ICSMA-21-322-02
Vulnerability from csaf_cisa - Published: 2021-11-18 00:00 - Updated: 2021-11-18 00:00Summary
Philips Patient Information Center iX (PIC iX) and Efficia CM Series
Notes
CISA Disclaimer
This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov
Legal Notice
All information products included in https://us-cert.cisa.gov/ics are provided "as is" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.
Risk evaluation
Successful exploitation of these vulnerabilities may allow an attacker unauthorized access to data (including patient data) and create a denial of service resulting in temporary interruption of viewing physiological data at the central station. Exploitation does not enable modification or change to point-of-care devices.
Critical infrastructure sectors
Healthcare and Public Health
Countries/areas deployed
Worldwide
Company headquarters location
Netherlands
Recommended Practices
CISA recommends users take defensive measures to minimize the risk of exploitation of this vulnerability. Specifically, users should: Minimize network exposure for all control system devices and/or systems, and ensure they are not accessible from the Internet; Locate control system networks and remote devices behind firewalls and isolate them from the business network; When remote access is required, use secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as its connected devices.
Recommended Practices
CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.
CISA also provides a section for control systems security recommended practices on the ICS webpage onus-cert.cisa.gov. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.
Recommended Practices
Additional mitigation guidance and recommended practices are publicly available on the ICS webpage on us-cert.cisa.gov in the Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.
Organizations observing any suspected malicious activity should follow their established internal procedures and report their findings to CISA for tracking and correlation against other incidents.
Exploitability
No known public exploits specifically target these vulnerabilities. These vulnerabilities are not exploitable remotely.
{
"document": {
"acknowledgments": [
{
"names": [
"Younes Dragoni",
"Andrea Palanca",
"Ivan Speziale"
],
"organization": "Nozomi Networks",
"summary": "reporting these vulnerabilities to CISA"
}
],
"category": "csaf_security_advisory",
"csaf_version": "2.0",
"distribution": {
"text": "Disclosure is not limited",
"tlp": {
"label": "WHITE",
"url": "https://us-cert.cisa.gov/tlp/"
}
},
"lang": "en-US",
"notes": [
{
"category": "general",
"text": "This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov",
"title": "CISA Disclaimer"
},
{
"category": "legal_disclaimer",
"text": "All information products included in https://us-cert.cisa.gov/ics are provided \"as is\" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.",
"title": "Legal Notice"
},
{
"category": "summary",
"text": "Successful exploitation of these vulnerabilities may allow an attacker unauthorized access to data (including patient data) and create a denial of service resulting in temporary interruption of viewing physiological data at the central station. Exploitation does not enable modification or change to point-of-care devices.",
"title": "Risk evaluation"
},
{
"category": "other",
"text": "Healthcare and Public Health",
"title": "Critical infrastructure sectors"
},
{
"category": "other",
"text": "Worldwide",
"title": "Countries/areas deployed"
},
{
"category": "other",
"text": "Netherlands",
"title": "Company headquarters location"
},
{
"category": "general",
"text": "CISA recommends users take defensive measures to minimize the risk of exploitation of this vulnerability. Specifically, users should: Minimize network exposure for all control system devices and/or systems, and ensure they are not accessible from the Internet; Locate control system networks and remote devices behind firewalls and isolate them from the business network; When remote access is required, use secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as its connected devices.",
"title": "Recommended Practices"
},
{
"category": "general",
"text": "CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.\nCISA also provides a section for control systems security recommended practices on the ICS webpage onus-cert.cisa.gov. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.",
"title": "Recommended Practices"
},
{
"category": "general",
"text": "Additional mitigation guidance and recommended practices are publicly available on the ICS webpage on us-cert.cisa.gov in the Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.\nOrganizations observing any suspected malicious activity should follow their established internal procedures and report their findings to CISA for tracking and correlation against other incidents.",
"title": "Recommended Practices"
},
{
"category": "other",
"text": "No known public exploits specifically target these vulnerabilities. These vulnerabilities are not exploitable remotely.",
"title": "Exploitability"
}
],
"publisher": {
"category": "coordinator",
"contact_details": "Email: CISAservicedesk@cisa.dhs.gov;\n Toll Free: 1-888-282-0870",
"name": "CISA",
"namespace": "https://www.cisa.gov/"
},
"references": [
{
"category": "self",
"summary": "ICS Advisory ICSMA-21-322-02 JSON",
"url": "https://raw.githubusercontent.com/cisagov/CSAF/develop/csaf_files/OT/white/2021/icsma-21-322-02.json"
},
{
"category": "self",
"summary": "ICS Advisory ICSMA-21-322-02 Web Version",
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-21-322-02"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://us-cert.cisa.gov/ics/alerts/ICS-ALERT-10-301-01"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://us-cert.cisa.gov/sites/default/files/recommended_practices/NCCIC_ICS-CERT_Defense_in_Depth_2016_S508C.pdf"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://us-cert.cisa.gov/ics/tips/ICS-TIP-12-146-01B"
}
],
"title": "Philips Patient Information Center iX (PIC iX) and Efficia CM Series",
"tracking": {
"current_release_date": "2021-11-18T00:00:00.000000Z",
"generator": {
"engine": {
"name": "CISA CSAF Generator",
"version": "1.0.0"
}
},
"id": "ICSMA-21-322-02",
"initial_release_date": "2021-11-18T00:00:00.000000Z",
"revision_history": [
{
"date": "2021-11-18T00:00:00.000000Z",
"legacy_version": "Initial",
"number": "1",
"summary": "ICSMA-21-322-02 Philips Patient Information Center iX (PIC iX) and Efficia CM Series"
}
],
"status": "final",
"version": "1"
}
},
"product_tree": {
"branches": [
{
"branches": [
{
"branches": [
{
"category": "product_version_range",
"name": "A.01 \u003c= C.0x | 4.0",
"product": {
"name": "Efficia CM Series: Revisions A.01 to C.0x and 4.0",
"product_id": "CSAFPID-0001"
}
}
],
"category": "product_name",
"name": "Efficia CM Series"
},
{
"branches": [
{
"category": "product_version",
"name": "B.02 | C.02 | C.03",
"product": {
"name": "Patient Information Center iX (PIC iX): Versions B.02 C.02 C.03",
"product_id": "CSAFPID-0002"
}
}
],
"category": "product_name",
"name": "Patient Information Center iX (PIC iX)"
}
],
"category": "vendor",
"name": "Philips"
}
]
},
"vulnerabilities": [
{
"cve": "CVE-2021-43548",
"cwe": {
"id": "CWE-20",
"name": "Improper Input Validation"
},
"notes": [
{
"category": "summary",
"text": "Patient Information Center iX (PIC iX) Versions C.02 and C.03 receives input or data, but does not validate or incorrectly validates that the input has the properties required to process the data safely and correctly. CVE-2021-43548 has been assigned to this vulnerability. A CVSS v3 base score of 6.5 has been calculated; the CVSS vector string is (AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2021-43548"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "Philips released a remediation for CVE-2021-43548 in Q3 2021 in PIC iX C.03.06. Philips plans to remediate CVE-2021-43552 and CVE-2021-43550 by end of Q2 of 2023. Users should operate all Philips deployed and supported products within Philips authorized specifications, including Philips approved software, software configuration, system services, and security configuration.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
{
"category": "mitigation",
"details": "As an interim mitigation, Philips recommends the following actions outlined in the Philips Patient Monitoring System Security for Clinical Networks guide at InCenter.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
],
"url": "https://incenter.medical.philips.com/"
},
{
"category": "mitigation",
"details": "Philips provided hardware ships with Bitlocker Drive Encryption enabled by default to protect the data at rest stored on the system. It should not be disabled.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
{
"category": "mitigation",
"details": "Philips recommends customers follow NIST SP 800-88 for media sanitization prior to system disposal.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
{
"category": "mitigation",
"details": "By default, patient information is not included in archives. When exporting archives that contain patient information, users should store information securely with strong access controls.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
{
"category": "mitigation",
"details": "The Philips patient monitoring network is required to be physically or logically isolated from the hospital local area network (LAN). Philips recommends using a firewall or routers that can implement access control lists restricting access in and out of the patient monitoring network for only necessary ports and IP addresses.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
{
"category": "vendor_fix",
"details": "Users with questions regarding their specific Philips Patient Information Center (PIC iX) and Efficia CM Series new release eligibility should contact a Philips service support team or regional service support. Philips contact information is available at the Philips customer service website, or by calling 1-800-722-9377.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
],
"url": "https://www.usa.philips.com/healthcare/solutions/customer-service-solutions"
},
{
"category": "vendor_fix",
"details": "Please see the Philips product security website for the latest security information for Philips products.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
],
"url": "https://www.philips.com/productsecurity"
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 6.5,
"baseSeverity": "MEDIUM",
"vectorString": "CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
"version": "3.0"
},
"products": [
"CSAFPID-0001",
"CSAFPID-0002"
]
}
]
},
{
"cve": "CVE-2021-43552",
"cwe": {
"id": "CWE-321",
"name": "Use of Hard-coded Cryptographic Key"
},
"notes": [
{
"category": "summary",
"text": "The use of a hard-coded cryptographic key significantly increases the possibility encrypted data may be recovered from the Patient Information Center iX (PIC iX) Versions B.02, C.02, and C.03. CVE-2021-43552 has been assigned to this vulnerability. A CVSS v3 base score of 6.1 has been calculated; the CVSS vector string is (AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:L/A:N).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2021-43552"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:L/A:N"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "Philips released a remediation for CVE-2021-43548 in Q3 2021 in PIC iX C.03.06. Philips plans to remediate CVE-2021-43552 and CVE-2021-43550 by end of Q2 of 2023. Users should operate all Philips deployed and supported products within Philips authorized specifications, including Philips approved software, software configuration, system services, and security configuration.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
{
"category": "mitigation",
"details": "As an interim mitigation, Philips recommends the following actions outlined in the Philips Patient Monitoring System Security for Clinical Networks guide at InCenter.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
],
"url": "https://incenter.medical.philips.com/"
},
{
"category": "mitigation",
"details": "Philips provided hardware ships with Bitlocker Drive Encryption enabled by default to protect the data at rest stored on the system. It should not be disabled.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
{
"category": "mitigation",
"details": "Philips recommends customers follow NIST SP 800-88 for media sanitization prior to system disposal.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
{
"category": "mitigation",
"details": "By default, patient information is not included in archives. When exporting archives that contain patient information, users should store information securely with strong access controls.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
{
"category": "mitigation",
"details": "The Philips patient monitoring network is required to be physically or logically isolated from the hospital local area network (LAN). Philips recommends using a firewall or routers that can implement access control lists restricting access in and out of the patient monitoring network for only necessary ports and IP addresses.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
{
"category": "vendor_fix",
"details": "Users with questions regarding their specific Philips Patient Information Center (PIC iX) and Efficia CM Series new release eligibility should contact a Philips service support team or regional service support. Philips contact information is available at the Philips customer service website, or by calling 1-800-722-9377.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
],
"url": "https://www.usa.philips.com/healthcare/solutions/customer-service-solutions"
},
{
"category": "vendor_fix",
"details": "Please see the Philips product security website for the latest security information for Philips products.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
],
"url": "https://www.philips.com/productsecurity"
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 6.1,
"baseSeverity": "MEDIUM",
"vectorString": "CVSS:3.0/AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:L/A:N",
"version": "3.0"
},
"products": [
"CSAFPID-0001",
"CSAFPID-0002"
]
}
]
},
{
"cve": "CVE-2021-43550",
"cwe": {
"id": "CWE-327",
"name": "Use of a Broken or Risky Cryptographic Algorithm"
},
"notes": [
{
"category": "summary",
"text": "The use of a broken or risky cryptographic algorithm is an unnecessary risk that may result in the exposure of sensitive information, which affects the communications between Patient Information Center iX (PIC iX) Versions C.02 and C.03 and Efficia CM Series Revisions A.01 to C.0x and 4.0. CVE-2021-43550 has been assigned to this vulnerability. A CVSS v3 base score of 5.9 has been calculated; the CVSS vector string is (AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:L/A:N).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2021-43550"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:L/A:N"
}
],
"remediations": [
{
"category": "vendor_fix",
"details": "Philips released a remediation for CVE-2021-43548 in Q3 2021 in PIC iX C.03.06. Philips plans to remediate CVE-2021-43552 and CVE-2021-43550 by end of Q2 of 2023. Users should operate all Philips deployed and supported products within Philips authorized specifications, including Philips approved software, software configuration, system services, and security configuration.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
{
"category": "mitigation",
"details": "As an interim mitigation, Philips recommends the following actions outlined in the Philips Patient Monitoring System Security for Clinical Networks guide at InCenter.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
],
"url": "https://incenter.medical.philips.com/"
},
{
"category": "mitigation",
"details": "Philips provided hardware ships with Bitlocker Drive Encryption enabled by default to protect the data at rest stored on the system. It should not be disabled.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
{
"category": "mitigation",
"details": "Philips recommends customers follow NIST SP 800-88 for media sanitization prior to system disposal.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
{
"category": "mitigation",
"details": "By default, patient information is not included in archives. When exporting archives that contain patient information, users should store information securely with strong access controls.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
{
"category": "mitigation",
"details": "The Philips patient monitoring network is required to be physically or logically isolated from the hospital local area network (LAN). Philips recommends using a firewall or routers that can implement access control lists restricting access in and out of the patient monitoring network for only necessary ports and IP addresses.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
]
},
{
"category": "vendor_fix",
"details": "Users with questions regarding their specific Philips Patient Information Center (PIC iX) and Efficia CM Series new release eligibility should contact a Philips service support team or regional service support. Philips contact information is available at the Philips customer service website, or by calling 1-800-722-9377.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
],
"url": "https://www.usa.philips.com/healthcare/solutions/customer-service-solutions"
},
{
"category": "vendor_fix",
"details": "Please see the Philips product security website for the latest security information for Philips products.",
"product_ids": [
"CSAFPID-0001",
"CSAFPID-0002"
],
"url": "https://www.philips.com/productsecurity"
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 5.9,
"baseSeverity": "MEDIUM",
"vectorString": "CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:L/A:N",
"version": "3.0"
},
"products": [
"CSAFPID-0001",
"CSAFPID-0002"
]
}
]
}
]
}
Loading…
Loading…
Sightings
| Author | Source | Type | Date |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or observed by the user.
- Confirmed: The vulnerability has been validated from an analyst's perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
- Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
- Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
- Not confirmed: The user expressed doubt about the validity of the vulnerability.
- Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.