cve-2021-38394
Vulnerability from cvelistv5
Published
2021-10-04 17:34
Modified
2024-09-16 18:59
Severity ?
EPSS score ?
Summary
An attacker with physical access to the device can extract the binary that checks for the hardware key and reverse engineer it, which could be used to create a physical duplicate of a valid hardware key. The hardware key allows access to special settings when inserted.
References
▼ | URL | Tags | |
---|---|---|---|
ics-cert@hq.dhs.gov | https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01 | Third Party Advisory, US Government Resource | |
af854a3a-2127-422b-91ae-364da2661108 | https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01 | Third Party Advisory, US Government Resource |
Impacted products
Vendor | Product | Version | |
---|---|---|---|
▼ | Boston Scientific | ZOOM LATITUDE |
Version: Model 3120 |
|
{ "containers": { "adp": [ { "providerMetadata": { "dateUpdated": "2024-08-04T01:37:16.553Z", "orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE" }, "references": [ { "tags": [ "x_refsource_MISC", "x_transferred" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01" } ], "title": "CVE Program Container" } ], "cna": { "affected": [ { "product": "ZOOM LATITUDE", "vendor": "Boston Scientific", "versions": [ { "status": "affected", "version": "Model 3120" } ] } ], "credits": [ { "lang": "en", "value": "Endres Puschner - Max Planck Institute for Security and Privacy, Bochum, Christoph Saatjohann - FH M\u00fcnster University of Applied Sciences, Christian Dresen - FH M\u00fcnster University of Applied Sciences, and Markus Willing - University of Muenster, discovered these issues as part of broader academic research of cardiac devices and reported them to Boston Scientific." } ], "datePublic": "2021-09-30T00:00:00", "descriptions": [ { "lang": "en", "value": "An attacker with physical access to the device can extract the binary that checks for the hardware key and reverse engineer it, which could be used to create a physical duplicate of a valid hardware key. The hardware key allows access to special settings when inserted." } ], "metrics": [ { "cvssV3_1": { "attackComplexity": "HIGH", "attackVector": "PHYSICAL", "availabilityImpact": "LOW", "baseScore": 6.2, "baseSeverity": "MEDIUM", "confidentialityImpact": "LOW", "integrityImpact": "HIGH", "privilegesRequired": "NONE", "scope": "CHANGED", "userInteraction": "REQUIRED", "vectorString": "CVSS:3.1/AV:P/AC:H/PR:N/UI:R/S:C/C:L/I:H/A:L", "version": "3.1" } } ], "problemTypes": [ { "descriptions": [ { "cweId": "CWE-1278", "description": "CWE-1278", "lang": "en", "type": "CWE" } ] } ], "providerMetadata": { "dateUpdated": "2021-10-04T17:34:58", "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert" }, "references": [ { "tags": [ "x_refsource_MISC" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01" } ], "source": { "advisory": "ICSMA-21-273-01", "defect": [ "CWE-1278" ], "discovery": "EXTERNAL" }, "title": "Missing Protection against Hardware Reverse Engineering Using Integrated Circuit Imaging Techniques for Boston Scientific Zoom Latitude", "workarounds": [ { "lang": "en", "value": "Boston Scientific is in the process of transitioning all users to a replacement programmer with enhanced security, the LATITUDE Programming System, Model 3300. Boston Scientific will not issue a product update to address the identified vulnerabilities in the ZOOM LATITUDE Programming System, Model 3120." } ], "x_generator": { "engine": "Vulnogram 0.0.9" }, "x_legacyV4Record": { "CVE_data_meta": { "ASSIGNER": "ics-cert@hq.dhs.gov", "DATE_PUBLIC": "2021-09-30T21:02:00.000Z", "ID": "CVE-2021-38394", "STATE": "PUBLIC", "TITLE": "Missing Protection against Hardware Reverse Engineering Using Integrated Circuit Imaging Techniques for Boston Scientific Zoom Latitude" }, "affects": { "vendor": { "vendor_data": [ { "product": { "product_data": [ { "product_name": "ZOOM LATITUDE", "version": { "version_data": [ { "version_value": "Model 3120" } ] } } ] }, "vendor_name": "Boston Scientific" } ] } }, "credit": [ { "lang": "eng", "value": "Endres Puschner - Max Planck Institute for Security and Privacy, Bochum, Christoph Saatjohann - FH M\u00fcnster University of Applied Sciences, Christian Dresen - FH M\u00fcnster University of Applied Sciences, and Markus Willing - University of Muenster, discovered these issues as part of broader academic research of cardiac devices and reported them to Boston Scientific." } ], "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": { "description_data": [ { "lang": "eng", "value": "An attacker with physical access to the device can extract the binary that checks for the hardware key and reverse engineer it, which could be used to create a physical duplicate of a valid hardware key. The hardware key allows access to special settings when inserted." } ] }, "generator": { "engine": "Vulnogram 0.0.9" }, "impact": { "cvss": { "attackComplexity": "HIGH", "attackVector": "PHYSICAL", "availabilityImpact": "LOW", "baseScore": 6.2, "baseSeverity": "MEDIUM", "confidentialityImpact": "LOW", "integrityImpact": "HIGH", "privilegesRequired": "NONE", "scope": "CHANGED", "userInteraction": "REQUIRED", "vectorString": "CVSS:3.1/AV:P/AC:H/PR:N/UI:R/S:C/C:L/I:H/A:L", "version": "3.1" } }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "eng", "value": "CWE-1278" } ] } ] }, "references": { "reference_data": [ { "name": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01", "refsource": "MISC", "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01" } ] }, "source": { "advisory": "ICSMA-21-273-01", "defect": [ "CWE-1278" ], "discovery": "EXTERNAL" }, "work_around": [ { "lang": "en", "value": "Boston Scientific is in the process of transitioning all users to a replacement programmer with enhanced security, the LATITUDE Programming System, Model 3300. Boston Scientific will not issue a product update to address the identified vulnerabilities in the ZOOM LATITUDE Programming System, Model 3120." } ] } } }, "cveMetadata": { "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2021-38394", "datePublished": "2021-10-04T17:34:58.296378Z", "dateReserved": "2021-08-10T00:00:00", "dateUpdated": "2024-09-16T18:59:19.736Z", "state": "PUBLISHED" }, "dataType": "CVE_RECORD", "dataVersion": "5.1", "vulnerability-lookup:meta": { "nvd": "{\"cve\":{\"id\":\"CVE-2021-38394\",\"sourceIdentifier\":\"ics-cert@hq.dhs.gov\",\"published\":\"2021-10-04T18:15:09.167\",\"lastModified\":\"2024-11-21T06:16:59.353\",\"vulnStatus\":\"Modified\",\"cveTags\":[],\"descriptions\":[{\"lang\":\"en\",\"value\":\"An attacker with physical access to the device can extract the binary that checks for the hardware key and reverse engineer it, which could be used to create a physical duplicate of a valid hardware key. The hardware key allows access to special settings when inserted.\"},{\"lang\":\"es\",\"value\":\"Un atacante con acceso f\u00edsico al dispositivo puede extraer el binario que comprueba la clave de hardware y realizar ingenier\u00eda inversa, que podr\u00eda ser usado para crear un duplicado f\u00edsico de una clave de hardware v\u00e1lida. La llave de hardware permite acceder a configuraciones especiales cuando es insertada\"}],\"metrics\":{\"cvssMetricV31\":[{\"source\":\"ics-cert@hq.dhs.gov\",\"type\":\"Secondary\",\"cvssData\":{\"version\":\"3.1\",\"vectorString\":\"CVSS:3.1/AV:P/AC:H/PR:N/UI:R/S:C/C:L/I:H/A:L\",\"baseScore\":6.2,\"baseSeverity\":\"MEDIUM\",\"attackVector\":\"PHYSICAL\",\"attackComplexity\":\"HIGH\",\"privilegesRequired\":\"NONE\",\"userInteraction\":\"REQUIRED\",\"scope\":\"CHANGED\",\"confidentialityImpact\":\"LOW\",\"integrityImpact\":\"HIGH\",\"availabilityImpact\":\"LOW\"},\"exploitabilityScore\":0.4,\"impactScore\":5.3},{\"source\":\"nvd@nist.gov\",\"type\":\"Primary\",\"cvssData\":{\"version\":\"3.1\",\"vectorString\":\"CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H\",\"baseScore\":6.4,\"baseSeverity\":\"MEDIUM\",\"attackVector\":\"PHYSICAL\",\"attackComplexity\":\"HIGH\",\"privilegesRequired\":\"NONE\",\"userInteraction\":\"NONE\",\"scope\":\"UNCHANGED\",\"confidentialityImpact\":\"HIGH\",\"integrityImpact\":\"HIGH\",\"availabilityImpact\":\"HIGH\"},\"exploitabilityScore\":0.5,\"impactScore\":5.9}],\"cvssMetricV2\":[{\"source\":\"nvd@nist.gov\",\"type\":\"Primary\",\"cvssData\":{\"version\":\"2.0\",\"vectorString\":\"AV:L/AC:M/Au:N/C:C/I:C/A:C\",\"baseScore\":6.9,\"accessVector\":\"LOCAL\",\"accessComplexity\":\"MEDIUM\",\"authentication\":\"NONE\",\"confidentialityImpact\":\"COMPLETE\",\"integrityImpact\":\"COMPLETE\",\"availabilityImpact\":\"COMPLETE\"},\"baseSeverity\":\"MEDIUM\",\"exploitabilityScore\":3.4,\"impactScore\":10.0,\"acInsufInfo\":false,\"obtainAllPrivilege\":false,\"obtainUserPrivilege\":false,\"obtainOtherPrivilege\":false,\"userInteractionRequired\":false}]},\"weaknesses\":[{\"source\":\"ics-cert@hq.dhs.gov\",\"type\":\"Secondary\",\"description\":[{\"lang\":\"en\",\"value\":\"CWE-1278\"}]}],\"configurations\":[{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:bostonscientific:zoom_latitude_pogrammer\\\\/recorder\\\\/monitor_3120_firmware:*:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"F9D9AC21-FFE0-4A10-A236-04957BFCE2A3\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:bostonscientific:zoom_latitude_pogrammer\\\\/recorder\\\\/monitor_3120:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"3CA3DB8D-C7B6-4276-BD8B-16C2EFDB5474\"}]}]}],\"references\":[{\"url\":\"https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01\",\"source\":\"ics-cert@hq.dhs.gov\",\"tags\":[\"Third Party Advisory\",\"US Government Resource\"]},{\"url\":\"https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01\",\"source\":\"af854a3a-2127-422b-91ae-364da2661108\",\"tags\":[\"Third Party Advisory\",\"US Government Resource\"]}]}}" } }
Loading…
Loading…
Sightings
Author | Source | Type | Date |
---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or seen somewhere by the user.
- Confirmed: The vulnerability is confirmed from an analyst perspective.
- Exploited: This vulnerability was exploited and seen by the user reporting the sighting.
- Patched: This vulnerability was successfully patched by the user reporting the sighting.
- Not exploited: This vulnerability was not exploited or seen by the user reporting the sighting.
- Not confirmed: The user expresses doubt about the veracity of the vulnerability.
- Not patched: This vulnerability was not successfully patched by the user reporting the sighting.