{"uuid": "ff9b6b9b-871d-4be2-b747-734f1413c1d3", "vulnerability_lookup_origin": "1a89b78e-f703-45f3-bb86-59eb712668bd", "author": "9f56dd64-161d-43a6-b9c3-555944290a09", "vulnerability": "CVE-2019-6538", "type": "seen", "source": "https://t.me/DarkWebInformer_CVEAlerts/17347", "content": "\ud83d\udd17 DarkWebInformer.com - Cyber Threat Intelligence\n\ud83d\udccc CVE ID: CVE-2019-6538\n\ud83d\udd25 CVSS Score: 9.3 (cvssV3_1, Vector: CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:H)\n\ud83d\udd39 Description: The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product\u2019s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.\n\ud83d\udccf Published: 2019-03-25T21:26:03\n\ud83d\udccf Modified: 2025-05-22T19:26:15.126Z\n\ud83d\udd17 References:\n1. https://ics-cert.us-cert.gov/advisories/ICSMA-19-080-01\n2. http://www.securityfocus.com/bid/107544", "creation_timestamp": "2025-05-22T19:45:55.000000Z"}