cve-2021-38392
Vulnerability from cvelistv5
Published
2021-10-04 17:35
Modified
2024-09-17 03:33
Summary
A skilled attacker with physical access to the affected device can gain access to the hard disk drive of the device to change the telemetry region and could use this setting to interrogate or program an implantable device in any region in the world.
References
Impacted products
Vendor Product Version
Show details on NVD website


{
  "containers": {
    "adp": [
      {
        "providerMetadata": {
          "dateUpdated": "2024-08-04T01:37:16.580Z",
          "orgId": "af854a3a-2127-422b-91ae-364da2661108",
          "shortName": "CVE"
        },
        "references": [
          {
            "tags": [
              "x_refsource_MISC",
              "x_transferred"
            ],
            "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01"
          }
        ],
        "title": "CVE Program Container"
      }
    ],
    "cna": {
      "affected": [
        {
          "product": "ZOOM LATITUDE",
          "vendor": "Boston Scientific",
          "versions": [
            {
              "status": "affected",
              "version": "Model 3120"
            }
          ]
        }
      ],
      "credits": [
        {
          "lang": "en",
          "value": "Endres Puschner - Max Planck Institute for Security and Privacy, Bochum, Christoph Saatjohann - FH M\u00fcnster  University of Applied Sciences, Christian Dresen - FH M\u00fcnster University of Applied Sciences, and Markus  Willing - University of Muenster, discovered these issues as part of broader academic research of cardiac devices  and reported them to Boston Scientific."
        }
      ],
      "datePublic": "2021-09-30T00:00:00",
      "descriptions": [
        {
          "lang": "en",
          "value": "A skilled attacker with physical access to the affected device can gain access to the hard disk drive of the device to change the telemetry region and could use this setting to interrogate or program an implantable device in any region in the world."
        }
      ],
      "metrics": [
        {
          "cvssV3_1": {
            "attackComplexity": "LOW",
            "attackVector": "PHYSICAL",
            "availabilityImpact": "LOW",
            "baseScore": 6.5,
            "baseSeverity": "MEDIUM",
            "confidentialityImpact": "LOW",
            "integrityImpact": "HIGH",
            "privilegesRequired": "NONE",
            "scope": "CHANGED",
            "userInteraction": "REQUIRED",
            "vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:R/S:C/C:L/I:H/A:L",
            "version": "3.1"
          }
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-284",
              "description": "CWE-284 Improper Access Control",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2021-10-04T17:35:13",
        "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "shortName": "icscert"
      },
      "references": [
        {
          "tags": [
            "x_refsource_MISC"
          ],
          "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01"
        }
      ],
      "source": {
        "advisory": "ICSMA-21-273-01",
        "defect": [
          "CWE-284"
        ],
        "discovery": "EXTERNAL"
      },
      "title": "Improper Access Control for Boston Scientific Zoom Latitude",
      "workarounds": [
        {
          "lang": "en",
          "value": "Boston Scientific is in the process of transitioning all users to a replacement programmer with enhanced security,  the LATITUDE Programming System, Model 3300. Boston Scientific will not issue a product update to address the identified vulnerabilities in the ZOOM LATITUDE Programming System, Model 3120."
        }
      ],
      "x_generator": {
        "engine": "Vulnogram 0.0.9"
      },
      "x_legacyV4Record": {
        "CVE_data_meta": {
          "ASSIGNER": "ics-cert@hq.dhs.gov",
          "DATE_PUBLIC": "2021-09-30T21:02:00.000Z",
          "ID": "CVE-2021-38392",
          "STATE": "PUBLIC",
          "TITLE": "Improper Access Control for Boston Scientific Zoom Latitude"
        },
        "affects": {
          "vendor": {
            "vendor_data": [
              {
                "product": {
                  "product_data": [
                    {
                      "product_name": "ZOOM LATITUDE",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "Model 3120"
                          }
                        ]
                      }
                    }
                  ]
                },
                "vendor_name": "Boston Scientific"
              }
            ]
          }
        },
        "credit": [
          {
            "lang": "eng",
            "value": "Endres Puschner - Max Planck Institute for Security and Privacy, Bochum, Christoph Saatjohann - FH M\u00fcnster  University of Applied Sciences, Christian Dresen - FH M\u00fcnster University of Applied Sciences, and Markus  Willing - University of Muenster, discovered these issues as part of broader academic research of cardiac devices  and reported them to Boston Scientific."
          }
        ],
        "data_format": "MITRE",
        "data_type": "CVE",
        "data_version": "4.0",
        "description": {
          "description_data": [
            {
              "lang": "eng",
              "value": "A skilled attacker with physical access to the affected device can gain access to the hard disk drive of the device to change the telemetry region and could use this setting to interrogate or program an implantable device in any region in the world."
            }
          ]
        },
        "generator": {
          "engine": "Vulnogram 0.0.9"
        },
        "impact": {
          "cvss": {
            "attackComplexity": "LOW",
            "attackVector": "PHYSICAL",
            "availabilityImpact": "LOW",
            "baseScore": 6.5,
            "baseSeverity": "MEDIUM",
            "confidentialityImpact": "LOW",
            "integrityImpact": "HIGH",
            "privilegesRequired": "NONE",
            "scope": "CHANGED",
            "userInteraction": "REQUIRED",
            "vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:R/S:C/C:L/I:H/A:L",
            "version": "3.1"
          }
        },
        "problemtype": {
          "problemtype_data": [
            {
              "description": [
                {
                  "lang": "eng",
                  "value": "CWE-284 Improper Access Control"
                }
              ]
            }
          ]
        },
        "references": {
          "reference_data": [
            {
              "name": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01",
              "refsource": "MISC",
              "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01"
            }
          ]
        },
        "source": {
          "advisory": "ICSMA-21-273-01",
          "defect": [
            "CWE-284"
          ],
          "discovery": "EXTERNAL"
        },
        "work_around": [
          {
            "lang": "en",
            "value": "Boston Scientific is in the process of transitioning all users to a replacement programmer with enhanced security,  the LATITUDE Programming System, Model 3300. Boston Scientific will not issue a product update to address the identified vulnerabilities in the ZOOM LATITUDE Programming System, Model 3120."
          }
        ]
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
    "assignerShortName": "icscert",
    "cveId": "CVE-2021-38392",
    "datePublished": "2021-10-04T17:35:13.879926Z",
    "dateReserved": "2021-08-10T00:00:00",
    "dateUpdated": "2024-09-17T03:33:08.540Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",
  "dataVersion": "5.1",
  "vulnerability-lookup:meta": {
    "nvd": "{\"cve\":{\"id\":\"CVE-2021-38392\",\"sourceIdentifier\":\"ics-cert@hq.dhs.gov\",\"published\":\"2021-10-04T18:15:09.110\",\"lastModified\":\"2024-11-21T06:16:59.043\",\"vulnStatus\":\"Modified\",\"cveTags\":[],\"descriptions\":[{\"lang\":\"en\",\"value\":\"A skilled attacker with physical access to the affected device can gain access to the hard disk drive of the device to change the telemetry region and could use this setting to interrogate or program an implantable device in any region in the world.\"},{\"lang\":\"es\",\"value\":\"Un atacante h\u00e1bil con acceso f\u00edsico al dispositivo afectado puede acceder a la unidad de disco duro del dispositivo para cambiar la regi\u00f3n de telemetr\u00eda y podr\u00eda usar esta configuraci\u00f3n para interrogar o programar un dispositivo implantable en cualquier regi\u00f3n del mundo\"}],\"metrics\":{\"cvssMetricV31\":[{\"source\":\"ics-cert@hq.dhs.gov\",\"type\":\"Secondary\",\"cvssData\":{\"version\":\"3.1\",\"vectorString\":\"CVSS:3.1/AV:P/AC:L/PR:N/UI:R/S:C/C:L/I:H/A:L\",\"baseScore\":6.5,\"baseSeverity\":\"MEDIUM\",\"attackVector\":\"PHYSICAL\",\"attackComplexity\":\"LOW\",\"privilegesRequired\":\"NONE\",\"userInteraction\":\"REQUIRED\",\"scope\":\"CHANGED\",\"confidentialityImpact\":\"LOW\",\"integrityImpact\":\"HIGH\",\"availabilityImpact\":\"LOW\"},\"exploitabilityScore\":0.7,\"impactScore\":5.3},{\"source\":\"nvd@nist.gov\",\"type\":\"Primary\",\"cvssData\":{\"version\":\"3.1\",\"vectorString\":\"CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H\",\"baseScore\":7.6,\"baseSeverity\":\"HIGH\",\"attackVector\":\"PHYSICAL\",\"attackComplexity\":\"LOW\",\"privilegesRequired\":\"NONE\",\"userInteraction\":\"NONE\",\"scope\":\"CHANGED\",\"confidentialityImpact\":\"HIGH\",\"integrityImpact\":\"HIGH\",\"availabilityImpact\":\"HIGH\"},\"exploitabilityScore\":0.9,\"impactScore\":6.0}],\"cvssMetricV2\":[{\"source\":\"nvd@nist.gov\",\"type\":\"Primary\",\"cvssData\":{\"version\":\"2.0\",\"vectorString\":\"AV:L/AC:L/Au:N/C:C/I:C/A:C\",\"baseScore\":7.2,\"accessVector\":\"LOCAL\",\"accessComplexity\":\"LOW\",\"authentication\":\"NONE\",\"confidentialityImpact\":\"COMPLETE\",\"integrityImpact\":\"COMPLETE\",\"availabilityImpact\":\"COMPLETE\"},\"baseSeverity\":\"HIGH\",\"exploitabilityScore\":3.9,\"impactScore\":10.0,\"acInsufInfo\":false,\"obtainAllPrivilege\":false,\"obtainUserPrivilege\":false,\"obtainOtherPrivilege\":false,\"userInteractionRequired\":false}]},\"weaknesses\":[{\"source\":\"ics-cert@hq.dhs.gov\",\"type\":\"Secondary\",\"description\":[{\"lang\":\"en\",\"value\":\"CWE-284\"}]},{\"source\":\"nvd@nist.gov\",\"type\":\"Primary\",\"description\":[{\"lang\":\"en\",\"value\":\"NVD-CWE-noinfo\"}]}],\"configurations\":[{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:bostonscientific:zoom_latitude_pogrammer\\\\/recorder\\\\/monitor_3120_firmware:*:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"F9D9AC21-FFE0-4A10-A236-04957BFCE2A3\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:bostonscientific:zoom_latitude_pogrammer\\\\/recorder\\\\/monitor_3120:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"3CA3DB8D-C7B6-4276-BD8B-16C2EFDB5474\"}]}]}],\"references\":[{\"url\":\"https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01\",\"source\":\"ics-cert@hq.dhs.gov\",\"tags\":[\"Third Party Advisory\",\"US Government Resource\"]},{\"url\":\"https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01\",\"source\":\"af854a3a-2127-422b-91ae-364da2661108\",\"tags\":[\"Third Party Advisory\",\"US Government Resource\"]}]}}"
  }
}


Log in or create an account to share your comment.




Tags
Taxonomy of the tags.


Loading…

Loading…

Loading…

Sightings

Author Source Type Date

Nomenclature

  • Seen: The vulnerability was mentioned, discussed, or seen somewhere by the user.
  • Confirmed: The vulnerability is confirmed from an analyst perspective.
  • Exploited: This vulnerability was exploited and seen by the user reporting the sighting.
  • Patched: This vulnerability was successfully patched by the user reporting the sighting.
  • Not exploited: This vulnerability was not exploited or seen by the user reporting the sighting.
  • Not confirmed: The user expresses doubt about the veracity of the vulnerability.
  • Not patched: This vulnerability was not successfully patched by the user reporting the sighting.