Max CVSS 7.5 Min CVSS 4.3 Total Count2
IDCVSSSummaryLast (major) updatePublished
CVE-2017-12718 6.8
A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump does not verify input buffer size prior to copying, leading to a buffe
09-10-2019 - 23:23 15-02-2018 - 10:29
CVE-2017-12720 6.8
An Improper Access Control issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The FTP server on the pump does not require authentication if the pump is configured to allow FTP connections.
03-10-2019 - 00:03 15-02-2018 - 10:29
CVE-2017-12721 4.3
An Improper Certificate Validation issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump does not validate host certificates, leaving the pump vulnerable to a man-in-the-middle (MITM
02-03-2018 - 16:10 15-02-2018 - 10:29
CVE-2017-12722 5.0
An Out-of-bounds Read issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump reads memory out of bounds, causing the communications module to crash. Sm
02-03-2018 - 16:09 15-02-2018 - 10:29
CVE-2017-12723 4.3
A Password in Configuration File issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump stores some passwords in the configuration file, which are accessible if the pump is configured
02-03-2018 - 14:48 15-02-2018 - 10:29
CVE-2017-12725 6.8
A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wire
02-03-2018 - 14:40 15-02-2018 - 10:29
CVE-2017-12724 6.8
A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The FTP server on the pump contains hardcoded credentials, which are not fully initialized. The FTP serve
02-03-2018 - 14:40 15-02-2018 - 10:29
CVE-2017-12726 7.5
A Use of Hard-coded Password issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. Telnet on the pump uses hardcoded credentials, which can be used if the pump is configured to allow external
02-03-2018 - 14:39 15-02-2018 - 10:29
Back to Top Mark selected