ID CVE-2020-12048
Summary Phoenix Hemodialysis Delivery System SW 3.36 and 3.40, The Phoenix Hemodialysis device does not support data-in-transit encryption (e.g., TLS/SSL) when transmitting treatment and prescription data on the network between the Phoenix system and the Exalis dialysis data management tool. An attacker with access to the network could observe sensitive treatment and prescription data sent between the Phoenix system and the Exalis tool.
References
Vulnerable Configurations
CVSS
Base: 5.0
Impact:
Exploitability:
Access
VectorComplexityAuthentication
Impact
ConfidentialityIntegrityAvailability
Last major update 29-06-2020 - 14:59
Published 29-06-2020 - 14:15
Last modified 29-06-2020 - 14:59
Back to Top