IDCVSSSummaryLast (major) updatePublished
CVE-2014-5431 4.6
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuratio
09-10-2019 - 23:11 26-03-2019 - 16:29
CVE-2014-5432 7.5
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication. A remote attacker may be able to make unauthorized configuration changes
09-10-2019 - 23:11 26-03-2019 - 16:29
CVE-2014-5433 7.5
An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16,
09-10-2019 - 23:11 26-03-2019 - 16:29
CVE-2014-5434 5.0
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 has a default account with hard-coded credentials used with the FTP protocol. Baxter asserts no files can be transferred to or from the
09-10-2019 - 23:11 26-03-2019 - 15:29