Refine your search
3 vulnerabilities found for Central Monitor CNS-6201 by NIHON KOHDEN CORPORATION
jvndb-2025-014793
Vulnerability from jvndb
Published
2025-10-01 11:35
Modified
2025-10-27 12:28
Severity ?
Summary
NIHON KOHDEN Central Monitor CNS-6201 vulnerable to NULL pointer dereference
Details
Central Monitor CNS-6201 provided by NIHON KOHDEN CORPORATION contains the following vulnerability.
<ul><li>NULL pointer dereference (CWE-476) - CVE-2025-59668</li></ul>
Jared P. Quinn of QuinnTech.ai discovered and reported the vulnerability to the developer and CISA. Cooperating with CISA, JPCERT/CC coordinated between the reporter and the developer.
References
| Type | URL | |||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
||||||||||||||
Impacted products
| Vendor | Product | |
|---|---|---|
{
"@rdf:about": "https://jvndb.jvn.jp/en/contents/2025/JVNDB-2025-014793.html",
"dc:date": "2025-10-27T12:28+09:00",
"dcterms:issued": "2025-10-01T11:35+09:00",
"dcterms:modified": "2025-10-27T12:28+09:00",
"description": "Central Monitor CNS-6201 provided by NIHON KOHDEN CORPORATION contains the following vulnerability.\r\n\u003cul\u003e\u003cli\u003eNULL pointer dereference (CWE-476) - CVE-2025-59668\u003c/li\u003e\u003c/ul\u003e\r\nJared P. Quinn of QuinnTech.ai discovered and reported the vulnerability to the developer and CISA. Cooperating with CISA, JPCERT/CC coordinated between the reporter and the developer.",
"link": "https://jvndb.jvn.jp/en/contents/2025/JVNDB-2025-014793.html",
"sec:cpe": {
"#text": "cpe:/o:misc:nihon_kohden_central_monitor_cns-6201",
"@product": "Central Monitor CNS-6201",
"@vendor": "NIHON KOHDEN CORPORATION",
"@version": "2.2"
},
"sec:cvss": {
"@score": "7.5",
"@severity": "High",
"@type": "Base",
"@vector": "CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
"@version": "3.0"
},
"sec:identifier": "JVNDB-2025-014793",
"sec:references": [
{
"#text": "https://jvn.jp/en/vu/JVNVU96989989/index.html",
"@id": "JVNVU#96989989",
"@source": "JVN"
},
{
"#text": "https://www.cve.org/CVERecord?id=CVE-2025-59668",
"@id": "CVE-2025-59668",
"@source": "CVE"
},
{
"#text": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-296-01",
"@id": "ICSMA-25-296-01",
"@source": "ICS-CERT ADVISORY"
},
{
"#text": "http://cwe.mitre.org/data/definitions/476.html",
"@id": "CWE-476",
"@title": "NULL Pointer Dereference(CWE-476)"
}
],
"title": "NIHON KOHDEN Central Monitor CNS-6201 vulnerable to NULL pointer dereference"
}
CVE-2025-59668 (GCVE-0-2025-59668)
Vulnerability from nvd
Published
2025-09-30 04:06
Modified
2025-10-31 14:12
Severity ?
7.5 (High) - CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
8.7 (High) - CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N
8.7 (High) - CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N
VLAI Severity ?
EPSS score ?
CWE
- CWE-476 - NULL pointer dereference
Summary
Multiple versions of Central Monitor CNS-6201 contain a NULL pointer dereference vulnerability. When processing a crafted certain UDP packet, the affected device may abnormally terminate.
References
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| NIHON KOHDEN CORPORATION | Central Monitor CNS-6201 |
Version: 01-03 Version: 01-04 Version: 01-05 Version: 01-06 Version: 02-10 Version: 02-11 Version: and 02-40 |
{
"containers": {
"adp": [
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2025-59668",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "partial"
}
],
"role": "CISA Coordinator",
"timestamp": "2025-09-30T15:46:30.521096Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2025-09-30T15:46:39.862Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
},
{
"providerMetadata": {
"dateUpdated": "2025-10-31T14:12:10.064Z",
"orgId": "af854a3a-2127-422b-91ae-364da2661108",
"shortName": "CVE"
},
"references": [
{
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-296-01"
},
{
"url": "https://www.nihonkohden.com/security/main/01112/teaserItems3/0/linkList/0/link/NKcorporateResponse-CNS-6201_CentralMonitor_Vulnerability(CVE-2025-59668)_en_Rev2.pdf"
}
],
"title": "CVE Program Container",
"x_generator": {
"engine": "ADPogram 0.0.1"
}
}
],
"cna": {
"affected": [
{
"product": "Central Monitor CNS-6201",
"vendor": "NIHON KOHDEN CORPORATION",
"versions": [
{
"status": "affected",
"version": "01-03"
},
{
"status": "affected",
"version": "01-04"
},
{
"status": "affected",
"version": "01-05"
},
{
"status": "affected",
"version": "01-06"
},
{
"status": "affected",
"version": "02-10"
},
{
"status": "affected",
"version": "02-11"
},
{
"status": "affected",
"version": "and 02-40"
}
]
}
],
"descriptions": [
{
"lang": "en",
"value": "Multiple versions of Central Monitor CNS-6201 contain a NULL pointer dereference vulnerability. When processing a crafted certain UDP packet, the affected device may abnormally terminate."
}
],
"metrics": [
{
"cvssV3_0": {
"baseScore": 7.5,
"baseSeverity": "HIGH",
"vectorString": "CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
"version": "3.0"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en-US",
"value": "GENERAL"
}
]
},
{
"cvssV4_0": {
"baseScore": 8.7,
"baseSeverity": "HIGH",
"vectorString": "CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N",
"version": "4.0"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en-US",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-476",
"description": "NULL pointer dereference",
"lang": "en-US",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2025-09-30T04:06:11.279Z",
"orgId": "ede6fdc4-6654-4307-a26d-3331c018e2ce",
"shortName": "jpcert"
},
"references": [
{
"url": "https://www.nihonkohden.com/security.html"
},
{
"url": "https://jvn.jp/en/vu/JVNVU96989989/"
}
],
"tags": [
"unsupported-when-assigned"
]
}
},
"cveMetadata": {
"assignerOrgId": "ede6fdc4-6654-4307-a26d-3331c018e2ce",
"assignerShortName": "jpcert",
"cveId": "CVE-2025-59668",
"datePublished": "2025-09-30T04:06:11.279Z",
"dateReserved": "2025-09-18T07:00:43.823Z",
"dateUpdated": "2025-10-31T14:12:10.064Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.2"
}
CVE-2025-59668 (GCVE-0-2025-59668)
Vulnerability from cvelistv5
Published
2025-09-30 04:06
Modified
2025-10-31 14:12
Severity ?
7.5 (High) - CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
8.7 (High) - CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N
8.7 (High) - CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N
VLAI Severity ?
EPSS score ?
CWE
- CWE-476 - NULL pointer dereference
Summary
Multiple versions of Central Monitor CNS-6201 contain a NULL pointer dereference vulnerability. When processing a crafted certain UDP packet, the affected device may abnormally terminate.
References
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| NIHON KOHDEN CORPORATION | Central Monitor CNS-6201 |
Version: 01-03 Version: 01-04 Version: 01-05 Version: 01-06 Version: 02-10 Version: 02-11 Version: and 02-40 |
{
"containers": {
"adp": [
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2025-59668",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "partial"
}
],
"role": "CISA Coordinator",
"timestamp": "2025-09-30T15:46:30.521096Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2025-09-30T15:46:39.862Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
},
{
"providerMetadata": {
"dateUpdated": "2025-10-31T14:12:10.064Z",
"orgId": "af854a3a-2127-422b-91ae-364da2661108",
"shortName": "CVE"
},
"references": [
{
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-296-01"
},
{
"url": "https://www.nihonkohden.com/security/main/01112/teaserItems3/0/linkList/0/link/NKcorporateResponse-CNS-6201_CentralMonitor_Vulnerability(CVE-2025-59668)_en_Rev2.pdf"
}
],
"title": "CVE Program Container",
"x_generator": {
"engine": "ADPogram 0.0.1"
}
}
],
"cna": {
"affected": [
{
"product": "Central Monitor CNS-6201",
"vendor": "NIHON KOHDEN CORPORATION",
"versions": [
{
"status": "affected",
"version": "01-03"
},
{
"status": "affected",
"version": "01-04"
},
{
"status": "affected",
"version": "01-05"
},
{
"status": "affected",
"version": "01-06"
},
{
"status": "affected",
"version": "02-10"
},
{
"status": "affected",
"version": "02-11"
},
{
"status": "affected",
"version": "and 02-40"
}
]
}
],
"descriptions": [
{
"lang": "en",
"value": "Multiple versions of Central Monitor CNS-6201 contain a NULL pointer dereference vulnerability. When processing a crafted certain UDP packet, the affected device may abnormally terminate."
}
],
"metrics": [
{
"cvssV3_0": {
"baseScore": 7.5,
"baseSeverity": "HIGH",
"vectorString": "CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
"version": "3.0"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en-US",
"value": "GENERAL"
}
]
},
{
"cvssV4_0": {
"baseScore": 8.7,
"baseSeverity": "HIGH",
"vectorString": "CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N",
"version": "4.0"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en-US",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-476",
"description": "NULL pointer dereference",
"lang": "en-US",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2025-09-30T04:06:11.279Z",
"orgId": "ede6fdc4-6654-4307-a26d-3331c018e2ce",
"shortName": "jpcert"
},
"references": [
{
"url": "https://www.nihonkohden.com/security.html"
},
{
"url": "https://jvn.jp/en/vu/JVNVU96989989/"
}
],
"tags": [
"unsupported-when-assigned"
]
}
},
"cveMetadata": {
"assignerOrgId": "ede6fdc4-6654-4307-a26d-3331c018e2ce",
"assignerShortName": "jpcert",
"cveId": "CVE-2025-59668",
"datePublished": "2025-09-30T04:06:11.279Z",
"dateReserved": "2025-09-18T07:00:43.823Z",
"dateUpdated": "2025-10-31T14:12:10.064Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.2"
}