW",
"attackVector": "LOCAL",
"availabilityImpact": "HIGH",
"baseScore": 8.4,
"baseSeverity": "HIGH",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
"version": "3.1"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-798",
"description": "CWE-798: Use of Hard-coded Credentials",
"lang": "en",
"type": "CWE"
}
]
},
{
"descriptions": [
{
"cweId": "CWE-250",
"description": "CWE-250: Execution with Unnecessary Privileges",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2025-12-17T13:03:05.696Z",
"orgId": "46b595e9-1acc-41cb-9398-adaf98d37a9b",
"shortName": "Radiometer"
},
"references": [
{
"url": "https://www.radiometer.com/myradiometer"
}
],
"solutions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "Local Radiometer representatives will contact all affected customers to discuss a permanent solution.\n\n\n\n\u003cbr\u003e"
}
],
"value": "Local Radiometer representatives will contact all affected customers to discuss a permanent solution."
}
],
"source": {
"discovery": "EXTERNAL"
},
"timeline": [
{
"lang": "en",
"time": "2025-12-17T13:01:00.000Z",
"value": "CVE Published"
}
],
"title": "Credential Disclosure vulnerability in Radiometer Products",
"workarounds": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "Only authorized people can physically access the analyzer.\n\n\n\n\u003cbr\u003e"
}
],
"value": "Only authorized people can physically access the analyzer."
}
],
"x_generator": {
"engine": "Vulnogram 0.5.0"
}
}
},
"cveMetadata": {
"assignerOrgId": "46b595e9-1acc-41cb-9398-adaf98d37a9b",
"assignerShortName": "Radiometer",
"cveId": "CVE-2025-14096",
"datePublished": "2025-12-17T12:19:27.489Z",
"dateReserved": "2025-12-05T10:50:00.566Z",
"dateUpdated": "2025-12-17T14:43:16.523Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.2",
"vulnerability-lookup:meta": {
"epss": {
"cve": "CVE-2025-14096",
"date": "2026-09-20",
"epss": "0.00135",
"percentile": "0.03337"
},
"nvd": {
"cve": {
"affected": [
{
"affectedData": [
{
"defaultStatus": "unaffected",
"platforms": [
"Windows XP",
"Windows 7"
],
"product": "ABL90 FLEX and ABL90 FLEX PLUS Analyzers",
"vendor": "Radiometer Medical Aps",
"versions": [
{
"status": "affected",
"version": "Windows 7 Operating system",
"versionType": "Radiometer Customized Windows Operating System"
},
{
"status": "affected",
"version": "Windows XP Operating system",
"versionType": "Radiometer Customized Windows Operating System"
}
]
},
{
"defaultStatus": "unaffected",
"platforms": [
"Windows 7",
"Windows XP"
],
"product": "AQT90 FLEX Analyzers",
"vendor": "Radiometer Medical Aps",
"versions": [
{
"status": "affected",
"version": "Windows 7 Operating system",
"versionType": "Radiometer Customized Windows Operating System"
},
{
"status": "affected",
"version": "Windows XP Operating system",
"versionType": "Radiometer Customized Windows Operating System"
}
]
},
{
"defaultStatus": "unaffected",
"platforms": [
"Windows 7",
"Windows XP"
],
"product": "ABL800 BASIC and ABL800 FLEX analyzers",
"vendor": "Radiometer Medical Aps",
"versions": [
{
"status": "affected",
"version": "Windows 7 Operating system",
"versionType": "Radiometer Customized Windows Operating System"
},
{
"status": "affected",
"version": "Windows XP Operating system",
"versionType": "Radiometer Customized Windows Operating System"
}
]
}
],
"source": "46b595e9-1acc-41cb-9398-adaf98d37a9b"
}
],
"cveTags": [],
"descriptions": [
{
"lang": "en",
"value": "A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.\n\nOther related CVE\u0027s are CVE-2025-14095 \u0026 CVE-2025-14097.\n\n\n\nAffected customers have been informed about this vulnerability. This CVE is being published to provide transparency.\n\n\n\n\n\nRequired Configuration for Exposure:\n\n\nAttacker requires physical access to the analyzer.\n\n\n\nTemporary work Around:\nOnly authorized people can physically access the analyzer.\n\nPermanent solution:\nLocal Radiometer representatives will contact all affected customers to discuss a permanent solution.\n\nExploit Status:\n\n\nResearchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication."
}
],
"id": "CVE-2025-14096",
"lastModified": "2026-06-17T08:35:19.237",
"metrics": {
"cvssMetricV31": [
{
"cvssData": {
"attackComplexity": "LOW",
"attackVector": "LOCAL",
"availabilityImpact": "HIGH",
"baseScore": 8.4,
"baseSeverity": "HIGH",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
"version": "3.1"
},
"exploitabilityScore": 2.5,
"impactScore": 5.9,
"source": "46b595e9-1acc-41cb-9398-adaf98d37a9b",
"type": "Secondary"
}
],
"ssvcV203": [
{
"source": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"ssvcData": {
"id": "CVE-2025-14096",
"options": [
{
"exploitation": "none"
},
{
"automatable": "no"
},
{
"technicalImpact": "total"
}
],
"role": "CISA Coordinator",
"timestamp": "2025-12-17T14:42:57.843796Z",
"version": "2.0.3"
}
}
]
},
"published": "2025-12-17T13:15:57.890",
"references": [
{
"source": "46b595e9-1acc-41cb-9398-adaf98d37a9b",
"url": "https://www.radiometer.com/myradiometer"
}
],
"sourceIdentifier": "46b595e9-1acc-41cb-9398-adaf98d37a9b",
"vulnStatus": "Deferred",
"weaknesses": [
{
"description": [
{
"lang": "en",
"value": "CWE-250"
},
{
"lang": "en",
"value": "CWE-798"
}
],
"source": "46b595e9-1acc-41cb-9398-adaf98d37a9b",
"type": "Secondary"
}
]
}
},
"redhat_vex": {
"current_release_date": "2025-12-18T12:39:55+00:00",
"cve": "CVE-2025-14096",
"id": "CVE-2025-14096",
"initial_release_date": "2025-01-01T00:00:00+00:00",
"product_status:known_not_affected": "1",
"source": "Red Hat CSAF VEX",
"status": "final",
"title": "Credential Disclosure vulnerability in Radiometer Products",
"url": "https://security.access.redhat.com/data/csaf/v2/vex/2025/cve-2025-14096.json",
"version": "3"
},
"vulnrichment": {
"containers": {
"adp": [
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2025-14096",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "total"
}
],
"role": "CISA Coordinator",
"timestamp": "2025-12-17T14:42:57.843796Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2025-12-17T14:43:12.844Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "unaffected",
"platforms": [
"Windows XP",
"Windows 7"
],
"product": "ABL90 FLEX and ABL90 FLEX PLUS Analyzers",
"vendor": "Radiometer Medical Aps",
"versions": [
{
"status": "affected",
"version": "Windows 7 Operating system",
"versionType": "Radiometer Customized Windows Operating System"
},
{
"status": "affected",
"version": "Windows XP Operating system",
"versionType": "Radiometer Customized Windows Operating System"
}
]
},
{
"defaultStatus": "unaffected",
"platforms": [
"Windows 7",
"Windows XP"
],
"product": "AQT90 FLEX Analyzers",
"vendor": "Radiometer Medical Aps",
"versions": [
{
"status": "affected",
"version": "Windows 7 Operating system",
"versionType": "Radiometer Customized Windows Operating System"
},
{
"status": "affected",
"version": "Windows XP Operating system",
"versionType": "Radiometer Customized Windows Operating System"
}
]
},
{
"defaultStatus": "unaffected",
"platforms": [
"Windows 7",
"Windows XP"
],
"product": "ABL800 BASIC and ABL800 FLEX analyzers",
"vendor": "Radiometer Medical Aps",
"versions": [
{
"status": "affected",
"version": "Windows 7 Operating system",
"versionType": "Radiometer Customized Windows Operating System"
},
{
"status": "affected",
"version": "Windows XP Operating system",
"versionType": "Radiometer Customized Windows Operating System"
}
]
}
],
"configurations": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "Requires physical access to the Analyzer.\n\n\u003cbr\u003e"
}
],
"value": "Requires physical access to the Analyzer."
}
],
"credits": [
{
"lang": "en",
"type": "reporter",
"value": "Florian Hauser and Fabian Weber from CODE WHITE GmbH"
}
],
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.\u003cbr\u003e\u003cbr\u003eOther related CVE\u0027s are CVE-2025-14095 \u0026amp; CVE-2025-14097.\u003cbr\u003e\u003cbr\u003e\n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eAffected customers have been informed about this vulnerability. This CVE is being published to provide transparency.\u003c/span\u003e\n\n\u003cbr\u003e\u003cbr\u003e\n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eRequired Configuration for Exposure:\u003cbr\u003e\u003c/span\u003e\n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eAttacker requires physical access to the analyzer.\u003c/span\u003e\n\n\u003cbr\u003e\u003cbr\u003eTemporary work Around:\u003cbr\u003eOnly authorized people can physically access the analyzer.\u003cbr\u003e\u003cbr\u003ePermanent solution:\u003cbr\u003eLocal Radiometer representatives will contact all affected customers to discuss a permanent solution.\u003cbr\u003e\u003cbr\u003eExploit Status:\u003cbr\u003e\n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eResearchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.\u003c/span\u003e\n\n\u003cbr\u003e\u003cbr\u003e"
}
],
"value": "A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.\n\nOther related CVE\u0027s are CVE-2025-14095 \u0026 CVE-2025-14097.\n\n\n\nAffected customers have been informed about this vulnerability. This CVE is being published to provide transparency.\n\n\n\n\n\nRequired Configuration for Exposure:\n\n\nAttacker requires physical access to the analyzer.\n\n\n\nTemporary work Around:\nOnly authorized people can physically access the analyzer.\n\nPermanent solution:\nLocal Radiometer representatives will contact all affected customers to discuss a permanent solution.\n\nExploit Status:\n\n\nResearchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication."
}
],
"exploits": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.\n\n\u003cbr\u003e"
}
],
"value": "Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication."
}
],
"impacts": [
{
"capecId": "CAPEC-69",
"descriptions": [
{
"lang": "en",
"value": "CAPEC-69 \u2014 Target Programs with Elevated Privileges"
}
]
}
],
"metrics": [
{
"cvssV3_1": {
"attackComplexity": "LOW",
"attackVector": "LOCAL",
"availabilityImpact": "HIGH",
"baseScore": 8.4,
"baseSeverity": "HIGH",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
"version": "3.1"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-798",
"description": "CWE-798: Use of Hard-coded Credentials",
"lang": "en",
"type": "CWE"
}
]
},
{
"descriptions": [
{
"cweId": "CWE-250",
"description": "CWE-250: Execution with Unnecessary Privileges",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2025-12-17T13:03:05.696Z",
"orgId": "46b595e9-1acc-41cb-9398-adaf98d37a9b",
"shortName": "Radiometer"
},
"references": [
{
"url": "https://www.radiometer.com/myradiometer"
}
],
"solutions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "Local Radiometer representatives will contact all affected customers to discuss a permanent solution.\n\n\n\n\u003cbr\u003e"
}
],
"value": "Local Radiometer representatives will contact all affected customers to discuss a permanent solution."
}
],
"source": {
"discovery": "EXTERNAL"
},
"timeline": [
{
"lang": "en",
"time": "2025-12-17T13:01:00.000Z",
"value": "CVE Published"
}
],
"title": "Credential Disclosure vulnerability in Radiometer Products",
"workarounds": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "Only authorized people can physically access the analyzer.\n\n\n\n\u003cbr\u003e"
}
],
"value": "Only authorized people can physically access the analyzer."
}
],
"x_generator": {
"engine": "Vulnogram 0.5.0"
}
}
},
"cveMetadata": {
"assignerOrgId": "46b595e9-1acc-41cb-9398-adaf98d37a9b",
"assignerShortName": "Radiometer",
"cveId": "CVE-2025-14096",
"datePublished": "2025-12-17T12:19:27.489Z",
"dateReserved": "2025-12-05T10:50:00.566Z",
"dateUpdated": "2025-12-17T14:43:16.523Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.2"
}
}
}
Loading…
Loading…
Experimental. This forecast is provided for visualization only and may change without notice. Do not use it for operational decisions.
Forecast uses a logistic model when the trend is rising, or an exponential decay model when the trend is falling. Fitted via linearized least squares.
Sightings
| Author | Source | Type | Date | Other |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or observed by the user.
- Confirmed: The vulnerability has been validated from an analyst's perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
- Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
- Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
- Not confirmed: The user expressed doubt about the validity of the vulnerability.
- Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.
Loading…
The MITRE ATT&CK techniques below are AI-generated suggestions, inferred from the description of the
vulnerability by the CIRCL/vulnerability-attack-technique-classification-roberta-base
model, served locally by ML-Gateway.
They have not been verified by an analyst and are provided for guidance only.
The approach is described in our paper Mapping CVEs to MITRE ATT&CK Techniques: A Curated Gold-Set Classifier and the Limits of LLM-Assisted Label Expansion.
Browse all ATT&CK techniques and the vulnerabilities related to each.
The approach is described in our paper Mapping CVEs to MITRE ATT&CK Techniques: A Curated Gold-Set Classifier and the Limits of LLM-Assisted Label Expansion.
Browse all ATT&CK techniques and the vulnerabilities related to each.
Loading…
Related by attack behaviour
Vulnerabilities whose description is nearest to this one in the vector space of the CIRCL/vulnerability-attack-technique-biencoder model. This is a similarity search over the bi-encoder space (plain cosine), not a classification, and it has no measured accuracy.
Loading…