ID CVE-2017-5149
Summary An issue was discovered in St. Jude Medical Merlin@home, versions prior to Version 8.2.2 (RF models: EX1150; Inductive models: EX1100; and Inductive models: EX1100 with MerlinOnDemand capability). The identities of the endpoints for the communication channel between the transmitter and St. Jude Medical's web site, Merlin.net, are not verified. This may allow a man-in-the-middle attacker to access or influence communications between the identified endpoints.
References
Vulnerable Configurations
  • cpe:2.3:o:abbott:merlin\@home_firmware:*:*:*:*:*:*:*:*
    cpe:2.3:o:abbott:merlin\@home_firmware:*:*:*:*:*:*:*:*
  • cpe:2.3:h:abbott:merlin\@home_ex1100:-:*:*:*:*:*:*:*
    cpe:2.3:h:abbott:merlin\@home_ex1100:-:*:*:*:*:*:*:*
  • cpe:2.3:h:abbott:merlin\@home_ex1150:-:*:*:*:*:*:*:*
    cpe:2.3:h:abbott:merlin\@home_ex1150:-:*:*:*:*:*:*:*
CVSS
Base: 6.8 (as of 26-06-2023 - 19:38)
Impact:
Exploitability:
CWE CWE-476
CAPEC
Access
VectorComplexityAuthentication
NETWORK MEDIUM NONE
Impact
ConfidentialityIntegrityAvailability
PARTIAL PARTIAL PARTIAL
cvss-vector via4 AV:N/AC:M/Au:N/C:P/I:P/A:P
refmap via4
bid 95331
misc https://ics-cert.us-cert.gov/advisories/ICSMA-17-009-01A
Last major update 26-06-2023 - 19:38
Published 13-02-2017 - 22:59
Last modified 26-06-2023 - 19:38