Vulnerabilites related to medtronic - n\'vision_8870_firmware
CVE-2018-10631 (GCVE-0-2018-10631)
Vulnerability from cvelistv5
Published
2018-07-13 19:00
Modified
2025-08-26 14:26
Severity ?
VLAI Severity ?
EPSS score ?
CWE
Summary
The 8840 Clinician Programmer executes the application program from the 8870 Application Card. An attacker with physical access to an 8870 Application Card and sufficient technical capability can modify the contents of this card, including the binary executables. If modified to bypass protection mechanisms, this malicious code will be run when the card is inserted into an 8840 Clinician Programmer.
References
Impacted products
Vendor | Product | Version | |||||||
---|---|---|---|---|---|---|---|---|---|
▼ | Medtronic | 8840 N’Vision Clinician Programmer |
Version: All versions |
||||||
|
{ "containers": { "adp": [ { "providerMetadata": { "dateUpdated": "2024-08-05T07:46:46.502Z", "orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE" }, "references": [ { "tags": [ "x_refsource_MISC", "x_transferred" ], "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-137-01" }, { "tags": [ "x_refsource_MISC", "x_transferred" ], "url": "https://www.medtronic.com/security" } ], "title": "CVE Program Container" }, { "affected": [ { "cpes": [ "cpe:2.3:a:ics_cert:medtronic_n_vision_clinician_programmer:*:*:*:*:*:*:*:*" ], "defaultStatus": "unknown", "product": "medtronic_n_vision_clinician_programmer", "vendor": "ics_cert", "versions": [ { "lessThanOrEqual": "*", "status": "affected", "version": "8840_n_vision_clinician_programmer", "versionType": "custom" }, { "status": "affected", "version": "8870_n_vision_removable_application_card" } ] } ], "metrics": [ { "other": { "content": { "id": "CVE-2018-10631", "options": [ { "Exploitation": "none" }, { "Automatable": "no" }, { "Technical Impact": "total" } ], "role": "CISA Coordinator", "timestamp": "2024-08-19T18:07:51.945413Z", "version": "2.0.3" }, "type": "ssvc" } } ], "providerMetadata": { "dateUpdated": "2025-08-26T14:26:52.840Z", "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0", "shortName": "CISA-ADP" }, "title": "CISA ADP Vulnrichment" } ], "cna": { "affected": [ { "defaultStatus": "unaffected", "product": "8840 N\u2019Vision Clinician Programmer", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "All versions" } ] }, { "defaultStatus": "unaffected", "product": "8870 N\u2019Vision removable Application Card", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "All versions" } ] } ], "credits": [ { "lang": "en", "type": "finder", "value": "Billy Rios and Jonathan Butts of Whitescope LLC reported this vulnerability" } ], "descriptions": [ { "lang": "en", "supportingMedia": [ { "base64": false, "type": "text/html", "value": "The 8840 Clinician Programmer executes the application program from the 8870 Application Card. An attacker with physical access to an 8870 Application Card and sufficient technical capability can modify the contents of this card, including the binary executables. If modified to bypass protection mechanisms, this malicious code will be run when the card is inserted into an 8840 Clinician Programmer." } ], "value": "The 8840 Clinician Programmer executes the application program from the 8870 Application Card. An attacker with physical access to an 8870 Application Card and sufficient technical capability can modify the contents of this card, including the binary executables. If modified to bypass protection mechanisms, this malicious code will be run when the card is inserted into an 8840 Clinician Programmer." } ], "metrics": [ { "cvssV3_1": { "attackComplexity": "HIGH", "attackVector": "PHYSICAL", "availabilityImpact": "HIGH", "baseScore": 6.3, "baseSeverity": "MEDIUM", "confidentialityImpact": "HIGH", "integrityImpact": "HIGH", "privilegesRequired": "NONE", "scope": "UNCHANGED", "userInteraction": "REQUIRED", "vectorString": "CVSS:3.1/AV:P/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:H", "version": "3.1" }, "format": "CVSS", "scenarios": [ { "lang": "en", "value": "GENERAL" } ] } ], "problemTypes": [ { "descriptions": [ { "cweId": "CWE-693", "description": "CWE-693", "lang": "en", "type": "CWE" } ] } ], "providerMetadata": { "dateUpdated": "2025-06-27T16:19:26.179Z", "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert" }, "references": [ { "url": "https://global.medtronic.com/xg-en/product-security/security-bulletins/nvision.html" }, { "url": "https://www.medtronic.com/security" }, { "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-137-01" }, { "url": "http://www.securityfocus.com/bid/104213" } ], "source": { "advisory": "ICSMA-18-137-01", "discovery": "EXTERNAL" }, "title": "Medtronic N\u0027Vision Clinician Programmer Protection Mechanism Failure", "workarounds": [ { "lang": "en", "supportingMedia": [ { "base64": false, "type": "text/html", "value": "\u003cp\u003eMedtronic has not developed a product update to address the vulnerabilities, but is reinforcing security reminders within this advisory to help reduce the risk associated with the vulnerabilities.\u003c/p\u003e\u003cp\u003eThe 8870 Therapy Application card stores PHI and PII as part of its normal operating procedure and should be handled, managed and secured in a manner consistent with the applicable laws for patient data privacy.\u003c/p\u003e\u003cp\u003eMedtronic recommends users take additional defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, hospitals and clinicians should:\u003c/p\u003e\u003cul\u003e\u003cli\u003eMaintain strict physical control of the 8870 application card.\u003c/li\u003e\u003cli\u003eUse only legitimately obtained 8870 cards and not cards provided by any third party as firmware and system updates are provided directly by Medtronic using new 8870 application cards.\u003c/li\u003e\u003cli\u003e8840 Programmers and 8870 Therapy Application compact flash cards are the property of Medtronic and should be returned to Medtronic when no longer in use. If that is not an option, you should securely dispose of them.\u003c/li\u003e\u003c/ul\u003e\u003cp\u003eMedtronic has released additional patient focused information, at the following location:\u003c/p\u003e\u003cp\u003e\u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.medtronic.com/security\"\u003ehttps://www.medtronic.com/security\u003c/a\u003e\u003c/p\u003e" } ], "value": "Medtronic has not developed a product update to address the vulnerabilities, but is reinforcing security reminders within this advisory to help reduce the risk associated with the vulnerabilities.\n\nThe 8870 Therapy Application card stores PHI and PII as part of its normal operating procedure and should be handled, managed and secured in a manner consistent with the applicable laws for patient data privacy.\n\nMedtronic recommends users take additional defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, hospitals and clinicians should:\n\n * Maintain strict physical control of the 8870 application card.\n * Use only legitimately obtained 8870 cards and not cards provided by any third party as firmware and system updates are provided directly by Medtronic using new 8870 application cards.\n * 8840 Programmers and 8870 Therapy Application compact flash cards are the property of Medtronic and should be returned to Medtronic when no longer in use. If that is not an option, you should securely dispose of them.\n\n\nMedtronic has released additional patient focused information, at the following location:\n\n https://www.medtronic.com/security" } ], "x_ConverterErrors": { "DATE_PUBLIC": { "error": "v4 DATE_PUBLIC is invalid", "message": "month must be in 1..12" } }, "x_generator": { "engine": "Vulnogram 0.2.0" }, "x_legacyV4Record": { "CVE_data_meta": { "ASSIGNER": "ics-cert@hq.dhs.gov", "DATE_PUBLIC": "2017-50-17T00:00:00", "ID": "CVE-2018-10631", "STATE": "PUBLIC" }, "affects": { "vendor": { "vendor_data": [ { "product": { "product_data": [ { "product_name": "Medtronic N\u0027Vision Clinician Programmer", "version": { "version_data": [ { "version_value": "8840 N\u0027Vision Clinician Programmer, all versions, and 8870 N\u0027Vision removable Application Card, all versions." } ] } } ] }, "vendor_name": "ICS-CERT" } ] } }, "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": { "description_data": [ { "lang": "eng", "value": "Medtronic N\u0027Vision Clinician Programmer 8840 N\u0027Vision Clinician Programmer, all versions, and 8870 N\u0027Vision removable Application Card, all versions. The 8840 Clinician Programmer executes the application program from the 8870 Application Card. An attacker with physical access to an 8870 Application Card and sufficient technical capability can modify the contents of this card, including the binary executables. If modified to bypass protection mechanisms, this malicious code will be run when the card is inserted into an 8840 Clinician Programmer." } ] }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "eng", "value": "PROTECTION MECHANISM FAILURE CWE-693" } ] } ] }, "references": { "reference_data": [ { "name": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-137-01", "refsource": "MISC", "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-137-01" }, { "name": "https://www.medtronic.com/security", "refsource": "MISC", "url": "https://www.medtronic.com/security" } ] } } } }, "cveMetadata": { "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2018-10631", "datePublished": "2018-07-13T19:00:00.000Z", "dateReserved": "2018-05-01T00:00:00.000Z", "dateUpdated": "2025-08-26T14:26:52.840Z", "state": "PUBLISHED" }, "dataType": "CVE_RECORD", "dataVersion": "5.1" }
CVE-2018-8849 (GCVE-0-2018-8849)
Vulnerability from cvelistv5
Published
2018-05-18 13:00
Modified
2025-06-27 16:24
Severity ?
VLAI Severity ?
EPSS score ?
CWE
- CWE-311 - MISSING ENCRYPTION OF SENSITIVE DATA
Summary
Medtronic N'Vision Clinician Programmer 8840 N'Vision Clinician Programme and 8870 N'Vision removable Application Card do not encrypt PII and PHI while at rest.
References
▼ | URL | Tags |
---|---|---|
http://www.securityfocus.com/bid/104213 | vdb-entry, x_refsource_BID | |
https://www.cisa.gov/news-events/ics-medical-advisories/icsma-18-137-01 | x_refsource_MISC | |
http://www.medtronic.com/content/dam/medtronic-com/us-en/corporate/documents/Medtronic-NVision-8840_Security-Bulletin_FINAL.pdf | x_refsource_CONFIRM | |
https://www.medtronic.com/security |
Impacted products
Vendor | Product | Version | |||||||
---|---|---|---|---|---|---|---|---|---|
▼ | Medtronic | N'Vision Clinician Programmer |
Version: all versions |
||||||
|
{ "containers": { "adp": [ { "providerMetadata": { "dateUpdated": "2024-08-05T07:10:45.899Z", "orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE" }, "references": [ { "name": "104213", "tags": [ "vdb-entry", "x_refsource_BID", "x_transferred" ], "url": "http://www.securityfocus.com/bid/104213" }, { "tags": [ "x_refsource_MISC", "x_transferred" ], "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-137-01" }, { "tags": [ "x_refsource_CONFIRM", "x_transferred" ], "url": "http://www.medtronic.com/content/dam/medtronic-com/us-en/corporate/documents/Medtronic-NVision-8840_Security-Bulletin_FINAL.pdf" } ], "title": "CVE Program Container" } ], "cna": { "affected": [ { "defaultStatus": "unaffected", "product": "N\u0027Vision Clinician Programmer", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "all versions" } ] }, { "defaultStatus": "unaffected", "product": "8870 N\u2019Vision removable Application Card", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "all versions" } ] } ], "credits": [ { "lang": "en", "type": "finder", "value": "Billy Rios and Jonathan Butts of Whitescope LLC reported these vulnerabilities to CISA." } ], "datePublic": "2018-05-17T06:00:00.000Z", "descriptions": [ { "lang": "en", "supportingMedia": [ { "base64": false, "type": "text/html", "value": "\u003cp\u003eMedtronic N\u0027Vision Clinician Programmer 8840 N\u0027Vision Clinician Programme and 8870 N\u0027Vision removable Application Card do not encrypt PII and PHI while at rest.\u003c/p\u003e" } ], "value": "Medtronic N\u0027Vision Clinician Programmer 8840 N\u0027Vision Clinician Programme and 8870 N\u0027Vision removable Application Card do not encrypt PII and PHI while at rest." } ], "metrics": [ { "cvssV3_1": { "attackComplexity": "LOW", "attackVector": "PHYSICAL", "availabilityImpact": "NONE", "baseScore": 4.6, "baseSeverity": "MEDIUM", "confidentialityImpact": "HIGH", "integrityImpact": "NONE", "privilegesRequired": "NONE", "scope": "UNCHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N", "version": "3.1" }, "format": "CVSS", "scenarios": [ { "lang": "en", "value": "GENERAL" } ] } ], "problemTypes": [ { "descriptions": [ { "cweId": "CWE-311", "description": "MISSING ENCRYPTION OF SENSITIVE DATA CWE-311", "lang": "en", "type": "CWE" } ] } ], "providerMetadata": { "dateUpdated": "2025-06-27T16:24:54.910Z", "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert" }, "references": [ { "name": "104213", "tags": [ "vdb-entry", "x_refsource_BID" ], "url": "http://www.securityfocus.com/bid/104213" }, { "tags": [ "x_refsource_MISC" ], "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-18-137-01" }, { "tags": [ "x_refsource_CONFIRM" ], "url": "http://www.medtronic.com/content/dam/medtronic-com/us-en/corporate/documents/Medtronic-NVision-8840_Security-Bulletin_FINAL.pdf" }, { "url": "https://www.medtronic.com/security" } ], "source": { "advisory": "ICSMA-18-137-01", "discovery": "EXTERNAL" }, "title": "Medtronic N\u0027Vision Clinician Programmer Missing Encryption of Sensitive Data", "workarounds": [ { "lang": "en", "supportingMedia": [ { "base64": false, "type": "text/html", "value": "\u003cp\u003eMedtronic has not developed a product update to address the \nvulnerabilities, but is reinforcing security reminders within this \nadvisory to help reduce the risk associated with the vulnerabilities.\u003c/p\u003e\n\u003cp\u003eThe 8870 Therapy Application card stores PHI and PII as part of its \nnormal operating procedure and should be handled, managed and secured in\n a manner consistent with the applicable laws for patient data privacy.\u003c/p\u003e\n\u003cp\u003eMedtronic recommends users take additional defensive measures to \nminimize the risk of exploitation of these vulnerabilities. \nSpecifically, hospitals and clinicians should:\u003c/p\u003e\n\u003cul\u003e\n\u003cli\u003eMaintain strict physical control of the 8870 application card.\u003c/li\u003e\n\u003cli\u003eUse only legitimately obtained 8870 cards and not cards provided by \nany third party as firmware and system updates are provided directly by \nMedtronic using new 8870 application cards.\u003c/li\u003e\n\u003cli\u003e8840 Programmers and 8870 Therapy Application compact flash cards \nare the property of Medtronic and should be returned to Medtronic when \nno longer in use. If that is not an option, you should securely dispose \nof them.\u003c/li\u003e\n\u003c/ul\u003e\n\u003cp\u003eMedtronic has released additional patient focused information, at the following location:\u003c/p\u003e\n\u003cp\u003e\u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.medtronic.com/security\"\u003ehttps://www.medtronic.com/security\u003c/a\u003e\u003c/p\u003e\n\n\u003cbr\u003e" } ], "value": "Medtronic has not developed a product update to address the \nvulnerabilities, but is reinforcing security reminders within this \nadvisory to help reduce the risk associated with the vulnerabilities.\n\n\nThe 8870 Therapy Application card stores PHI and PII as part of its \nnormal operating procedure and should be handled, managed and secured in\n a manner consistent with the applicable laws for patient data privacy.\n\n\nMedtronic recommends users take additional defensive measures to \nminimize the risk of exploitation of these vulnerabilities. \nSpecifically, hospitals and clinicians should:\n\n\n\n * Maintain strict physical control of the 8870 application card.\n\n * Use only legitimately obtained 8870 cards and not cards provided by \nany third party as firmware and system updates are provided directly by \nMedtronic using new 8870 application cards.\n\n * 8840 Programmers and 8870 Therapy Application compact flash cards \nare the property of Medtronic and should be returned to Medtronic when \nno longer in use. If that is not an option, you should securely dispose \nof them.\n\n\n\n\nMedtronic has released additional patient focused information, at the following location:\n\n\n https://www.medtronic.com/security" } ], "x_generator": { "engine": "Vulnogram 0.2.0" }, "x_legacyV4Record": { "CVE_data_meta": { "ASSIGNER": "ics-cert@hq.dhs.gov", "DATE_PUBLIC": "2018-05-17T00:00:00", "ID": "CVE-2018-8849", "STATE": "PUBLIC" }, "affects": { "vendor": { "vendor_data": [ { "product": { "product_data": [ { "product_name": "N\u0027Vision Clinician Programmer", "version": { "version_data": [ { "version_value": "8840 N\u0027Vision Clinician Programmer, all versions" }, { "version_value": "8870 N\u0027Vision removable Application Card, all versions" } ] } } ] }, "vendor_name": "Medtronic" } ] } }, "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": { "description_data": [ { "lang": "eng", "value": "Medtronic N\u0027Vision Clinician Programmer 8840 N\u0027Vision Clinician Programmer, all versions, and 8870 N\u0027Vision removable Application Card, all versions does not encrypt PII and PHI while at rest." } ] }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "eng", "value": "MISSING ENCRYPTION OF SENSITIVE DATA CWE-311" } ] } ] }, "references": { "reference_data": [ { "name": "104213", "refsource": "BID", "url": "http://www.securityfocus.com/bid/104213" }, { "name": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-137-01", "refsource": "MISC", "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-137-01" }, { "name": "http://www.medtronic.com/content/dam/medtronic-com/us-en/corporate/documents/Medtronic-NVision-8840_Security-Bulletin_FINAL.pdf", "refsource": "CONFIRM", "url": "http://www.medtronic.com/content/dam/medtronic-com/us-en/corporate/documents/Medtronic-NVision-8840_Security-Bulletin_FINAL.pdf" } ] } } } }, "cveMetadata": { "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2018-8849", "datePublished": "2018-05-18T13:00:00Z", "dateReserved": "2018-03-20T00:00:00", "dateUpdated": "2025-06-27T16:24:54.910Z", "state": "PUBLISHED" }, "dataType": "CVE_RECORD", "dataVersion": "5.1" }
Vulnerability from fkie_nvd
Published
2018-05-18 13:29
Modified
2025-06-27 17:15
Severity ?
Summary
Medtronic N'Vision Clinician Programmer 8840 N'Vision Clinician Programme and 8870 N'Vision removable Application Card do not encrypt PII and PHI while at rest.
References
Impacted products
Vendor | Product | Version | |
---|---|---|---|
medtronic | n\'vision_8840_firmware | - | |
medtronic | n\'vision_8840 | - | |
medtronic | n\'vision_8870_firmware | - | |
medtronic | n\'vision_8870 | - |
{ "configurations": [ { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:medtronic:n\\\u0027vision_8840_firmware:-:*:*:*:*:*:*:*", "matchCriteriaId": "0CDE1DCE-A7D1-415B-8B50-CEC490D250DA", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:medtronic:n\\\u0027vision_8840:-:*:*:*:*:*:*:*", "matchCriteriaId": "C8999A64-FA2F-48B6-8EE2-35DC311CBEB4", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" }, { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:medtronic:n\\\u0027vision_8870_firmware:-:*:*:*:*:*:*:*", "matchCriteriaId": "568A12CA-DAB3-4797-8223-DC74FA4E8492", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:medtronic:n\\\u0027vision_8870:-:*:*:*:*:*:*:*", "matchCriteriaId": "40349019-CB0F-490B-A767-64E0C38F0E33", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" } ], "cveTags": [], "descriptions": [ { "lang": "en", "value": "Medtronic N\u0027Vision Clinician Programmer 8840 N\u0027Vision Clinician Programme and 8870 N\u0027Vision removable Application Card do not encrypt PII and PHI while at rest." }, { "lang": "es", "value": "Medtronic N\u0027Vision Clinician Programmer 8840 N\u0027Vision Clinician Programmer, en todas las versiones, y 8870 N\u0027Vision removable Application Card, en todas las versiones, no cifran PII y PHI mientras est\u00e1n en reposo." } ], "id": "CVE-2018-8849", "lastModified": "2025-06-27T17:15:32.103", "metrics": { "cvssMetricV2": [ { "acInsufInfo": false, "baseSeverity": "LOW", "cvssData": { "accessComplexity": "LOW", "accessVector": "LOCAL", "authentication": "NONE", "availabilityImpact": "NONE", "baseScore": 2.1, "confidentialityImpact": "PARTIAL", "integrityImpact": "NONE", "vectorString": "AV:L/AC:L/Au:N/C:P/I:N/A:N", "version": "2.0" }, "exploitabilityScore": 3.9, "impactScore": 2.9, "obtainAllPrivilege": false, "obtainOtherPrivilege": false, "obtainUserPrivilege": false, "source": "nvd@nist.gov", "type": "Primary", "userInteractionRequired": false } ], "cvssMetricV30": [ { "cvssData": { "attackComplexity": "LOW", "attackVector": "PHYSICAL", "availabilityImpact": "NONE", "baseScore": 4.6, "baseSeverity": "MEDIUM", "confidentialityImpact": "HIGH", "integrityImpact": "NONE", "privilegesRequired": "NONE", "scope": "UNCHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N", "version": "3.0" }, "exploitabilityScore": 0.9, "impactScore": 3.6, "source": "nvd@nist.gov", "type": "Primary" } ], "cvssMetricV31": [ { "cvssData": { "attackComplexity": "LOW", "attackVector": "PHYSICAL", "availabilityImpact": "NONE", "baseScore": 4.6, "baseSeverity": "MEDIUM", "confidentialityImpact": "HIGH", "integrityImpact": "NONE", "privilegesRequired": "NONE", "scope": "UNCHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N", "version": "3.1" }, "exploitabilityScore": 0.9, "impactScore": 3.6, "source": "ics-cert@hq.dhs.gov", "type": "Secondary" } ] }, "published": "2018-05-18T13:29:00.427", "references": [ { "source": "ics-cert@hq.dhs.gov", "tags": [ "Vendor Advisory" ], "url": "http://www.medtronic.com/content/dam/medtronic-com/us-en/corporate/documents/Medtronic-NVision-8840_Security-Bulletin_FINAL.pdf" }, { "source": "ics-cert@hq.dhs.gov", "tags": [ "Third Party Advisory", "VDB Entry" ], "url": "http://www.securityfocus.com/bid/104213" }, { "source": "ics-cert@hq.dhs.gov", "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-18-137-01" }, { "source": "ics-cert@hq.dhs.gov", "url": "https://www.medtronic.com/security" }, { "source": "af854a3a-2127-422b-91ae-364da2661108", "tags": [ "Vendor Advisory" ], "url": "http://www.medtronic.com/content/dam/medtronic-com/us-en/corporate/documents/Medtronic-NVision-8840_Security-Bulletin_FINAL.pdf" }, { "source": "af854a3a-2127-422b-91ae-364da2661108", "tags": [ "Third Party Advisory", "VDB Entry" ], "url": "http://www.securityfocus.com/bid/104213" }, { "source": "af854a3a-2127-422b-91ae-364da2661108", "tags": [ "Third Party Advisory", "US Government Resource" ], "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-137-01" } ], "sourceIdentifier": "ics-cert@hq.dhs.gov", "vulnStatus": "Modified", "weaknesses": [ { "description": [ { "lang": "en", "value": "CWE-311" } ], "source": "ics-cert@hq.dhs.gov", "type": "Secondary" }, { "description": [ { "lang": "en", "value": "CWE-311" } ], "source": "nvd@nist.gov", "type": "Primary" } ] }
Vulnerability from fkie_nvd
Published
2018-07-13 19:29
Modified
2025-08-26 15:15
Severity ?
Summary
The 8840 Clinician Programmer executes the application program from the 8870 Application Card. An attacker with physical access to an 8870 Application Card and sufficient technical capability can modify the contents of this card, including the binary executables. If modified to bypass protection mechanisms, this malicious code will be run when the card is inserted into an 8840 Clinician Programmer.
References
▼ | URL | Tags | |
---|---|---|---|
ics-cert@hq.dhs.gov | http://www.securityfocus.com/bid/104213 | ||
ics-cert@hq.dhs.gov | https://global.medtronic.com/xg-en/product-security/security-bulletins/nvision.html | ||
ics-cert@hq.dhs.gov | https://ics-cert.us-cert.gov/advisories/ICSMA-18-137-01 | Third Party Advisory, US Government Resource | |
ics-cert@hq.dhs.gov | https://www.medtronic.com/security | Vendor Advisory | |
af854a3a-2127-422b-91ae-364da2661108 | https://ics-cert.us-cert.gov/advisories/ICSMA-18-137-01 | Third Party Advisory, US Government Resource | |
af854a3a-2127-422b-91ae-364da2661108 | https://www.medtronic.com/security | Vendor Advisory |
Impacted products
Vendor | Product | Version | |
---|---|---|---|
medtronic | n\'vision_8840_firmware | - | |
medtronic | n\'vision_8840 | - | |
medtronic | n\'vision_8870_firmware | - | |
medtronic | n\'vision_8870 | - |
{ "configurations": [ { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:medtronic:n\\\u0027vision_8840_firmware:-:*:*:*:*:*:*:*", "matchCriteriaId": "0CDE1DCE-A7D1-415B-8B50-CEC490D250DA", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:medtronic:n\\\u0027vision_8840:-:*:*:*:*:*:*:*", "matchCriteriaId": "C8999A64-FA2F-48B6-8EE2-35DC311CBEB4", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" }, { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:medtronic:n\\\u0027vision_8870_firmware:-:*:*:*:*:*:*:*", "matchCriteriaId": "568A12CA-DAB3-4797-8223-DC74FA4E8492", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:medtronic:n\\\u0027vision_8870:-:*:*:*:*:*:*:*", "matchCriteriaId": "40349019-CB0F-490B-A767-64E0C38F0E33", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" } ], "cveTags": [], "descriptions": [ { "lang": "en", "value": "The 8840 Clinician Programmer executes the application program from the 8870 Application Card. An attacker with physical access to an 8870 Application Card and sufficient technical capability can modify the contents of this card, including the binary executables. If modified to bypass protection mechanisms, this malicious code will be run when the card is inserted into an 8840 Clinician Programmer." }, { "lang": "es", "value": "Medtronic N\u0027Vision Clinician Programmer 8840 N\u0027Vision Clinician Programmer, en todas las versiones, y 8870 N\u0027Vision Application Card extra\u00edble, en todas las versiones. 8840 Clinician Programmer ejecuta el programa de la aplicaci\u00f3n desde la Application Card 8870. Un atacante con acceso f\u00edsico a la Application Card 8870 y la suficiente capacidad t\u00e9cnica puede modificar el contenido de esta tarjeta, incluyendo los ejecutables binarios. Si se modifica para omitir los mecanismos de protecci\u00f3n, este c\u00f3digo malicioso se ejecutar\u00e1 cuando la tarjeta se inserta en un 8840 Clinician Programmer." } ], "id": "CVE-2018-10631", "lastModified": "2025-08-26T15:15:38.060", "metrics": { "cvssMetricV2": [ { "acInsufInfo": false, "baseSeverity": "MEDIUM", "cvssData": { "accessComplexity": "LOW", "accessVector": "LOCAL", "authentication": "NONE", "availabilityImpact": "PARTIAL", "baseScore": 4.6, "confidentialityImpact": "PARTIAL", "integrityImpact": "PARTIAL", "vectorString": "AV:L/AC:L/Au:N/C:P/I:P/A:P", "version": "2.0" }, "exploitabilityScore": 3.9, "impactScore": 6.4, "obtainAllPrivilege": false, "obtainOtherPrivilege": false, "obtainUserPrivilege": false, "source": "nvd@nist.gov", "type": "Primary", "userInteractionRequired": false } ], "cvssMetricV30": [ { "cvssData": { "attackComplexity": "LOW", "attackVector": "PHYSICAL", "availabilityImpact": "HIGH", "baseScore": 6.8, "baseSeverity": "MEDIUM", "confidentialityImpact": "HIGH", "integrityImpact": "HIGH", "privilegesRequired": "NONE", "scope": "UNCHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H", "version": "3.0" }, "exploitabilityScore": 0.9, "impactScore": 5.9, "source": "nvd@nist.gov", "type": "Primary" } ], "cvssMetricV31": [ { "cvssData": { "attackComplexity": "HIGH", "attackVector": "PHYSICAL", "availabilityImpact": "HIGH", "baseScore": 6.3, "baseSeverity": "MEDIUM", "confidentialityImpact": "HIGH", "integrityImpact": "HIGH", "privilegesRequired": "NONE", "scope": "UNCHANGED", "userInteraction": "REQUIRED", "vectorString": "CVSS:3.1/AV:P/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:H", "version": "3.1" }, "exploitabilityScore": 0.4, "impactScore": 5.9, "source": "ics-cert@hq.dhs.gov", "type": "Secondary" } ] }, "published": "2018-07-13T19:29:00.213", "references": [ { "source": "ics-cert@hq.dhs.gov", "url": "http://www.securityfocus.com/bid/104213" }, { "source": "ics-cert@hq.dhs.gov", "url": "https://global.medtronic.com/xg-en/product-security/security-bulletins/nvision.html" }, { "source": "ics-cert@hq.dhs.gov", "tags": [ "Third Party Advisory", "US Government Resource" ], "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-137-01" }, { "source": "ics-cert@hq.dhs.gov", "tags": [ "Vendor Advisory" ], "url": "https://www.medtronic.com/security" }, { "source": "af854a3a-2127-422b-91ae-364da2661108", "tags": [ "Third Party Advisory", "US Government Resource" ], "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-137-01" }, { "source": "af854a3a-2127-422b-91ae-364da2661108", "tags": [ "Vendor Advisory" ], "url": "https://www.medtronic.com/security" } ], "sourceIdentifier": "ics-cert@hq.dhs.gov", "vulnStatus": "Modified", "weaknesses": [ { "description": [ { "lang": "en", "value": "CWE-693" } ], "source": "ics-cert@hq.dhs.gov", "type": "Primary" }, { "description": [ { "lang": "en", "value": "CWE-693" } ], "source": "nvd@nist.gov", "type": "Secondary" } ] }