Vulnerabilites related to Becton Dickinson & Co - BD Alarisâ„¢ Point-of-Care Unit (PCU) Model 8015
CVE-2023-30561 (GCVE-0-2023-30561)
Vulnerability from cvelistv5
Published
2023-07-13 19:03
Modified
2024-10-22 16:07
Severity ?
VLAI Severity ?
EPSS score ?
CWE
- CWE-311 - Missing Encryption of Sensitive Data
Summary
The data flowing between the PCU and its modules is insecure. A threat actor with physical access could potentially read or modify data by attaching a specially crafted device while an infusion is running.
References
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Becton Dickinson & Co | BD Alarisâ„¢ Point-of-Care Unit (PCU) Model 8015 |
Version: 0 < |
{
"containers": {
"adp": [
{
"providerMetadata": {
"dateUpdated": "2024-08-02T14:28:51.672Z",
"orgId": "af854a3a-2127-422b-91ae-364da2661108",
"shortName": "CVE"
},
"references": [
{
"tags": [
"x_transferred"
],
"url": "https://www.bd.com/en-us/about-bd/cybersecurity/bulletin/bd-alaris-system-with-guardrails-suite-mx"
}
],
"title": "CVE Program Container"
},
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2023-30561",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "partial"
}
],
"role": "CISA Coordinator",
"timestamp": "2024-10-22T15:49:18.852817Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2024-10-22T16:07:16.822Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "unaffected",
"product": "BD Alaris\u00e2\u201e\u00a2 Point-of-Care Unit (PCU) Model 8015",
"vendor": "Becton Dickinson \u0026 Co",
"versions": [
{
"lessThanOrEqual": "12.1.3",
"status": "affected",
"version": "0",
"versionType": "custom"
}
]
}
],
"datePublic": "2023-07-13T18:56:00.000Z",
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "The data flowing between the PCU and its modules is insecure. A threat actor with physical access could potentially read or modify data by attaching a specially crafted device while an infusion is running."
}
],
"value": "The data flowing between the PCU and its modules is insecure. A threat actor with physical access could potentially read or modify data by attaching a specially crafted device while an infusion is running."
}
],
"impacts": [
{
"capecId": "CAPEC-390",
"descriptions": [
{
"lang": "en",
"value": "CAPEC-390 Bypassing Physical Security"
}
]
},
{
"capecId": "CAPEC-94",
"descriptions": [
{
"lang": "en",
"value": "CAPEC-94 Man in the Middle Attack"
}
]
}
],
"metrics": [
{
"cvssV3_1": {
"attackComplexity": "LOW",
"attackVector": "PHYSICAL",
"availabilityImpact": "HIGH",
"baseScore": 6.1,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "HIGH",
"integrityImpact": "NONE",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:H",
"version": "3.1"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-311",
"description": "CWE-311 Missing Encryption of Sensitive Data",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2023-07-13T19:03:17.356Z",
"orgId": "2325d071-eabf-4b7b-a4ea-0819b6629a18",
"shortName": "BD"
},
"references": [
{
"url": "https://www.bd.com/en-us/about-bd/cybersecurity/bulletin/bd-alaris-system-with-guardrails-suite-mx"
}
],
"source": {
"discovery": "INTERNAL"
},
"title": "Lack of Cryptographic Security of IUI Bus ",
"x_generator": {
"engine": "Vulnogram 0.1.0-dev"
}
}
},
"cveMetadata": {
"assignerOrgId": "2325d071-eabf-4b7b-a4ea-0819b6629a18",
"assignerShortName": "BD",
"cveId": "CVE-2023-30561",
"datePublished": "2023-07-13T19:03:17.356Z",
"dateReserved": "2023-04-12T16:30:07.537Z",
"dateUpdated": "2024-10-22T16:07:16.822Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}
CVE-2023-30560 (GCVE-0-2023-30560)
Vulnerability from cvelistv5
Published
2023-07-13 18:53
Modified
2024-10-31 17:33
Severity ?
VLAI Severity ?
EPSS score ?
CWE
- CWE-287 - Improper Authentication
Summary
The configuration from the PCU can be modified without authentication using physical connection to the PCU.
References
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Becton Dickinson & Co | BD Alarisâ„¢ Point-of-Care Unit (PCU) Model 8015 |
Version: 0 < |
{
"containers": {
"adp": [
{
"providerMetadata": {
"dateUpdated": "2024-08-02T14:28:51.941Z",
"orgId": "af854a3a-2127-422b-91ae-364da2661108",
"shortName": "CVE"
},
"references": [
{
"tags": [
"x_transferred"
],
"url": "https://www.bd.com/en-us/about-bd/cybersecurity/bulletin/bd-alaris-system-with-guardrails-suite-mx"
}
],
"title": "CVE Program Container"
},
{
"affected": [
{
"cpes": [
"cpe:2.3:h:becton_dickinson_and_co:bd_alarisa_point_of_care_unit_model_8015:*:*:*:*:*:*:*:*"
],
"defaultStatus": "unknown",
"product": "bd_alarisa_point_of_care_unit_model_8015",
"vendor": "becton_dickinson_and_co",
"versions": [
{
"status": "affected",
"version": "0"
}
]
}
],
"metrics": [
{
"other": {
"content": {
"id": "CVE-2023-30560",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "total"
}
],
"role": "CISA Coordinator",
"timestamp": "2024-10-31T17:29:20.439171Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2024-10-31T17:33:02.007Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "unaffected",
"product": "BD Alaris\u00e2\u201e\u00a2 Point-of-Care Unit (PCU) Model 8015",
"vendor": "Becton Dickinson \u0026 Co ",
"versions": [
{
"lessThanOrEqual": "12.1.3",
"status": "affected",
"version": "0",
"versionType": "custom"
}
]
}
],
"datePublic": "2023-07-13T18:00:00.000Z",
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "\u003cp\u003eThe configuration from the PCU can be modified without authentication using physical connection to the PCU. \u003c/p\u003e\n\n\n\n\n\n"
}
],
"value": "The configuration from the PCU can be modified without authentication using physical connection to the PCU. \n\n\n\n\n\n\n\n"
}
],
"impacts": [
{
"capecId": "CAPEC-114",
"descriptions": [
{
"lang": "en",
"value": "CAPEC-114 Authentication Abuse"
}
]
}
],
"metrics": [
{
"cvssV3_1": {
"attackComplexity": "LOW",
"attackVector": "PHYSICAL",
"availabilityImpact": "HIGH",
"baseScore": 6.8,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
"version": "3.1"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-287",
"description": "CWE-287 Improper Authentication",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2023-07-13T18:53:49.951Z",
"orgId": "2325d071-eabf-4b7b-a4ea-0819b6629a18",
"shortName": "BD"
},
"references": [
{
"url": "https://www.bd.com/en-us/about-bd/cybersecurity/bulletin/bd-alaris-system-with-guardrails-suite-mx"
}
],
"source": {
"discovery": "INTERNAL"
},
"title": " PCU Configuration Lacks Authentication",
"x_generator": {
"engine": "Vulnogram 0.1.0-dev"
}
}
},
"cveMetadata": {
"assignerOrgId": "2325d071-eabf-4b7b-a4ea-0819b6629a18",
"assignerShortName": "BD",
"cveId": "CVE-2023-30560",
"datePublished": "2023-07-13T18:53:49.951Z",
"dateReserved": "2023-04-12T16:30:07.536Z",
"dateUpdated": "2024-10-31T17:33:02.007Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}