cve-2024-9834
Vulnerability from cvelistv5
Published
2024-11-14 20:57
Modified
2024-11-15 15:10
Severity ?
Summary
Improper data protection on Life2000 ventilator serial interface
Impacted products
Vendor Product Version
Show details on NVD website


{
  "containers": {
    "adp": [
      {
        "affected": [
          {
            "cpes": [
              "cpe:2.3:o:baxter:life2000_ventilator_firmware:*:*:*:*:*:*:*:*"
            ],
            "defaultStatus": "unaffected",
            "product": "life2000_ventilator_firmware",
            "vendor": "baxter",
            "versions": [
              {
                "lessThanOrEqual": "06.08.00.00",
                "status": "affected",
                "version": "0",
                "versionType": "custom"
              }
            ]
          }
        ],
        "metrics": [
          {
            "other": {
              "content": {
                "id": "CVE-2024-9834",
                "options": [
                  {
                    "Exploitation": "none"
                  },
                  {
                    "Automatable": "no"
                  },
                  {
                    "Technical Impact": "total"
                  }
                ],
                "role": "CISA Coordinator",
                "timestamp": "2024-11-15T15:09:40.431805Z",
                "version": "2.0.3"
              },
              "type": "ssvc"
            }
          }
        ],
        "providerMetadata": {
          "dateUpdated": "2024-11-15T15:10:40.157Z",
          "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
          "shortName": "CISA-ADP"
        },
        "title": "CISA ADP Vulnrichment"
      }
    ],
    "cna": {
      "affected": [
        {
          "defaultStatus": "unaffected",
          "product": "Life2000 Ventilation System",
          "vendor": "Baxter",
          "versions": [
            {
              "status": "affected",
              "version": "06.08.00.00 and prior"
            }
          ]
        }
      ],
      "datePublic": "2024-11-14T20:43:00.000Z",
      "descriptions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "Improper data protection on the ventilator\u0027s serial interface could allow an attacker to send and receive messages that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.\u003cbr\u003e"
            }
          ],
          "value": "Improper data protection on the ventilator\u0027s serial interface could allow an attacker to send and receive messages that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance."
        }
      ],
      "impacts": [
        {
          "capecId": "CAPEC-117",
          "descriptions": [
            {
              "lang": "en",
              "value": "CAPEC-117 Interception"
            }
          ]
        },
        {
          "capecId": "CAPEC-441",
          "descriptions": [
            {
              "lang": "en",
              "value": "CAPEC-441 Malicious Logic Insertion"
            }
          ]
        }
      ],
      "metrics": [
        {
          "cvssV3_1": {
            "attackComplexity": "LOW",
            "attackVector": "LOCAL",
            "availabilityImpact": "HIGH",
            "baseScore": 9.3,
            "baseSeverity": "CRITICAL",
            "confidentialityImpact": "HIGH",
            "integrityImpact": "HIGH",
            "privilegesRequired": "NONE",
            "scope": "CHANGED",
            "userInteraction": "NONE",
            "vectorString": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H",
            "version": "3.1"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-319",
              "description": "CWE-319 Cleartext Transmission of Sensitive Information",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2024-11-14T21:47:25.133Z",
        "orgId": "dba971b9-eb30-4121-91e1-3b45611354aa",
        "shortName": "Baxter"
      },
      "references": [
        {
          "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01"
        }
      ],
      "source": {
        "discovery": "UNKNOWN"
      },
      "title": "Improper data protection on Life2000 ventilator serial interface",
      "x_generator": {
        "engine": "Vulnogram 0.2.0"
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "dba971b9-eb30-4121-91e1-3b45611354aa",
    "assignerShortName": "Baxter",
    "cveId": "CVE-2024-9834",
    "datePublished": "2024-11-14T20:57:22.734Z",
    "dateReserved": "2024-10-10T19:24:48.834Z",
    "dateUpdated": "2024-11-15T15:10:40.157Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",
  "dataVersion": "5.1",
  "meta": {
    "nvd": "{\"cve\":{\"id\":\"CVE-2024-9834\",\"sourceIdentifier\":\"productsecurity@baxter.com\",\"published\":\"2024-11-14T21:15:22.803\",\"lastModified\":\"2024-11-15T13:58:08.913\",\"vulnStatus\":\"Awaiting Analysis\",\"cveTags\":[],\"descriptions\":[{\"lang\":\"en\",\"value\":\"Improper data protection on the ventilator\u0027s serial interface could allow an attacker to send and receive messages that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.\"},{\"lang\":\"es\",\"value\":\"La protecci\u00f3n de datos inadecuada en la interfaz serial del ventilador podr\u00eda permitir que un atacante env\u00ede y reciba mensajes que resulten en una divulgaci\u00f3n no autorizada de informaci\u00f3n y/o tengan impactos no deseados en la configuraci\u00f3n y el rendimiento del dispositivo.\"}],\"metrics\":{\"cvssMetricV31\":[{\"source\":\"productsecurity@baxter.com\",\"type\":\"Secondary\",\"cvssData\":{\"version\":\"3.1\",\"vectorString\":\"CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H\",\"baseScore\":9.3,\"baseSeverity\":\"CRITICAL\",\"attackVector\":\"LOCAL\",\"attackComplexity\":\"LOW\",\"privilegesRequired\":\"NONE\",\"userInteraction\":\"NONE\",\"scope\":\"CHANGED\",\"confidentialityImpact\":\"HIGH\",\"integrityImpact\":\"HIGH\",\"availabilityImpact\":\"HIGH\"},\"exploitabilityScore\":2.5,\"impactScore\":6.0}]},\"weaknesses\":[{\"source\":\"productsecurity@baxter.com\",\"type\":\"Secondary\",\"description\":[{\"lang\":\"en\",\"value\":\"CWE-319\"}]}],\"references\":[{\"url\":\"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01\",\"source\":\"productsecurity@baxter.com\"}]}}"
  }
}


Log in or create an account to share your comment.




Tags
Taxonomy of the tags.


Loading…

Loading…

Loading…

Sightings

Author Source Type Date

Nomenclature

  • Seen: The vulnerability was mentioned, discussed, or seen somewhere by the user.
  • Confirmed: The vulnerability is confirmed from an analyst perspective.
  • Exploited: This vulnerability was exploited and seen by the user reporting the sighting.
  • Patched: This vulnerability was successfully patched by the user reporting the sighting.
  • Not exploited: This vulnerability was not exploited or seen by the user reporting the sighting.
  • Not confirmed: The user expresses doubt about the veracity of the vulnerability.
  • Not patched: This vulnerability was not successfully patched by the user reporting the sighting.