CVE-2015-3953 (GCVE-0-2015-3953)
Vulnerability from cvelistv5
Published
2019-03-25 16:02
Modified
2024-08-06 06:04
Severity ?
VLAI Severity ?
EPSS score ?
CWE
- CWE-259 - Use of hard-coded password
Summary
Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
References
| URL | Tags | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
|
|||||||||||
Impacted products
| Vendor | Product | Version | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Hospira | Plum A+ Infusion System |
Version: <= 13.4 |
||||||||||||
|
||||||||||||||
{
"containers": {
"adp": [
{
"providerMetadata": {
"dateUpdated": "2024-08-06T06:04:00.957Z",
"orgId": "af854a3a-2127-422b-91ae-364da2661108",
"shortName": "CVE"
},
"references": [
{
"tags": [
"x_refsource_MISC",
"x_transferred"
],
"url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01"
}
],
"title": "CVE Program Container"
}
],
"cna": {
"affected": [
{
"product": "Plum A+ Infusion System",
"vendor": "Hospira",
"versions": [
{
"status": "affected",
"version": "\u003c= 13.4"
}
]
},
{
"product": "Plum A+3 Infusion System",
"vendor": "Hospira",
"versions": [
{
"status": "affected",
"version": "\u003c= 13.6"
}
]
},
{
"product": "Symbiq Infusion System",
"vendor": "Hospira",
"versions": [
{
"status": "affected",
"version": "\u003c= 3.13"
}
]
}
],
"datePublic": "2015-06-10T00:00:00",
"descriptions": [
{
"lang": "en",
"value": "Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue."
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-259",
"description": "Use of hard-coded password CWE-259",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2019-03-25T16:02:25",
"orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"shortName": "icscert"
},
"references": [
{
"tags": [
"x_refsource_MISC"
],
"url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01"
}
],
"x_legacyV4Record": {
"CVE_data_meta": {
"ASSIGNER": "ics-cert@hq.dhs.gov",
"ID": "CVE-2015-3953",
"STATE": "PUBLIC"
},
"affects": {
"vendor": {
"vendor_data": [
{
"product": {
"product_data": [
{
"product_name": "Plum A+ Infusion System",
"version": {
"version_data": [
{
"version_value": "\u003c= 13.4"
}
]
}
},
{
"product_name": "Plum A+3 Infusion System",
"version": {
"version_data": [
{
"version_value": "\u003c= 13.6"
}
]
}
},
{
"product_name": "Symbiq Infusion System",
"version": {
"version_data": [
{
"version_value": "\u003c= 3.13"
}
]
}
}
]
},
"vendor_name": "Hospira"
}
]
}
},
"data_format": "MITRE",
"data_type": "CVE",
"data_version": "4.0",
"description": {
"description_data": [
{
"lang": "eng",
"value": "Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue."
}
]
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "Use of hard-coded password CWE-259"
}
]
}
]
},
"references": {
"reference_data": [
{
"name": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01",
"refsource": "MISC",
"url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01"
}
]
}
}
}
},
"cveMetadata": {
"assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"assignerShortName": "icscert",
"cveId": "CVE-2015-3953",
"datePublished": "2019-03-25T16:02:25",
"dateReserved": "2015-05-12T00:00:00",
"dateUpdated": "2024-08-06T06:04:00.957Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1",
"vulnerability-lookup:meta": {
"nvd": "{\"cve\":{\"id\":\"CVE-2015-3953\",\"sourceIdentifier\":\"ics-cert@hq.dhs.gov\",\"published\":\"2019-03-25T17:29:00.623\",\"lastModified\":\"2024-11-21T02:30:08.053\",\"vulnStatus\":\"Modified\",\"cveTags\":[],\"descriptions\":[{\"lang\":\"en\",\"value\":\"Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.\"},{\"lang\":\"es\",\"value\":\"Las cuentas embebidas podr\u00edan ser usadas para acceder a Hospira Plum A+ Infusion System versi\u00f3n 13.4 y anteriores, Plum A+3 Infusion System versi\u00f3n 13.6 y anteriores y Symbiq Infusion System, versi\u00f3n 3.13 y anteriores. Hospira recomienda que los usuarios cierren el Puerto 20/FTP y el Puerto 23/TELNET en los dispositivos afectados. Hospira ha lanzado tambi\u00e9n el Plum 360 Infusion System que no es vulnerable a este problema.\"}],\"metrics\":{\"cvssMetricV30\":[{\"source\":\"nvd@nist.gov\",\"type\":\"Primary\",\"cvssData\":{\"version\":\"3.0\",\"vectorString\":\"CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H\",\"baseScore\":9.8,\"baseSeverity\":\"CRITICAL\",\"attackVector\":\"NETWORK\",\"attackComplexity\":\"LOW\",\"privilegesRequired\":\"NONE\",\"userInteraction\":\"NONE\",\"scope\":\"UNCHANGED\",\"confidentialityImpact\":\"HIGH\",\"integrityImpact\":\"HIGH\",\"availabilityImpact\":\"HIGH\"},\"exploitabilityScore\":3.9,\"impactScore\":5.9}],\"cvssMetricV2\":[{\"source\":\"nvd@nist.gov\",\"type\":\"Primary\",\"cvssData\":{\"version\":\"2.0\",\"vectorString\":\"AV:N/AC:L/Au:N/C:C/I:C/A:C\",\"baseScore\":10.0,\"accessVector\":\"NETWORK\",\"accessComplexity\":\"LOW\",\"authentication\":\"NONE\",\"confidentialityImpact\":\"COMPLETE\",\"integrityImpact\":\"COMPLETE\",\"availabilityImpact\":\"COMPLETE\"},\"baseSeverity\":\"HIGH\",\"exploitabilityScore\":10.0,\"impactScore\":10.0,\"acInsufInfo\":false,\"obtainAllPrivilege\":false,\"obtainUserPrivilege\":false,\"obtainOtherPrivilege\":false,\"userInteractionRequired\":false}]},\"weaknesses\":[{\"source\":\"ics-cert@hq.dhs.gov\",\"type\":\"Secondary\",\"description\":[{\"lang\":\"en\",\"value\":\"CWE-259\"}]},{\"source\":\"nvd@nist.gov\",\"type\":\"Primary\",\"description\":[{\"lang\":\"en\",\"value\":\"CWE-798\"}]}],\"configurations\":[{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:pifzer:plum_a\\\\+_infusion_system_firmware:*:*:*:*:*:*:*:*\",\"versionEndIncluding\":\"13.4\",\"matchCriteriaId\":\"61C1A9DD-F143-4D0C-871C-B6CD7AF9DAB2\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:pifzer:plum_a\\\\+_infusion_system:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"1DB2490B-0318-4770-BF45-CD7527F15D7F\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:pifzer:plum_a\\\\+3_infusion_system_firmware:*:*:*:*:*:*:*:*\",\"versionEndIncluding\":\"13.6\",\"matchCriteriaId\":\"8536E705-89E8-47CB-9567-6AD65FBA0F1B\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:pifzer:plum_a\\\\+3_infusion_system:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"423AA561-8E38-4378-814B-1008B96F27A6\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:pifzer:symbiq_infusion_system_firmware:*:*:*:*:*:*:*:*\",\"versionEndIncluding\":\"3.13\",\"matchCriteriaId\":\"C6DD5141-72AB-4694-8595-B4BED8EC7773\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:pifzer:symbiq_infusion_system:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"5C295E1A-BF60-476D-B972-5C5C28D7633B\"}]}]}],\"references\":[{\"url\":\"https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01\",\"source\":\"ics-cert@hq.dhs.gov\",\"tags\":[\"Mitigation\",\"Third Party Advisory\",\"US Government Resource\"]},{\"url\":\"https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01\",\"source\":\"af854a3a-2127-422b-91ae-364da2661108\",\"tags\":[\"Mitigation\",\"Third Party Advisory\",\"US Government Resource\"]}]}}"
}
}
Loading…
Loading…
Sightings
| Author | Source | Type | Date |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or seen somewhere by the user.
- Confirmed: The vulnerability is confirmed from an analyst perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: This vulnerability was exploited and seen by the user reporting the sighting.
- Patched: This vulnerability was successfully patched by the user reporting the sighting.
- Not exploited: This vulnerability was not exploited or seen by the user reporting the sighting.
- Not confirmed: The user expresses doubt about the veracity of the vulnerability.
- Not patched: This vulnerability was not successfully patched by the user reporting the sighting.
Loading…
Loading…